Assessment of mechanical properties of WaveOne Gold Primary reciprocating instruments
Source abstract
Abstract
This study aimed to evaluate cyclic fatigue resistance and bending properties and torque/force generation of WaveOne Gold (WOG) Primary in comparison with WaveOne (WO) Primary and Reciproc (RE) R25. A cyclic fatigue test revealed that the WOG Primary took significantly longer time to fracture compared with the WO Primary (p<0.05). The WOG Primary had the smallest load values at a deflection of 0.5 and 2 mm (p<0.05), as measured with a cantilever bending test. Torque/force measurement demonstrated that maximum upward force and maximum counterclockwise torque values in the WOG Primary were significantly lower than those in the RE R25 (p<0.05). Under the present experimental condition, the WOG Primary showed a higher cyclic fatigue resistance compared with the WO Primary, a higher flexibility compared with the WO Primary and RE R25, and generated a significantly lower maximum torque compared with the RE R25.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Endodontics and Root Canal Treatments · Drilling and Well Engineering · Ultrasonics and Acoustic Wave Propagation
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Endodontic Department, Stomatological Hospital of Southern Medical University
Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Science, Tokyo Medical and Dental University (TMDU)
References (24)
Show source-provided references
1. 1) Bürklein S, Hinschitza K, Dammaschke T, Schäfer E. Shaping ability and cleaning effectiveness of two single-filesystems in severely curved root canals of extracted teeth: Reciproc and WaveOne versus Mtwo and ProTaper. Int Endod J 2012; 45: 449-461.10.1111/j.1365-2591.2011.01996.x ↗
2. 2) Capar ID, Ertas H, Ok E, Arslan H, Ertas ET. Comparative study of different novel nickel-titaniumrotary systems for root canal preparation in severely curved root canals. J Endod 2014; 40: 852-856.10.1016/j.joen.2013.10.010 ↗
3. 4) De-Deus G, Moreira EJ, Lopes HP, Elias CN. Extended cyclic fatigue life of F2 ProTaper instruments used in reciprocating movement. Int Endod J 2010; 43: 1063-1068.10.1111/j.1365-2591.2010.01756.x ↗
4. 5) Capar ID, Arslan H. A review of instrumentation kinematics of engine-driven nickel-titanium instruments. Int Endod J 2016; 49: 119-135.10.1111/iej.12432 ↗
5. 6) Elnaghy AM, Elsaka SE. Assessment of the mechanical properties of ProTaper Next Nickel-titanium rotary files. J Endod 2014; 40: 1830-1834.10.1016/j.joen.2014.06.011 ↗
6. 7) Keskin C, Inan U, Demiral M. Effect of interrupted motion on the cyclic fatigue resistance of reciprocating nickel-titanium instruments. Int Endod J 2018; 51: 549-555.10.1111/iej.12790 ↗
7. 8) Elsaka SE, Elnaghy AM, Badr AE. Torsional and bending resistance of WaveOne Gold, Reciproc and Twisted File Adaptive instruments. Int Endod J 2017; 50: 1077-1083.10.1111/iej.12728 ↗
8. 9) Hieawy A, Haapasalo M, Zhou H, Wang ZJ, Shen Y. Phase transformation behavior and resistance to bending and cyclic fatigue of ProTaper Gold and ProTaper Universal instruments. J Endod 2015; 41: 1134-1138.10.1016/j.joen.2015.02.030 ↗
9. 11) Piao J, Miyara K, Ebihara A, Nomura N, Hanawa T, Suda H. Correlation between cyclic fatigue and the bending properties of nickel titanium endodontic instruments. Dent Mater J 2014; 33: 539-544.10.4012/dmj.2013-318 ↗
