A new method for concentration analysis of bacterial endotoxins in perfluorocarbon
Source abstract
Abstract
This communication demonstrates the feasibility of the gel-clot method for the analysis of bacterial endotoxins in water extracts of perfluorocarbon which is a water insoluble liquid medical device. Perfluorocarbon (10 mL) was shaken with 10mL water for 15 min at 2000 r/min and the endotoxin present was extracted to the aqueous phase without interference inhibition/enhancement of the product and the recovery of endotoxin added to perfluorocarbon was determined. A validation study confirmed that endotoxins presented in perfluorocarbon pass over into the aqueous phase at concentrations of 20, 10 and 5 EU/mL with recoveries from 86.8% to 96.8%. Therefore, the gel-clot test is suitable for detecting bacterial endotoxins in perfluorocarbon which is a water insoluble medical device.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Analytical Chemistry and Sensors · Odor and Emission Control Technologies · Nitric Oxide and Endothelin Effects
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (9)
Show source-provided references
1. ISO 16672-2003. Ophthalmic implants — ocular endotamponades
2. United States Pharmacopeia. Chapter No. 85: Analysis of bacterial endotoxin
3. United States Pharmacopeia. Chapter No. 161: Blood and other liquid transfusion device and similar medical device
4. Chinese Pharmacopoeia. Appendix XIE, 2010, 99–102 (in Chinese)
5. Zhang Y, Nu S-Y, Gao W-J, et al. New method for bacterial endotoxin test of absorptive medical film. Modern Preventive Medicine, 2007, 34(18): 3442–3446 (in Chinese)
6. Zheng X, Zhao Y M, Liu C. Bacterial endotoxin test for puerarin injection. China Pharmacy, 2007, 18: 1254–1255 (in Chinese)
7. Pan Z-X, Song Q. Study on determination of bacterial endotoxin of omeprazole sodium for injection. Chinese Journal of Pharmaceutical Analysis, 2008, 28(2): 275–277 (in Chinese)
8. Gao J, Fu C, Ji Y, et al. Study on bacterial endotoxin test for neostigmine methylsulfate injection. Chinese Journal of Pharmaceutical Analysis, 2008, 28(7): 1141–1143 (in Chinese)
9. Hu Y-C, Li N, Zhou J-P. Study on gelation clot method of bacterial endotoxin test in testorerone undecanoate injection. Chinese Journal of Pharmaceutical Analysis, 2010, 30(12): 2425–2429 (in Chinese)