journal article·2022-09-15Open access

Requirements of Scintillation Crystals with the Development of PET Scanners

Xin Yu, Xi Zhang, Heng Zhang, Hao Peng, Qiushi Ren, Jianfeng Xu, Qiyu Peng, Siwei Xie

Crystals· Volume 12· Issue 9

Abstract

Positron emission tomography (PET) is widely used in the diagnosis of tumors, cardiovascular system diseases, and neurological diseases. Scintillation crystals are an important part of PET scanners; they can convert γ photons into fluorescent photons to obtain their energy, time, and position information. Currently, an important research goal in PET is to find scintillation crystals with better performance. In this work, the principle of scintillation crystals is introduced, and the properties and requirements of scintillation crystals in different PET scanners are analyzed. At present, Lu2(1−x)Y2xSiO5 (LYSO) is the scintillation crystal with the best comprehensive properties. LaBr3 performs even better regarding the timing characteristics and light output; however, LaBr3 has not been used in any PET scanner because of its deliquescence. Detectors made of Gd3(Ga, Al)5O12 (GAGG) exhibit a high depth of interaction (DOI) resolution and have considerable application potential. The application fields of PET are constantly expanding, and its future development aims to achieve high spatial resolution and high sensitivity, which require scintillation crystals with better performance.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Radiation Detection and Scintillator Technologies · Medical Imaging Techniques and Applications · Atomic and Subatomic Physics Research

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

School of Mechanical Science and Engineering, Huazhong University of Science and Technology, 1037 Luoyu Road, Wuhan 430074, China

School of Physics and Technology, Wuhan University, 299 Bayi Road, Wuhan 430074, China

Institute of Biomedical Engineering, Shenzhen Bay Laboratory, Shenzhen 518107, China

Department of Biomedical Engineering, College of Future Technology, Peking University, Beijing 100871, China

Peking University Shenzhen Graduate School, Shenzhen 518055, China

Shenzhen Alpha Laboratory Technology Ltd., Shenzhen 518052, China

Funding

Shenzhen Science and Technology Program · Guangdong Basic and Applied Basic Research Foundation · State Key Lab of Digital Manufacturing Equipment & Technology Open Project Funding Project · Nanshan District Science and Technology Program

References (100+)

