A preliminary specimen-level mechanical and compositional characterization of two commercial zirconia dental implant materials
Source abstract
Abstract
This preliminary specimen-level study characterized single specimens from two commercially available zirconia dental implant materials using nanoindentation, SEM-EDX, and XRD. The aim was to document local mechanical response, elemental composition, and qualitative phase-related features, and to provide preliminary implant-specific material information that may be considered in future simulation-based studies. One implant specimen from each commercial system was evaluated. Repeated nanoindentation measurements were obtained from different locations on metallographically prepared sections, while surface morphology and elemental composition were examined by SEM-EDX using a ZEISS EVO10 scanning electron microscope equipped with a Bruker XFlash Detector 730M-250 EDX system. Phase-related features were assessed qualitatively by XRD in the 2θ range of 20°–70°. Repeated local nanoindentation measurements yielded descriptive hardness and indentation elastic modulus values for the two specimens. Z1 showed a hardness of 17.23 ± 0.87 GPa and an indentation elastic modulus of 223.7 ± 7.6 GPa based on 10 measurements, whereas Z2 showed a hardness of 17.96 ± 1.01 GPa and an indentation elastic modulus of 246.2 ± 7.4 GPa based on eight measurements; in both cases, the SD reflects local spatial variability within a single specimen rather than between-specimen variability. SEM-EDX analyses showed Zr–O–Y-based elemental profiles in both specimens, while a trace Al signal was observed only in Z1 by EDX spectrum and elemental mapping. The XRD patterns of both specimens showed broadly similar zirconia-based reflections within the limits of qualitative comparison. These findings should be interpreted as preliminary specimen-level observations rather than statistically generalizable commercial-system data. The measured values may be useful as preliminary material inputs for future sensitivity-based simulation studies, provided that they are considered together with validated literature data and appropriate model-specific assumptions.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Dental Implant Techniques and Outcomes · Dental materials and restorations · Bone Tissue Engineering Materials
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Oral and Maxillofacial Surgery, Institute of Health Sciences, Bezmialem Vakif University
Department of Metallurgical and Materials Engineering, Marmara University
Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Bezmialem Vakif University
References (20)
Show source-provided references
1. Influence of the microstructure on the life prediction of hydrothermal degraded 3Y-TZP bioceramics10.1016/j.jmrt.2020.07.059 ↗
2. Are ceramic implants a viable alternative to titanium implants? A systematic literature review10.1111/j.1600-0501.2009.01785.x ↗
3. From mechanical models to clinical reality: a systematic review of finite element advances in dental implant design, biomechanics, and optimization10.1016/j.mtcomm.2025.114314 ↗
4. Classification and properties of dental zirconia as implant fixtures and superstructures10.3390/ma14174879 ↗
5. Single-tooth implant rehabilitation: a retrospective study of 67 implants
6. Alumina-doped zirconia submicro-particles: synthesis, thermal stability, and microstructural characterization10.3390/ma12182856 ↗
7. Optimizing SEM-EDX for fast, high-quality and non-destructive elemental analysis of glass10.1039/D4JA00212A ↗
8. The influence of alumina airborne-particle abrasion on the properties of zirconia-based dental ceramics (3Y-TZP)10.3390/coatings13101691 ↗
9. Synthesis and thermal stability of zirconia and yttria-stabilized zirconia microspheres10.1016/j.jcis.2015.02.049 ↗
10. Titanium alloys in total joint replacement—a materials science perspective10.1016/s0142-9612(97)00146-4 ↗
11. Prospective controlled clinical study investigating long-term clinical parameters, patient satisfaction, and microbial contamination of zirconia implants10.1111/cid.12720 ↗
12. Factors affecting the decision to replace failed implants: a retrospective study10.1902/jop.2008.080255 ↗
13. Cubic-formation and grain-growth mechanisms in tetragonal zirconia polycrystal10.1111/j.1151-2916.2003.tb03483.x ↗
14. Sintering kinetics at constant rates of heating: effect of Al2O3 on the initial sintering stage of fine zirconia powder10.1111/j.1551-2916.2005.00620.x ↗
15. Ceramic implant rehabilitation: consensus statements from joint congress for ceramic implantology: consensus statements on ceramic implant10.1563/aaid-joi-D-23-00083 ↗
16. Zirconia dental implants: a closer look at surface condition and intrinsic composition by SEM-EDX10.3390/bioengineering10091102 ↗
17. Choosing between energy-dispersive or wavelength-dispersive SEM spectroscopy methods for elemental distribution in deacidified paper10.1007/s10570-025-06771-w ↗
18. Bacterial adherence and biofilm formation on medical implants: a review10.1177/0954411914556137 ↗
19. Is titanium alloy Ti-6Al-4V cytotoxic to gingival fibroblasts—A systematic review10.1002/cre2.444 ↗
20. Fracture analysis of one/two-piece clinically failed zirconia dental implants10.1016/j.dental.2022.08.004 ↗