Screening dental implant design parameters for effect on the fatigue limit of reduced-diameter implants
Source abstract
Abstract
Objectives This study screened the design parameters of a reduced-diameter implant to determine which parameters have the most significant effect on the implant fatigue limit. Methods A dental implant assembly, which included an implant body (Biomet 3i), an abutment (GingiHue®), and an abutment screw (Gold-Tite Square screw) was scanned using micro-computed tomography (SkyScan 1172) and was measured using Mimics (Materialise) and an optical microscope (VHX-1000, Keyence). Sixteen design parameters were measured, and the values of the commercial design were taken as reference level for each design parameter. Values up to 20 % lower and 20 % higher than the reference were explored using a Taguchi orthogonal array (DOE++, Reliasoft), which varies more than one design parameter at a time to efficiently explore all main effects and lower order interactions across few implant designs. Solid models of these 27 implant designs and the reference design were constructed using SOLIDWORKS (Dassault Systèmes). Each solid model was loaded according to ISO 14801. fe-safe (Dassault Systèmes) was used to estimate the fatigue limits. ANOVA statistical test in DOE++ was used to screen the design parameters. Results Interaction between the coronal and apical tapers of the implant body had a significant effect on the fatigue limit (p ≤ 0.05), where fatigue limit was low for designs with a constant taper. Conversely, the combination of high degree of apical taper and low degree of coronal taper lead to the highest fatigue limit. Significance Using a Taguchi orthogonal array proved to be an efficient strategy for screening implant design parameters for effect on fatigue limit. The modified implant designed by manipulating the most influential parameters is predicted to have much greater fatigue limit compared to the commercially available design.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Dental Implant Techniques and Outcomes · Dental materials and restorations · Orthodontics and Dentofacial Orthopedics
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (52)
Show source-provided references
1. Bone "Mass" and the "Mechanostat": a proposal10.1002/ar.1092190104 ↗
2. Single-tooth implants and their role in preserving remaining teeth: a 10-year survival study
3. A prospective clinical study evaluating the safety and effectiveness of narrow-diameter threaded implants in the anterior region of the maxilla
4. Early dental implant failure: risk factors10.1016/j.bjoms.2011.04.074 ↗
5. Dental Implants10.1016/j.cden.2017.06.007 ↗
6. Influence of diameter and length of implant on early dental implant failure10.1016/j.joms.2009.10.002 ↗
7. The influence of implant diameter on its survival: a meta-analysis based on prospective clinical trials10.1902/jop.2013.130043 ↗
8. Long-term retrospective study of narrow implants for fixed dental prostheses10.1111/j.1600-0501.2012.02472.x ↗
9. The clinical performance of narrow diameter implants versus regular diameter implants: a meta-analysis10.1563/aaid-joi-D-19-00025 ↗
10. Narrow-diameter implants: a systematic review and meta-analysis10.1111/clr.13272 ↗
11. A comparative analysis of implants presenting different diameters: extra-narrow, narrow and conventional10.3390/ma13081888 ↗
12. A 1-year clinical report of a one-piece implant abutment
13. Prosthodontic complications in a prospective clinical trial of single-stage implants at 36 months
14. Clinical complications with implants and implant prostheses10.1016/S0022-3913(03)00212-9 ↗
15. Failures and complications in 391 consecutively inserted fixed prostheses supported by Branemark implants in edentulous jaws: a study of treatment from the time of prosthesis placement to the first annual checkup
16. Loose gold screws frequently occur in full-arch fixed prostheses supported by osseointegrated implants after 5 years
17. A study of 589 consecutive implants supporting complete fixed prostheses. Part II: prosthetic aspects10.1016/0022-3913(92)90557-Q ↗
18. Cement- and screw-retained implant-supported prostheses: up to 10 years of follow-up of a new design
19. Single tooth replacement of missing molars: a retrospective study of 78 implants
20. Tissue-integrated prosthesis complications
21. Single implant-supported molar and premolar crowns: a ten-year retrospective clinical report10.1016/j.prosdent.2003.08.025 ↗
22. Effect of dental implant diameter on fatigue performance. Part I: mechanical behavior10.1111/j.1708-8208.2012.00477.x ↗
23. Effect of dental implant diameter on fatigue performance. Part II: failure analysis10.1111/j.1708-8208.2012.00476.x ↗
24. Effect of thread depth and implant shape on stress distribution in anterior and posterior regions of mandible bone: a finite element analysis10.4103/1735-3327.255745 ↗
25. Evaluation of the mechanical characteristics of the implant-abutment complex of a reduced-diameter morse-taper implant - a nonlinear finite element stress analysis10.1034/j.1600-0501.2003.00828.x ↗
26. The impact of loads on standard diameter, small diameter and mini implants: a comparative laboratory study10.1111/j.1600-0501.2007.01395.x ↗
27. The role of cortical zone level and prosthetic platform angle in dental implant mechanical response: a 3D finite element analysis10.1016/j.dental.2021.08.022 ↗
28. Freitas Junior, A.C., E.A. Bonfante, N.R. Silva, L. Marotta, and P.G. Coelho, Effect of implant-abutment connection design on reliability of crowns: regular vs. horizontal mismatched platform. Clin Oral Implants Res. 23(9): p. 1123-1126.10.1111/j.1600-0501.2011.02257.x ↗
29. Freitas-Junior, A.C., E.A. Bonfante, L.M. Martins, N.R. Silva, L. Marotta, and P.G. Coelho, Effect of implant diameter on reliability and failure modes of molar crowns. International Journal of Prosthodontics. 24(6): p. 557-561.