Show source-provided references
1. Scintillation properties of organic–inorganic layered perovskite nanocrystals in glass10.1063/5.0007084 ↗
2. A solution-processed zero-dimensional all-inorganic perovskite scintillator for high resolution gamma-ray spectroscopy detection10.1039/D0NR00772B ↗
3. Development of inorganic scintillators10.1016/S0168-9002(97)00239-8 ↗
4. Note on the effect of extreme cold on the emanations of radium10.1098/rspl.1903.0018 ↗
5. Inorganic crystals for the detection of high energy particles and quanta10.1103/PhysRev.73.1210 ↗
6. The detection of gamma-rays with thallium-activated sodium iodide crystals10.1103/PhysRev.75.796 ↗
7. Alkali halide scintillation counters10.1103/PhysRev.74.100 ↗
8. A photoelectric method for the measurement of spectra of light sources of rapidly varying intensities10.1016/0003-2697(60)90019-1 ↗
9. Gektin, A., and Korzhik, M. (2017). Inorganic Scintillators for Detector Systems, Springer.
10. Development of new scintillators for medical applications10.1016/j.nima.2015.08.041 ↗
11. Inorganic scintillators: Today and tomorrow10.1016/S0022-2313(02)00423-4 ↗
12. Williams, R.T., Grim, J.Q., Li, Q., Ucer, K.B., Bizarri, G.A., and Burger, A. (2015). Scintillation Detectors of Radiation: Excitations at High Densities and Strong Gradients. Excitonic and Photonic Processes in Materials, Springer.10.1007/978-981-287-131-2_10 ↗
13. Robertson, J.S., Marr, R.B., Rosenblum, M., Radeka, V., and Yamamoto, Y.L. (1972). 32-Crystal Positron Transverse Section Detector (No. BNL-17237; CONF-720959-1), Brookhaven National Lab.10.2172/4622519 ↗
14. Instrumentation trends in nuclear medicine10.1016/S0001-2998(77)80037-8 ↗
15. Attenuation correction for a combined 3D PET/CT scanner10.1118/1.598392 ↗
16. Scintillator materials10.1021/cm00045a002 ↗
17. Positron emission tomography10.1088/0031-9155/51/13/R08 ↗
18. Scintillation mechanisms of inorganic materials: From crystal characteristics to scintillation properties10.1016/j.jcrysgro.2009.12.063 ↗
19. Scintillation detectors for X-rays10.1088/0957-0233/17/4/R01 ↗
20. Recent R&D trends in inorganic single-crystal scintillator materials for radiation detection10.1002/adom.201400571 ↗
21. Efficiency of electron-hole pair production in scintillators10.1016/0022-2313(96)00026-9 ↗
22. Rodnyi, P.A. (1997). Physical Processes in Inorganic Scintillators, CRC Press.
23. Inorganic scintillating materials and scintillation detectors10.2183/pjab.94.007 ↗
24. The quest for the ideal inorganic scintillator10.1016/S0168-9002(03)01031-3 ↗
25. Improving timing performance of double-ended readout in TOF-PET detectors10.1088/1748-0221/15/01/P01003 ↗
26. A GSO depth of interaction detector for PET10.1109/23.681982 ↗
27. Melcher, C.L. (1991). Lutetium orthosilicate single crystal scintillator detector. (5,025,151), U.S. Patent.
28. Recent developments in PET detector technology10.1088/0031-9155/53/17/R01 ↗
29. High-energy-resolution scintillator: Ce3+ activated LaBr 310.1063/1.1385342 ↗
30. 2 inch diameter single crystal growth and scintillation properties of Ce: Gd3Al2Ga3O1210.1016/j.jcrysgro.2011.11.085 ↗
31. Judd-Ofelt Analysis of Spectroscopic Properties of Tm3+ Doped Lu2SiO5 Crystals10.3788/CJL20083504.0601 ↗
32. Prostate cancer: PET with 18F-FDG, 18F-or 11C-acetate, and 18F-or 11C-choline10.2967/jnumed.110.077941 ↗
33. Chemistry for positron emission tomography: Recent advances in 11C, 18F, 13N, and 15O labeling reactions10.1002/anie.201805501 ↗
34. A γ-photon detector based on liquid light guide for whole-body PET10.1088/1748-0221/12/11/P11012 ↗
35. Diagnosis of bone metastases: A meta-analysis comparing 18FDG PET, CT, MRI and bone scintigraphy10.1007/s00330-011-2221-4 ↗
36. Comparison of PET, MRI, and CT with pathology in a proven case of Alzheimer’s disease10.1212/WNL.36.12.1569 ↗
37. Novel preclinical PET geometrical concept using a monolithic scintillator crystal offering concurrent enhancement in spatial resolution and detection sensitivity: A simulation study10.1088/1361-6560/ab63ef ↗
38. Spatial resolution and sensitivity of the Inveon small-animal PET scanner10.2967/jnumed.108.055152 ↗
