Corrosion resistance measurements of dental alloys, are they correlated?
Source abstract
Abstract
Objectives The aim was to assess in vitro the resistance to corrosion of eight commercial dental alloys by two quantitative methods, electrochemical and immersion tests, then to statistically test the hypothesis of possible correlation between the polarization resistance (R(p)) and the elemental release. Methods Two quantitative methods; electrochemical and immersion test, were used. From the first, after recording the OCP during 24h immersion in acidified artificial saliva (pH 2.3), R(p) was obtained using the linear polarization in anodic path and applying the Mansfeld's method. From the static immersion test, using the same test solution, the elemental release from was analysed and determined using the ICP-AES. Thereafter, the two measurements were used to plot the regression line and to determine the correlation coefficient. The significance of the correlation was tested using F-test at a confidence interval of 0.99. Results : The resistance to corrosion results obtained from the two methods were ranked and compared; an inverse relation between them was evident. Then, the obtained coefficient of correlation (R(2)) was 0.886. With the F-test at 0.99 confidence interval, the hypothesis was accepted as the calculated F was about 44 against critical F=13.7. Conclusion The correlation between the two measurements, R(p) and mass loss, was proved statistically significant. This result may provide a new approach to predict the corrosion behaviour of dental alloys by firstly using the easier methods.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Dental Erosion and Treatment · Dental materials and restorations · Endodontics and Root Canal Treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (22)
Show source-provided references
1. Bockris · 1981 · vol. 4
2. In vitro corrosion behaviour and metallic ion release of different prosthodobtic alloys10.1111/j.1875-595X.1999.tb00538.x ↗
3. Corrosion behaviour of dental metals and alloys in different media10.1046/j.1365-2842.1998.00305.x ↗
4. In vitro corrosion behavior of 13 prosthodontic alloys
5. The effect of pH on the corrosion of dental metal alloys10.1046/j.1365-2842.2000.00549.x ↗
6. Biocompatibility of dental casting alloys: a review10.1016/S0022-3913(00)80016-5 ↗
7. ISO 10271:2001. Dental metallic materials—corrosion test methods.
8. ASTM G 16-95 (reapproved 2004). Standard guide for applying statistics to analysis of corrosion data.
9. Dunn · 1974
10. Biological interactions of dental cast alloys with oral tissues10.1016/S0109-5641(01)00063-X ↗
11. Electrochemical characteristics of high-Pd alloys in relation to Pd-allergy10.1016/S0109-5641(00)00016-6 ↗
12. Craig · 1975
13. Corrosion and biocompatibility testing of palladium alloy castings10.1016/S0109-5641(00)00033-6 ↗
14. Effect of chemical composition on the corrosion behavior of Ni–Cr–Mo dental casting alloys10.1002/jbm.10080 ↗
15. Potentiodynamic polarization study of the in vitro corrosion behavior of 3 high-palladium alloys and a gold-palladium alloy in 5 media10.1067/mpr.2002.121239 ↗
16. Dealloying and electoformation in high-Pd dental alloys10.1016/S0109-5641(00)00045-2 ↗
17. In vitro corrosion resistance of high-palladium dental casting alloys10.1016/S0109-5641(99)00037-8 ↗
18. Static immersion and irritation tests of dental metal–ceramic alloys10.1111/j.1600-0722.2005.00187.x ↗
19. Comparison of metal release from various metallic biomaterials in vitro10.1016/j.biomaterials.2004.02.005 ↗
20. Corrosion resistance of cobalt–chromium and palladium–silver alloys used in fixed prosthetic restorations10.1111/j.1600-0722.2005.00190.x ↗
21. Electrochemical studies of the corrosion behaviour of titanium and the Ti–6Al–4V alloy using electrochemical impedance spectroscopy
22. A model for alloy film corrosion and prediction of resulting passive layer structures