10. 12) Tokita D, Ebihara A, Nishijo M, Miyara K, Okiji T. Dynamic torque and vertical force analysis during nickel-titanium rotary root canal preparation with different modes of reciprocal rotation. J Endod 2017; 43: 1706-1710.10.1016/j.joen.2017.05.010 ↗
11. 13) Plotino G, Grande NM, Testarelli L, Gambarini G. Cyclic fatigue of Reciproc and WaveOne reciprocating instruments. Int Endod J 2012; 45: 614-618.10.1111/j.1365-2591.2012.02015.x ↗
12. 14) Özyürek T. Cyclic fatigue resistance of Reciproc, WaveOne, and WaveOne Gold nickel-titanium instruments. J Endod 2016; 42: 1536-1539.10.1016/j.joen.2016.06.019 ↗
13. 15) Arias A, Perez-Higueras JJ, de la Macorra JC. Influence of clinical usage of GT and GTX files on cyclic fatigue resistance. Int Endod J 2014; 47: 257-263.10.1111/iej.12141 ↗
14. 16) He R, Ni J. Design improvement and failure reduction of endodontic files through finite element analysis: application to V-Taper file designs. J Endod 2010; 36: 1552-1557.10.1016/j.joen.2010.06.002 ↗
15. 17) Capar ID, Kaval ME, Ertas H, Sen BH. Comparison of the cyclic fatigue resistance of 5 different rotary pathfinding instruments made of conventional nickel-titanium wire, M-wire, and controlled memory wire. J Endod 2015; 41: 535-538.10.1016/j.joen.2014.11.008 ↗
16. 18) Cheung GS, Darvell BW. Low-cycle fatigue of NiTi rotary instruments of various cross-sectional shapes. Int Endod J 2007; 40: 626-632.10.1111/j.1365-2591.2007.01257.x ↗
17. 19) Grande NM, Plotino G, Pecci R, Bedini R, Malagnino VA, Somma F. Cyclic fatigue resistance and three-dimensional analysis of instruments from two nickel-titanium rotary systems. Int Endod J 2006; 39: 755-763.10.1111/j.1365-2591.2006.01143.x ↗
18. 20) Tripi TR, Bonaccorso A, Condorelli GG. Cyclic fatigue of different nickel-titanium endodontic rotary instruments. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2006; 102: e106-e114.10.1016/j.tripleo.2005.12.012 ↗
19. 21) Yilmaz K, Özyürek T. Cyclic fatigue life of Tango-Endo, WaveOne GOLD, and Reciproc NiTi instruments. Restor Dent Endod 2017; 42: 134-139.10.5395/rde.2017.42.2.134 ↗
20. 22) Arias A, Perez-Higueras JJ, de la Macorra JC. Differences in cyclic fatigue resistance at apical and coronal levels of Reciproc and WaveOne new files. J Endod 2012; 38: 1244-1248.10.1016/j.joen.2012.05.022 ↗
21. 23) Kim HC, Kwak SW, Cheung GS, Ko DH, Chung SM, Lee W. Cyclic fatigue and torsional resistance of two new nickel-titanium instruments used in reciprocation motion: Reciproc versus WaveOne. J Endod 2012; 38: 541-544.10.1016/j.joen.2011.11.014 ↗
22. 24) Lopes HP, Gambarra-Soares T, Elias CN, Siqueira JF Jr, Inojosa IF, Lopes WS, Vieira VT. Comparison of the mechanical properties of rotary instruments made of conventional nickel-titanium wire, M-wire, or nickel-titanium alloy in R-phase. J Endod 2013; 39: 516-520.10.1016/j.joen.2012.12.006 ↗
23. 25) Ha JH, Kwak SW, Kim SK, Kim HC. Screw-in forces during instrumentation by various file systems. Restor Dent Endod 2016; 41: 304-309.10.5395/rde.2016.41.4.304 ↗
24. 26) Ha JH, Cheung GS, Versluis A, Lee CJ, Kwak SW, Kim HC. ‘Screw-in’ tendency of rotary nickel-titanium files due to design geometry. Int Endod J 2015; 48: 666-672.10.1111/iej.12363 ↗