39. Progress and future trends in PET/CT and PET/MRI molecular imaging approaches for breast cancer10.3389/fonc.2020.01301 ↗
40. An investigation of sensitivity limits in PET scanners10.1016/j.nima.2007.06.112 ↗
41. Recent developments in PET instrumentation10.2174/138920110792246555 ↗
42. Dahlbom, M. (2021). PET Imaging: Basic and New Trends. Handbook of Particle Detection and Imaging, Springer International Publishing.10.1007/978-3-319-93785-4_38 ↗
43. Salomoni, M., Pots, R., Auffray, E., and Lecoq, P. (2018). Enhancing light extraction of inorganic scintillators using photonic crystal. Crystals, 8.10.3390/cryst8020078 ↗
44. Improving light output and coincidence time resolution of scintillating crystals using nanoimprinted photonic crystal slabs10.1016/j.nima.2019.06.026 ↗
45. Fundamental limits of spatial resolution in PET10.1016/j.nima.2010.11.092 ↗
46. Design study of a PET detector with 0.5 mm crystal pitch for high-resolution preclinical imaging10.1088/1361-6560/ac0b82 ↗
47. Breast cancer: PET imaging of estrogen receptors10.1148/radiology.169.1.3262228 ↗
48. PET in lung cancer
49. Charged hadron tumour therapy monitoring by means of PET10.1016/j.nima.2004.03.128 ↗
50. Cardiac PET imaging for the detection and monitoring of coronary artery disease and microvascular health10.1016/j.jcmg.2010.04.007 ↗
51. Phelps, M.E. (2004). PET: Molecular Imaging and its Biological Applications, Springer Science Business.
52. Assessment of cardiac sympathetic neuronal function using PET imaging10.1016/j.nuclcard.2004.06.133 ↗
53. The use of PET in Alzheimer disease10.1038/nrneurol.2009.217 ↗
54. FDG-PET improves tumour detection in patients with paraneoplastic neurological syndromes10.1093/brain/awh247 ↗
55. Performance evaluation of the new whole-body PET/CT scanner: Discovery ST10.1007/s00259-003-1444-2 ↗
56. Imaging characteristics of a three-dimensional GSO whole-body PET camera
57. 2011 Physical and clinical performance of the mCT time-of-flight PET/CT scanner10.1088/0031-9155/56/8/004 ↗
58. Why is TOF PET reconstruction a more robust method in the presence of inconsistent data?10.1088/0031-9155/56/1/010 ↗
59. Factors influencing time resolution of scintillators and ways to improve them10.1109/TNS.2010.2049860 ↗
60. Vinke, R., Seifert, S., Schaart, D.R., Schreuder, F.P., de Boer, M.R., van Dam, H.T., Beekman, F.J., Löhner, H., and Dendooven, P. (November, January 24). Optimization of digital time pickoff methods for LaBr 3-SiPM TOF-PET detectors. Proceedings of the 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC), Orlando, FL, USA.10.1109/NSSMIC.2009.5401598 ↗
61. Design of a lanthanum bromide detector for time-of-flight PET10.1109/TNS.2004.835777 ↗
62. Karp, J.S., Wiener, R., Surti, S., Schmall, J.P., Ferri, A., Gola, A., Tarolli, A., and Piemonte, C. (November, January 27). Timing and energy resolution of new near-UV SiPMs coupled to LaBr3: Ce for TOF-PET. Proceedings of the 2013 IEEE Nuclear Science Symposium and Medical Imaging Conference (2013 NSS/MIC), Seoul, Korea.
63. Xie, S., Zhang, X., Zhang, Y., Ying, G., Huang, Q., Xu, J., and Peng, Q. (2020). Evaluation of various scintillator materials in radiation detector design for positron emission tomography (PET). Crystals, 10.10.3390/cryst10100869 ↗
64. The imaging performance of a LaBr3-based PET scanner
65. Preclinical Evaluation and Pilot Clinical Study of Al18F-PSMA-BCH for Prostate Cancer PET Imaging10.2967/jnumed.118.221671 ↗
66. Small-animal PET imaging of human epidermal growth factor receptor positive tumor with a 64Cu labeled affibody protein10.1021/bc900515p ↗
67. Small-animal molecular imaging methods
68. Small animal PET scanner based on monolithic LYSO crystals: Performance evaluation10.1118/1.3673771 ↗
69. A prototype high-resolution small-animal PET scanner dedicated to mouse brain imaging10.2967/jnumed.115.165886 ↗
70. Xia, Y., Ma, T., Liu, Y., Wang, S., and Shao, Y. (2011, January 23–29). A modified spatial resolution formula for DOI-PET. Proceedings of the 2011 IEEE Nuclear Science Symposium Conference Record, Valencia, Spain.10.1109/NSSMIC.2011.6152705 ↗