30. Load fatigue performance of a single-tooth implant abutment system: effect of diameter
31. Shemtov-Yona, K., D. Rittel, L. Levin, and E.E. Machtei, Effect of Dental Implant Diameter on Fatigue Performance. Part I: Mechanical Behavior. Clin Implant Dent Relat Res.
32. Shemtov-Yona, K., D. Rittel, E.E. Machtei, and L. Levin, Effect of Dental Implant Diameter on Fatigue Performance. Part II: Failure Analysis. Clin Implant Dent Relat Res.
33. Effect of crown stiffness and prosthetic screw absence on the stress distribution in implant-supported restoration: a 3D finite element analysis10.1371/journal.pone.0285421 ↗
34. The effect of abutment material stiffness on the mechanical behavior of dental implant assemblies: a 3D finite element study10.1016/j.jmbbm.2023.105847 ↗
35. Effects of tightening torque on screw stress and formation of implant-abutment microgaps: a finite element analysis10.1016/j.prosdent.2020.11.026 ↗
36. Taguchi, G., Introduction to quality engineering: designing quality into products and processes. 1986.
37. The QC circle phenomenon
38. Loeb, M.A., J.A. Griggs, and Y. Duan, Design optimization of reduced-diameter dental implants, in 43rd General Meeting and Exhibition of the AADR. 2014: Charlotte, NC.
39. Effect of loading frequency on cyclic fatigue lifetime of a standard-diameter implant with an internal abutment connection10.1016/j.dental.2018.09.001 ↗
40. Effect of Bone Remodeling on dental implant fatigue limit predicted using 3D finite element analysis
41. Effects of abutment screw preload and preload simulation techniques on dental implant lifetime
42. ToolBox, E. Young's Modulus, Tensile Strength and Yield Strength Values for some Materials. 2003; Available from: http://www.engineeringtoolbox.com/young-modulus-d_417.html.
43. Mechanical and microstructural characterization of AS-CAST Ti-12Mo-xNb alloys for orthopedic application10.1590/1980-5373-mr-2019-0178 ↗
44. New Ti-alloys and surface modifications to improve the mechanical properties and the biological response to orthopedic and dental implants: a review10.1155/2016/2908570 ↗
45. Mechanical properties of vertebrate hard tissues10.1243/0954411981534178 ↗
46. Ultrasonic determination of the elastic modulus of human cortical bone10.1007/BF02522857 ↗
47. Fatigue lifetime prediction of a reduced-diameter dental implant system: numerical and experimental study10.1016/j.dental.2018.06.002 ↗
48. Corporation · 2008
49. Corporation, C.T., Technical Data Sheet: CP Titanium Grade 4. 2000: www.cartech.com. 1-10.
50. Fatigue lifetime of reduced-diameter implants placed in different bone models10.1016/j.dental.2024.09.003 ↗
51. Effect of implant diameter on fatigue strength10.1097/ID.0000000000000502 ↗
52. Comparison of the corrosion fatigue characteristics of CP Ti-grade 4, Ti-6Al-4V ELI, Ti-6Al-7Nb, and Ti-15Mo