journal article·2023-07-17Open access

Brain imaging with portable low-field MRI

W. Taylor Kimberly, Annabel J. Sorby-Adams, Andrew G. Webb, Ed X. Wu, Rachel Beekman, Ritvij Bowry, Steven J. Schiff, Adam de Havenon, Francis X. Shen, Gordon Sze, Pamela Schaefer, Juan Eugenio Iglesias, Matthew S. Rosen, Kevin N. Sheth

Nature Reviews Bioengineering· Volume 1· Issue 9

Abstract

The advent of portable, low-field MRI (LF-MRI) heralds new opportunities in neuroimaging. Low power requirements and transportability have enabled scanning outside the controlled environment of a conventional MRI suite, enhancing access to neuroimaging for indications that are not well suited to existing technologies. Maximizing the information extracted from the reduced signal-to-noise ratio of LF-MRI is crucial to developing clinically useful diagnostic images. Progress in electromagnetic noise cancellation and machine learning reconstruction algorithms from sparse k-space data as well as new approaches to image enhancement have now enabled these advancements. Coupling technological innovation with bedside imaging creates new prospects in visualizing the healthy brain and detecting acute and chronic pathological changes. Ongoing development of hardware, improvements in pulse sequences and image reconstruction, and validation of clinical utility will continue to accelerate this field. As further innovation occurs, portable LF-MRI will facilitate the democratization of MRI and create new applications not previously feasible with conventional systems.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Advanced MRI Techniques and Applications · Atomic and Subatomic Physics Research · Advanced X-ray and CT Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (100+)

Show source-provided references
1. Pfahler, G. E. In Proceedings of the 5th Annual Meeting of the American Roentgen Ray Society Vol. 4, 175–183 (St. Louis, MO, 1904).
2. Burr, C. W., Pfahler, G. E. & Camp, C. D. Thrombosis of the midcerebral artery, causing aphasia and hemiplegia, with remarks on cerebral skiagraphy. J. Nerv. Ment. Dis. 31, 558 (1904).10.1097/00005053-190408000-00013 ↗
3. Ambrose, J. Computerized transverse axial scanning (tomography). 2. Clinical application. Br. J. Radiol. 46, 1023–1047 (1973).10.1259/0007-1285-46-552-1023 ↗
4. Hounsfield, G. N. Computerized transverse axial scanning (tomography). 1. Description system. Br. J. Radiol. 46, 1016–1022 (1973).10.1259/0007-1285-46-552-1016 ↗
5. Lauterbur, P. C. Image formation by induced local interactions. Examples employing nuclear magnetic resonance. 1973.Clin. Orthop. Relat. Res. 244, 3–6 (1989).10.1097/00003086-198907000-00002 ↗
6. Mansfield, P. & Maudsley, A. A. Planar spin imaging by NMR. J. Magn. Reson. 27, 101–119 (1977).
7. Raich, H. & Blümler, P. Design and construction of a dipolar Halbach array with a homogeneous field from identical bar magnets: NMR Mandhalas. Concepts Magn. Reson. Part B Magn. Reson. Eng. 23B, 16–25 (2004).10.1002/cmr.b.20018 ↗
8. Lother, S., Schiff, S. J., Neuberger, T., Jakob, P. M. & Fidler, F. Design of a mobile, homogeneous, and efficient electromagnet with a large field of view for neonatal low-field MRI. Magn. Reson. Mater. Phys. