journal article·2024-01

Analysis of Cerium-Substituted Yttrium Iron Garnets Using Inductively Coupled Plasma Atomic Emission Spectrometry with Preliminary Decomposition in a Microwave System

N. A. Korotkova, K. V. Petrova, V. B. Baranovskaya

Inorganic Materials· Volume 60· Issue 1
Publisher page

Abstract

Abstract A method for analyzing iron garnets of the composition Y3–xCexFe5–yGayO12, where x = 0.4–0.5 and y = 2.4–2.6, using inductively coupled plasma atomic emission spectroscopy (ICP-AES) with preliminary decomposition of the sample in microwave system is described. During the investigation, conditions for microwave decomposition of samples are studied and selected—the composition of acid mixtures (HCl/HNO3) and modes of microwave heating (holding time and temperature) providing complete dissolution of the samples. Conditions for ICP-AES analysis of cerium-doped yttrium iron garnets and the effect of matrix components (Y, Ce, Fe, and Ga) on the determination of impurity elements are studied. The operating parameters of the spectrometer (high-frequency generator power and argon spray flow rate) making it possible to reduce the matrix effect when determining Na, Mg, Al, Si, P, K, Ca, Sc, Cr, Mn, Co, Ni, Cu, Zn, Se, Cd, Sn, Te, La, Pr, Nd, Sm, Eu, Gd, Tb, Dy, Ho, Er, Tm, Yb, Lu, and Pb impurities in yttrium iron garnets using ICP-AES are selected in order to minimize the matrix effect. ICP-AES studies are carried out in axial and radial modes for impurity elements and in radial modes for matrix elements. The limits of quantification (LOQs) of most analytes are in the range from n × 10–5 to n × 10–4 wt %. The accuracy is confirmed by the “spike–recovery” method, as well as by the analysis of samples with certified element contents. The standard deviation is in the range from 1 to 5% depending on the selected plasma viewing mode and the elements being detected. The developed method makes it possible to determine macro- and microcomponents in yttrium iron garnets in a wide range of concentrations with a high accuracy.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Magneto-Optical Properties and Applications · Mineralogy and Gemology Studies · Photorefractive and Nonlinear Optics

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (18)

