The Resolution in X-ray Crystallography and Single-Particle Cryogenic Electron Microscopy
Source abstract
Abstract
X-ray crystallography and single-particle analysis cryogenic electron microscopy are essential techniques for uncovering the three-dimensional structures of biological macromolecules. Both techniques rely on the Fourier transform to calculate experimental maps. However, one of the crucial parameters, resolution, is rather broadly defined. Here, the methods to determine the resolution in X-ray crystallography and single-particle analysis are summarized. In X-ray crystallography, it is becoming increasingly more common to include reflections discarded previously by traditionally used standards, allowing for the inclusion of incomplete and anisotropic reflections into the refinement process. In general, the resolution is the smallest lattice spacing given by Bragg’s law for a particular set of X-ray diffraction intensities; however, typically the resolution is truncated by the user during the data processing based on certain parameters and later it is used during refinement. However, at which resolution to perform such a truncation is not always clear and this makes it very confusing for the novices entering the structural biology field. Furthermore, it is argued that the effective resolution should be also reported as it is a more descriptive measure accounting for anisotropy and incompleteness of the data. In single particle cryo-EM, the situation is not much better, as multiple ways exist to determine the resolution, such as Fourier shell correlation, spectral signal-to-noise ratio and the Fourier neighbor correlation. The most widely accepted is the Fourier shell correlation using a threshold of 0.143 to define the resolution (so-called “gold-standard”), although it is still debated whether this is the correct threshold. Besides, the resolution obtained from the Fourier shell correlation is an estimate of varying resolution across the density map. In reality, the interpretability of the map is more important than the numerical value of the resolution.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Advanced Electron Microscopy Techniques and Applications · Enzyme Structure and Function · RNA modifications and cancer
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Faculty of Science and Engineering, University of Groningen, Nijenborgh 4, 9747 AG Groningen, The Netherlands
Groningen Biomolecular Sciences & Biotechnology Institute (GBB), University of Groningen, Nijenborgh 4, 9747 AG Groningen, The Netherlands
Moscow Institute of Physics and Technology, 141701 Dolgoprudny, Russia
References (61)
Show source-provided references
1. Crystal structure of small protein crambin at 0.48 Å resolution10.1107/S1744309110052607 ↗
2. CryoTEM with a Cold Field Emission Gun That Moves Structural Biology into a New Stage10.1017/S1431927619005725 ↗
3. The benefits of atomic resolution10.1016/S0959-440X(97)80078-4 ↗
4. Phase annealing in SHELX-90: Direct methods for larger structures10.1107/S0108767390000277 ↗
5. ‘Atomic resolution’: A badly abused term in structural biology10.1107/S205979831700225X ↗
6. XXXI Investigations in optics, with special reference to the spectroscope10.1080/14786447908639684 ↗
7. Solving structures of protein complexes by molecular replacement with Phaser10.1107/S0907444906045975 ↗
8. Pushing the boundaries of molecular replacement with maximum likelihood10.1107/S0907444901012471 ↗
9. [28] Phase determination from multiwavelength anomalous diffraction measurements10.1016/S0076-6879(97)76074-9 ↗
10. SAD phasing: History, current impact and future opportunities10.1016/j.abb.2016.03.018 ↗
11. [25] Overview of isomorphous replacement phasing10.1016/S0076-6879(97)76071-3 ↗
12. Crystallographic ab initio protein structure solution below atomic resolution10.1038/nmeth.1365 ↗
13. Sheldrick, G., Gilmore, C., Hauptman, H., Weeks, C., Miller, R., and Usón, I. (2006). Ab initio phasing. Int. Tables Crystallogr., 413–432.10.1107/97809553602060000689 ↗
14. Linking crystallographic model and data quality10.1126/science.1218231 ↗
15. How good are my data and what is the resolution?10.1107/S0907444913000061 ↗
16. Diamonds in the rough: A strong case for the inclusion of weak-intensity X-ray diffraction data10.1107/S1399004714005318 ↗
17. Better models by discarding data?10.1107/S0907444913001121 ↗
18. Weak data do not make a free lunch, only a cheap meal10.1107/S1399004713026680 ↗
19. Estimation of the quality of refined protein crystal structures10.1002/pro.2639 ↗
