Total hip stem failure due to weld procedure and material selection misuse
Source abstract
Abstract
Abstract Total hip arthroplasty has been one of the most successful techniques in orthopedics providing good results for long term. The rate of fatigue failure of total hip arthroplasty stem today is very low due to improvements of the technique along many decades. However, any rare case of stem fracture should be treated carefully due to the problems that can result in the revision for the patient. This paper presents failure analysis of three cemented hip stem which experienced fatigue with less than 5 years in use. The failure is associated to design concept which includes a weld procedure at the proximal third of the stem to attach a metallic wire mesh aiming to improve the bonding to the cement mantle. The failure also is associated with the misuse of AISI 316 L as implant material.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Advanced materials and composites
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (14)
Show source-provided references
1. Fatigue failure analysis of a specific total hip prosthesis stem design10.1016/j.ijfatigue.2007.11.005 ↗
2. Failure analysis of uncemented total hip stem due to microstructure and neck stress riser10.1016/j.engfailanal.2007.10.012 ↗
3. Design aspects involved in a cemented THA stem failure case10.1016/j.engfailanal.2008.06.016 ↗
4. Migration, stem shape, and surface finish in cemented total hip arthroplasty10.1097/00003086-199810000-00011 ↗
5. Charnley low-friction arthroplasty: a worldwide retrospective review at 1 to 20years10.1054/arth.2002.31973 ↗
6. Long-term results of Charnley total hip replacement: review of 92 patients at 15 to 20years10.1302/0301-620X.76B2.8113285 ↗
7. Twenty-five-year survivorship of two thousand consecutive primary Charnley total hip replacements: factors affecting survivorship of acetabular and femoral components10.2106/00004623-200202000-00002 ↗
8. Comparison of clinical outcomes in total hip arthroplasty using rough and polished cemented stems with essentially the same geometry10.2106/00004623-200204000-00012 ↗
9. Twenty – to twenty-six year radiographic review in patients 50 years of age or younger with cemented Charnley low-friction arthroplasty10.1016/S0883-5403(03)00296-1 ↗
10. Primary Exeter total hip arthroplasty in patients with small femurs, a minimal of 10years follow-up10.1016/j.arth.2004.01.014 ↗
11. Influence of stem geometry on the stability of polished tapered cemented femoral stems10.1302/0301-620X.87B7.16079 ↗
12. The Exeter universal cemented femoral component at 8 to 12 years – a study of the first 325 hips10.1302/0301-620X.84B3.12261 ↗
13. Failure of acetabular cups fixed with cement and thick embedded wire mesh
14. Influence of the ferrite rate on the tenacity of a welded joint in austenitic stainless steel: experimental study and numerical modeling10.1016/j.commatsci.2008.10.009 ↗