Express X-Ray Fluorescence Analysis of Technical-Grade Tantalum and Niobium: from Raw Materials to Products
Source abstract
Abstract
Abstract The available approaches to detecting impurities in Ta- and Nb-based materials involve transferring a sample into a solution with subsequent isolation of impurities. This procedure is quite complicated and time-consuming. Therefore, it is interesting to study the possibilities of direct, in particular, X-ray fluorescence (XRF), analysis of solid samples of these materials. The conventional XRF scheme, which involves the experimental construction of calibration characteristics for each element being determined, requires a large number of reference samples containing a very wide range of impurities. In this work, a preliminary characterization of samples of technical tantalum and niobium and products based on them has been carried out. It is shown that, for initial materials, the XRF method can establish only a significant absence of impurities; however, even for sintered niobium hydride and tantalum powder, XRF can be used for the express estimation of the composition. It is proposed to use a Spectroscan MAX-GVM crystal diffraction spectrometer for the analysis and to construct the calibration using the standard software that implements the fundamental parameter method. In this case, the resulting impurity contents can differ from reference values by one to two orders of magnitude. Such accuracy is often sufficient to adjust the technology. The limits of the XRF detection of impurities in Ta- and Nb-based materials have been established. For the elements determined by K-series lines (from Ti to Co), the detection limits range within 30–60 ppm; for the elements determined by M-series lines (Ta), the detection limit is approximately 200 ppm; and for L-series lines (Nb), it is from 100 to 150 ppm.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Electron and X-Ray Spectroscopy Techniques · X-ray Spectroscopy and Fluorescence Analysis
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (15)
Show source-provided references
1. Mosichev, V.I., Nikolaev, G.I., and Kalinin, B.D., Metally i splavy. Analiz i issledovanie. Metody atomnoi spektroskopii. Atomno-emissionnyi, atomno-absorbtsionnyi i rentgenofluorestsentnyi analiz (Metals and Alloys. Analysis and Research. Methods of Atomic Spectroscopy. Atomic Emission, Atomic Absorption and X-Ray Fluorescence Analysis, The Handbook), St. Petersburg: Professional, 2006, vol. 2.
2. Karpov, Yu.A. and Baranovskaya, V.B., Issues of standardization of the methods of chemical analysis in metallurgy, Zavod. Lab. Diagn. Mater., 2019, vol. 85, no. 1, part 2, pp. 5–14. https://doi.org/10.26896/1028-6861-2019-85-1-II-5-1410.26896/1028-6861-2019-85-1-II-5-14 ↗
3. Karpov, Yu.A. and Baranovskaya, V.B., Methods for the analysis of rare and precious metals ways of development, Analytika, 2019, vol. 9, no. 1, pp. 40–47. https://doi.org/10.22184/2227-572X.2019.09.1.40.4710.22184/2227-572X.2019.09.1.40.47 ↗
4. Suvorova, D., Khudonogova, E., and Revenko, A., Development of the XRF determination technique for the Ga, Hf, and Ta contents in rare earth ores, Analit. Kontrol’, 2016, vol. 20, no. 4, pp. 344–354. https://doi.org/10.15826/analitika.2016.20.4.00910.15826/analitika.2016.20.4.009 ↗
5. Suvorova, D., Khudonogova, E., and Revenko, A., X‑ray fluorescence determination of Cs, Ce, Nd, and Ta concentrations in rocks of various composition, X‑ray Spectrom., 2017, vol. 46, no. 3, pp. 200–208. https://doi.org/10.1002/xrs.274710.1002/xrs.2747 ↗
6. Losev, N.F., Kolichestvennyi rentgenospektral’nyi fluorestsentnyi analiz (Quantitative X-ray Fluorescent Analysis), Moscow: Nauka, 1969.
7. Revenko, A.G., Rentgenospektral’nyi fluorestsentnyi analiz prirodnykh materialov (X-ray Spectral Fluorescence Analysis of Natural Materials), Novosibirsk: Nauka, 1994.
8. Zhizhin, I.P., Kalinin, B.D., Liunski, A.V., et al., X-ray fluorescence spectrometers 'Spectroscan Maks’. Analytical parameters, Analit. Kontrol’, 2002, vol. 6, no. 4, pp. 463–469.
9. Borkhodoev, V.Ya., About a correlation between the limits of detection and determination in X-ray fluorescence analysis, J. Anal. Chem., 2015, vol. 70, no. 11, pp. 1307–1312. https://doi.org/10.1134/S106193481509004X10.1134/S106193481509004X ↗
10. Borkhodoev, V.Ya., About the limit of detection in X‑ray fluorescence analysis, J. Anal. Chem., 2014, vol. 69, no. 11, pp. 1041–1046. https://doi.org/10.1134/S106193481411002110.1134/S1061934814110021 ↗
11. Sherman, J. and Winifred, J.M., Adjustment of an inverse matrix corresponding to changes in the elements of a given column or a given row of the original matrix, Ann. Math. Stat., 1950, vol. 21, no. 1, pp. 124–127. https://doi.org/10.1214/aoms/117772989310.1214/aoms/1177729893 ↗
12. Bondarenko, A.V., Belonovsky, A.V., and Katzman, Ya.M., Application of fundamental parameter method in X-ray fluorescence analysis of pulp products in ore concentration, Gorn. Prom-st', 2021, no. 5-2, pp. 84–88.
13. Mantler, M., Willis, J., Lachance, G., et al., Quantitative analysis, in Handbook of Practical X-ray Fluorescence Analysis, Beckhoff, B., Kanngießer, B., Langhoff, N., Wedell, R., and Wolff, H., Eds., Berlin: Springer, 2006, pp. 309–410. https://doi.org/10.1007/978-3-540-36722-2_510.1007/978-3-540-36722-2_5 ↗
14. Borkhodoev, V.Ya., Rentgenofluorestsentnyi analiz gornykh porod sposobom fundamental’nykh parametrov (X-ray Fluorescence Analysis of Rocks by the Method of Fundamental Parameters), Magadan: SVKNII FEB RAN, 1999.
15. Rousseau, R., Corrections for matrix effects in X-ray fluorescence analysis: A tutorial, Spectrochim. Acta, Part B, 2006, vol. 7, pp. 759–777. https://doi.org/10.1016/j.sab.2006.06.01410.1016/j.sab.2006.06.014 ↗