Show source-provided references
1. Smirnova, M.N., Glazkova, I.S., Nikiforova, G.E., et al., Synthesis of Ce: YIG nanopowder by gel combustion, Nanosyst. Phys. Chem. Math., 2021, vol. 12, no. 2, pp. 210–217. https://doi.org/10.17586/2220-8054-2021-12-2-210-21710.17586/2220-8054-2021-12-2-210-217 ↗
2. Teterin, Yu.A., Smirnova, M.N., Maslakov, K.I., et al., Ionic and phase compositions of Y2.5Ce0.5Fe2.5Ga2.5O12 ferrogarnet powder produced by gel combustion, Dokl. Phys. Chem., 2022, vol. 503, pp. 45–49. https://doi.org/10.1134/s001250162204002910.1134/s0012501622040029 ↗
3. Balaram, V., Rare earth elements: A review of applications, occurrence, exploration, analysis, recycling, and environmental impact, Geosci. Front., 2019, vol. 10, pp. 1286–1303. https://doi.org/10.1016/j.gsf.2018.12.00510.1016/j.gsf.2018.12.005 ↗
4. Xu, Y., Li, M., Wang, C., et al., Impurities in large scale produced Nd-doped phosphate laser glasses. I. Cu ions, Opt. Mater.: X, 2019, vol. 4, pp. 1–7. https://doi.org/10.1016/j.omx.2019.10003310.1016/j.omx.2019.100033 ↗
5. Patil, A.B., Tarik, M., Schuler, A.J., et al., Insights about inductively coupled plasma optical emission spectroscopy interferences of major rare earth elements in complex e-waste feeds, Spectrochim. Acta, Part B, 2022, vol. 191, p. 106399. https://doi.org/10.1016/j.sab.2022.10639910.1016/j.sab.2022.106399 ↗
6. Lorenz, T. and Bertau, M., Recycling of rare earth elements from FeNdB-Magnets via solid-state chlorination, J. Cleaner Prod., 2019, vol. 215, pp. 131–143. https://doi.org/10.1016/j.jclepro.2019.01.05110.1016/j.jclepro.2019.01.051 ↗
7. Spivakov, A., Chun-Rong, L., Chang-Yen, T., and Ying-Zhen, C., Size-dependent magnetic and magneto-optical properties of bi-doped yttrium iron garnet nanopowders, Nanoscale Res. Lett., 2022, vol. 17, pp. 70. https://doi.org/10.1186/s11671-022-03709-010.1186/s11671-022-03709-0 ↗
8. Bonin, M., Fontaine, F.G., and Larivière, D., Comparative studies of digestion techniques for the dissolution of neodymium-based magnets, Metals, 2021, vol. 11, p. 1149. https://doi.org/10.3390/met1108114910.3390/met11081149 ↗
9. Balaram, V. and Subramanyam K.S.V., Sample preparation for geochemical analysis: Strategies and significance, Adv. Sample Prep., 2022, vol. 1, p. 100010. https://doi.org/10.1016/j.sampre.2022.10001010.1016/j.sampre.2022.100010 ↗
10. Korotkova, N.A., Baranovskaya, V.B., and Petrova, K.V., Microwave digestion and ICP-MS determination of major and trace elements in waste Sm-Co magnets, Metals, 2022, vol. 12, p. 1308. https://doi.org/10.3390/met1208130810.3390/met12081308 ↗
11. Smirnova, M.N., Goeva, L.V., Simonenko, N.P., et al., Gel formation specifics in the synthesis of Mg(Fe0.8Ga0.2)2O4 by the glycine-nitrate method, Russ. J. Inorg. Chem., 2016, vol. 61, no. 10, pp. 1354–1359. https://doi.org/10.7868/S0044457X1508019X10.7868/S0044457X1508019X ↗
12. Todoli, J.L. and Mermet, J.M., Acid interferences in atomic spectrometry: Analyte signal effects and subsequent reduction, Spectrochim. Acta, Part B, 1999, vol. 54, pp. 895–929. https://doi.org/10.1016/S0584-8547(99)00041-510.1016/S0584-8547(99)00041-5 ↗
13. Todoli, J.L. and Mermet, J.M., Matrix effects, in Liquid Sample Introduction in ICP Spectrometry. A Practical Guide, Todoli, J.-L. and Mermet, J.-M., Eds., Amsterdam: Elsevier, 2008, pp. 147–190. https://doi.org/10.1016/b978-0-444-53142-1.00006-610.1016/b978-0-444-53142-1.00006-6 ↗
14. Todoli, J.L., Gras, L., Hernandis, V., and Mora, J., Elemental matrix effects in ICP-AES, J. Anal. At. Spectrom., 2002, vol. 17, pp. 142–169. https://doi.org/10.1039/B009570M10.1039/B009570M ↗
15. Pupyshev, A.A., Spectral interferences and their correction in atomic emission spectral analysis, Zavod. Lab. Mater. Diagn., 2019, vol. 85, no. 1, part 2, pp. 15–32. https://doi.org/10.26896/1028-6861-2019-85-1-II-15-3210.26896/1028-6861-2019-85-1-II-15-32 ↗
16. Lagalante, A.F., Atomic emission spectroscopy: A tutorial review, Appl. Spectrosc. Rev., 2004, vol. 34, no. 3, pp. 191–207. https://doi.org/10.1081/ASR-10010084510.1081/ASR-100100845 ↗
17. Petrova, K.V., Baranovskaya, V.B., and Korotkova, N.A., Direct inductively coupled plasma optical emission spectrometry for analysis of waste samarium-cobalt magnets, Arab. J. Chem., 2022, vol. 15, p. 103501. https://doi.org/10.1016/j.arabjc.2021.10350110.1016/j.arabjc.2021.103501 ↗
18. Almeida, G., Filipini da Silveira, J.R., Nunis da Silva, A.L., et al., Additivity of optical emissions applied to neodymium and praseodymium quantification in metallic didymiumand (Nd, Pr)-Fe-B alloy samples by low-resolution atomic emission spectrometry: An evaluation of the mathematical approach used to solve spectral interferences, Anal. Chim. Acta, 2019, vol. 1085, pp. 21–28. https://doi.org/10.1016/j.aca.2019.07.04910.1016/j.aca.2019.07.049 ↗