journal article·2013-09

Metallurgical failure analysis of acetabular metal-backed screws

Sandro Griza, Thiago Figueiredo Azevedo, Silvando V. dos Santos, Eduardo K. Tentardini, Telmo R. Strohaecker

Publisher page

Abstract

Abstract Screws are often used in total hip arthroplasty to increase the cementless acetabular metal-backed fixation stability. This study aimed at identifying metallurgical factors related to metal-backed screw fractures in two cases of premature total hip arthroplasty revision. A preliminary analysis of radiographs and retrieved parts was performed. Fractured screws were subjected to fracture analysis, energy dispersive X-ray spectroscopy (EDS) and metallographic analysis. A finite element analysis was performed to predict the stresses experienced by the screws. Two factors were identified: titanium alloy microstructure, consisting of alpha phase plates with oxygen-rich alpha case and as cast Cr–Co alloy with dendritic microstructure and unfavorably aligned shrinkage formation. Finite element analysis has showed enough level of stress to predict screw fracture in the case of unbounded metal-backed and bone interface.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Hip disorders and treatments · Orthopedic Infections and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (34)

Show source-provided references
1. The outcome of Charnley total hip arthroplasty with cement after a minimum twenty-year follow-up. The results of one surgeon10.2106/00004623-199307000-00002 ↗
2. Acetabular fixation options: notes from the other side10.1016/j.arth.2006.02.152 ↗
3. The effect of interfacial parameters on cup bone relative micromotions, a finite element investigation10.1016/S0021-9290(00)00112-3 ↗
4. Analysis of the results of the C-Fit uncemented total hip arthroplasty in young patients with hydroxyapatite or porous coating of components
5. Berry DJ, Barnes CL, Scou RD, Cabanela ME, Poss R. Catastrophic failure of the polyethylene liner of uncemented acetabular components. Bone Joint Surg [Br] 1994;76-B:575–8.10.1302/0301-620X.76B4.8027143 ↗
6. Failure of an uncemented acetabular prosthesis – a case study10.1016/j.engfailanal.2004.10.010 ↗
7. Barrack RL, Burke DW, Cook SD, Skinner HB, Harris WH. Complications related to modularity of total hip components. J Bone Joint Surg [Br] 1993;75-B:688–92.10.1302/0301-620X.75B5.8376421 ↗
8. Stiehl JB, Mahfouz MR. Catastrophic failure of a modular revision total hip polyethylene insert. J Arthroplasty 2007;22(1).10.1016/j.arth.2006.02.077 ↗
9. Ito H, Minami A, Kondo E, Fujita M, Ubayama Y, Matsuno T. Destruction of acetabular bone caused by early failure of a constrained acetabular component. J Arthroplasty 2001;16(7).10.1054/arth.2001.26595 ↗
10. Dissociation of acetabular components after total hip arthroplasty – report of four cases10.2106/00004623-199072100-00019 ↗
11. Yun AG, Padgett D, Pellicci P, Dorr LD. Constrained acetabular liners mechanisms of failure. J Arthroplasty 2005;20(4).10.1016/j.arth.2004.12.054 ↗
12. Cheung K, Yung S, Wong K, Chiu K. Early failure of smooth hydroxyapatite-coated press-fit acetabular cup – 7 years of follow-up. J Arthroplasty 2005;20(5).10.1016/j.arth.2004.10.002 ↗
13. Ronald R. Hugate, Ian D. Dickey, Qingshan Chen, Christina M. Wood, Franklin H. Sim, Michael G. Rock. Fixed-angle screws vs standard screws in acetabular prosthesis fixation – a cadaveric biomechanical study. J Arthroplasty 2009;24(5).10.1016/j.arth.2008.04.019 ↗
14. Failure analysis of uncemented total hip stem due to microstructure and neck stress riser10.1016/j.engfailanal.2007.10.012 ↗
15. Acetabular metal-backed fatigue due to severe wear before revision10.1016/j.engfailanal.2009.01.004 ↗
16. Case study and analysis of a fatigue failure in a THA stem10.1016/j.engfailanal.2012.10.011 ↗
17. Low-friction arthroplasty of the hip. A 10–12-year follow-up study
18. Charnley low-friction arthroplasty – a worldwide retrospective review at 15–20 years10.1054/arth.2002.31973 ↗
19. Experience with the exeter total hip replacement since 1970
20. Neumann L, Freund KG, Sorenson KH. Long-term results of Charnley total hip replacement: review of 92 patients at 15 to 20 years. J Bone Joint Surg [Br] 1994;76-B:245–51.10.1302/0301-620X.76B2.8113285 ↗
21. Berry DJ, Harmsen WS, Cabanela ME, Morrey BF. Twenty-five-year survivorship of two thousand consecutive primary Charnley total hip replacements: factors affecting survivorship of acetabular and femoral components. J Bone Joint Surg Am 2002; 84-A(2):171–7 [Feb].10.2106/00004623-200202000-00002 ↗
22. Busch V, Klarenbeek R, Slooff T, Schreurs BW, Gardeniers J. Cemented hip designs are a reasonable option in young patients. Clin Orthop Relat Res 2010; 468(12):3214–20 [Apr. 20].10.1007/s11999-010-1355-z ↗
23. Primary exeter total hip arthroplasty in patients with small femurs, a minimal of 10years follow-up10.1016/j.arth.2004.01.014 ↗
24. Collis DK, Mohler CG. Comparison of clinical outcomes in total hip arthroplasty using rough and polished cemented stems with essentially the same geometry. J Bone Joint Surg 2002;84-a:586–92 [4].10.2106/00004623-200204000-00012 ↗
25. Charnley low-frictional torque arthroplasty in patients under the age of 51 years: follow-up to 33 years10.1302/0301-620X.84B4.10293 ↗
26. Fatigue of biomaterials: a review10.1016/S0142-1123(00)00052-9 ↗
27. Failure analysis of a Ti6Al4V cementless HIP prosthesis10.1016/j.engfailanal.2006.11.003 ↗
28. Fatigue characteristics of metallic biomaterials10.1016/j.ijfatigue.2006.09.021 ↗
29. Jensen Walter J. Failures of mechanical fasteners. ASM metals handbook. Failure analysis and prevention, vol. 11. 10th ed.; 1991. p. 2443.
30. Boyer Rodney R. Titanium and titanium alloys: metallographic techniques and microstructures. ASM metals handbook. Metallography and microstructures, vol. 9, 10th ed.; 1991. p. 1033.
31. The fatigue crack propagation resistance of Ti–6Al–4V under aqueous saline environments10.1016/S0142-1123(01)00141-4 ↗
32. Classification of environmentally assisted fatigue crack growth behavior10.1016/j.ijfatigue.2009.03.019 ↗
33. Case study of Ti6Al4V pedicle screw failures due to geometric and microstructural aspects10.1016/j.engfailanal.2012.05.009 ↗
34. Influence of femoral head size on impingement, dislocation and stress distribution in total hip replacement10.1016/j.medengphy.2006.07.001 ↗