71. Theoretical study of the benefit of long axial field-of-view PET on region of interest quantification10.1088/1361-6560/aac815 ↗
72. A four-layer DOI detector with a relative offset for use in an animal PET system10.1109/TNS.2010.2044892 ↗
73. Characterization of stacked-crystal PET detector designs for measurement of both TOF and DOI10.1088/0031-9155/60/9/3549 ↗
74. Effects of reflector and crystal surface on the performance of a depth-encoding PET detector with dual-ended readout10.1118/1.4881097 ↗
75. Design and simulation of a novel method for determining depth-of-interaction in a PET scintillation crystal array using a single-ended readout by a multi-anode PMT10.1088/0031-9155/55/13/017 ↗
76. A depth encoding PET detector using four-crystals-to-one-SiPM coupling and light-sharing window method10.1002/mp.13603 ↗
77. Performance of a high-resolution depth encoding PET detector using barium sulfate reflector10.1088/1361-6560/aa71f3 ↗
78. A PET detector prototype based on digital SiPMs and GAGG scintillators10.1088/0031-9155/60/4/1667 ↗
79. Recovery of inter-detector and inter-crystal scattering in brain PET based on LSO and GAGG crystals10.1088/1361-6560/ab9f5c ↗
80. Evaluation of dual-ended readout GAGG-based DOI-PET detectors with different surface treatments10.1002/mp.14874 ↗
81. The 2006 Henry, N. Wagner Lecture: Of mice and men (and positrons)—advances in PET imaging technology
82. First human imaging studies with the EXPLORER total-body PET scanner10.2967/jnumed.119.226498 ↗
83. Optimal whole-body PET scanner configurations for different volumes of LSO scintillator: A simulation study10.1088/0031-9155/57/13/4077 ↗
84. Quantitative image reconstruction for total-body PET imaging using the 2-meter long EXPLORER scanner10.1088/1361-6560/aa5e46 ↗
85. Quantitative accuracy in total-body imaging using the uEXPLORER PET/CT scanner10.1088/1361-6560/ac287c ↗
86. Performance evaluation of the uEXPLORER total-body PET/CT scanner based on NEMA NU 2-2018 with additional tests to characterize PET scanners with a long axial field of view10.2967/jnumed.120.250597 ↗
87. Performance characteristics of the digital biograph vision PET/CT system10.2967/jnumed.118.215418 ↗
88. Prasad, R., and Supanich, M. (2020, January 15). Performance Characterization of 3D PET System for High Count Dynamic Cardiac Imaging. Proceedings of the European Congress of Radiology-ECR 2020, Vienna, Austria.
89. Performance evaluation of the 5-Ring GE Discovery MI PET/CT system using the national electrical manufacturers association NU 2-2012 Standard10.1002/mp.13576 ↗
90. Comparison of NEMA characterizations for Discovery MI and Discovery MI-DR TOF PET/CT systems at different sites and with other commercial PET/CT systems10.1186/s40658-020-0271-x ↗
91. Discovery molecular imaging digital ready PET/CT performance evaluation according to the NEMA NU2-2012 standard10.1097/MNM.0000000000000962 ↗
92. Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities
93. Rapid feasibility studies of tracers for positron emission tomography: High-resolution PET in small animals with kinetic analysis10.1038/jcbfm.1991.157 ↗
94. Performance evaluation of microPET: A high-resolution lutetium oxyorthosilicate PET scanner for animal imaging
95. Performance evaluation of the Philips MOSAIC small animal PET scanner10.1007/s00259-006-0271-7 ↗
96. Raytest ClearPETTM, a new generation small animal PET scanner10.1016/j.nima.2006.10.143 ↗
97. Performance evaluation of the FLEX triumph X-PET scanner using the national electrical manufacturers association NU-4 standards10.2967/jnumed.110.076125 ↗
98. Performance evaluation of the Trans-PET® BioCaliburn® LH system: A large FOV small-animal PET system10.1088/0031-9155/60/1/137 ↗
99. Performance characteristics of a novel clustered multi-pinhole technology for simultaneous high-resolution SPECT/PET10.1007/s12149-015-0966-6 ↗
100. Performance evaluation of the MOLECUBES β-CUBE—a high spatial resolution and high sensitivity small animal PET scanner utilizing monolithic LYSO scintillation detectors10.1088/1361-6560/aacec3 ↗