20. A computer-linked cathode-ray tube microdensitometer for x-ray crystallography10.1088/0022-3735/1/5/303 ↗
21. Improved R-factors for diffraction data analysis in macromolecular crystallography10.1038/nsb0497-269 ↗
22. On the use of the merging R factor as a quality indicator for X-ray data10.1107/S0021889897003907 ↗
23. Phage T7 Gp2 inhibition of Escherichia coli RNA polymerase involves misappropriation of σ70 domain 1.110.1073/pnas.1314576110 ↗
24. Structure of a prokaryotic sodium channel pore reveals essential gating elements and an outer ion binding site common to eukaryotic channels10.1016/j.jmb.2013.10.010 ↗
25. Use of massively multiple merged data for low-resolution S-SAD phasing and refinement of flavivirus NS110.1107/S1399004714017556 ↗
26. Robust structural analysis of native biological macromolecules from multi-crystal anomalous diffraction data10.1107/S0907444913001479 ↗
27. Assessing and maximizing data quality in macromolecular crystallography10.1016/j.sbi.2015.07.003 ↗
28. The PDB_REDO server for macromolecular structure model optimization10.1107/S2052252514009324 ↗
29. On effective and optical resolutions of diffraction data sets10.1107/S0907444913016673 ↗
30. SFCHECK: A unified set of procedures for evaluating the quality of macromolecular structure-factor data and their agreement with the atomic model10.1107/S0907444998006684 ↗
31. The probability distribution of X-ray intensities10.1107/S0365110X49000813 ↗
32. Global indicators of X-ray data quality10.1107/S0021889800018227 ↗
33. EFRESOL: Effective resolution of a diffraction data set10.1107/S1600576715001648 ↗
34. Maximizing the potential of electron cryomicroscopy data collected using direct detectors10.1016/j.jsb.2013.09.003 ↗
35. Detective quantum efficiency of electron area detectors in electron microscopy10.1016/j.ultramic.2009.04.002 ↗
36. Active pixel sensor array as a detector for electron microscopy10.1016/j.ultramic.2005.03.006 ↗
37. Movies of ice-embedded particles enhance resolution in electron cryo-microscopy10.1016/j.str.2012.08.026 ↗
38. A Bayesian view on cryo-EM structure determination10.1016/j.jmb.2011.11.010 ↗
39. The correlation averaging of a regularly arranged bacterial cell envelope protein10.1111/j.1365-2818.1982.tb00405.x ↗
40. Arthropod hemocyanin structures studied by image analysis
41. A new resolution criterion based on spectral signal-to-noise ratios10.1016/0304-3991(87)90225-7 ↗
42. Spectral signal-to-noise ratio and resolution assessment of 3D reconstructions10.1016/j.jsb.2004.10.011 ↗
43. Three-dimensional spectral signal-to-noise ratio for a class of reconstruction algorithms10.1016/S1047-8477(02)00033-3 ↗
44. Ab initio resolution measurement for single particle structures10.1016/j.jsb.2006.08.003 ↗
45. A review of resolution measures and related aspects in 3D Electron Microscopy10.1016/j.pbiomolbio.2016.09.005 ↗
46. Exact filters for general geometry three dimensional reconstruction
47. Optimal determination of particle orientation, absolute hand, and contrast loss in single-particle electron cryomicroscopy10.1016/j.jmb.2003.07.013 ↗
48. Prevention of overfitting in cryo-EM structure determination10.1038/nmeth.2115 ↗
49. Outcome of the first electron microscopy validation task force meeting10.1016/j.str.2011.12.014 ↗
50. Mean phase error and the map-correlation coefficient10.1107/S0907444993005852 ↗
51. Fourier shell correlation threshold criteria10.1016/j.jsb.2005.05.009 ↗
52. van Heel, M., and Schatz, M. (2017). Reassessing the revolutions resolutions. BioRxiv, 224402.10.1101/224402 ↗
53. Resolution measurement in structures derived from single particles10.1107/S0907444900009549 ↗
54. The beginning of kinesin’s force-generating cycle visualized at 9-A resolution10.1083/jcb.200612090 ↗
55. Structure of intact Thermus thermophilus V-ATPase by cryo-EM reveals organization of the membrane-bound VO motor10.1073/pnas.0911085107 ↗
56. Structure of an apoptosome-procaspase-9 CARD complex10.1016/j.str.2010.04.001 ↗
57. One number does not fit all: Mapping local variations in resolution in cryo-EM reconstructions10.1016/j.jsb.2013.08.002 ↗
58. MonoRes: Automatic and accurate estimation of local resolution for electron microscopy maps10.1016/j.str.2017.12.018 ↗
59. Quantifying the local resolution of cryo-EM density maps10.1038/nmeth.2727 ↗
60. Wavelet steerability and the higher-order Riesz transform10.1109/TIP.2009.2038832 ↗
61. Measuring local-directional resolution and local anisotropy in cryo-EM maps10.1038/s41467-019-13742-w ↗