journal article·2008-08

Fractographic failure analysis of a Procera® AllCeram crown using stereo and scanning electron microscopy

Susanne S. Scherrer, George D. Quinn, Janet B. Quinn

Dental Materials· Volume 24· Issue 8
Publisher page

Abstract

Objectives Presentation of a methodological approach using stereo and scanning electron microscope examination for the failure analysis of an alumina all-ceramic premolar crown (Procera AllCeram). Methods The recovered part of a fractured Procera alumina crown was examined utilizing first a stereomicroscope and second a scanning electron microscope (SEM). The stereomicroscope analysis was performed at low magnifications with oblique lighting in order to enhance spatial relationships and gross detection of crack features. A preliminary fracture surface map of the stereo observations was drawn and used as a guide for the SEM analysis that followed. Specific sites of interest identified under the stereo microscope were analyzed using the SEM at high magnifications searching for small fracture features such as wake hackle and twist hackle within the veneering ceramic in order to confirm the direction of crack propagation. Results At low magnifications and oblique illumination, the stereomicroscope analysis provided an excellent overview of the fractured topography, showing sites of major interest such as a primary edge chip at a margin, a compression curl indicating the end of the fracture event as well as larger hackle lines distributed over the cracked surface. The greater magnifications with the SEM analysis of the sites of interest showed the presence of wake and twist hackle, indicators of the crack propagation direction. A general map of the fracture events could be reconstructed starting with a primary veneer edge chip at the mesial margin. Hackle and wake hackle of the crack front emanating from this margin arose from hoop stresses and propagated through the full crown thickness towards the distal end of the restoration where the compression curl was located. Additional occlusal surface damage in the form of veneer chipping containing arrest lines and twist hackle running in the opposite direction as the main crack path were observed, but occurred as a secondary event without penetrating the alumina core material. Significance Stereo and scanning electron microscopy are complementary analysis techniques useful for the mapping and interpretation of the fracture surface. This case examination is intended to guide the clinical researcher in using qualitative (descriptive) fractography as a tool for understanding the failure process in brittle restorative materials, as well as for assessing possible design inadequacies.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Dental materials and restorations · Endodontics and Root Canal Treatments · Advanced ceramic materials synthesis

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (10)

Show source-provided references
1. Quinn, GD, Fractography of ceramics and glasses. A NIST recommended practice guide; Special Publication 960-16; Washington, DC: National Institute of Standards and Technology; May 2007 (http://www.ceramics.nist.gov/pubs/practice.htm).10.6028/NBS.SP.960-16 ↗
2. Standard practice for fractography and characterization of fracture origins in advanced ceramics
3. Fraktographische schadensanalyse an dentalkeramik
4. Fractographic analyses of three ceramic whole crown restoration failures10.1016/j.dental.2005.01.006 ↗
5. Failure analysis of ceramic clinical cases using qualitative fractography
6. Fracture surface analysis of clinically failed fixed partial dentures10.1177/154405910608500314 ↗
7. Fractographic ceramic failure analysis using the replica technique10.1016/j.dental.2006.12.002 ↗
8. Descriptive fractography on all ceramic dental crown failures
9. The increasing role of fractography in the dental community
10. Fracture surface analysis of dental ceramics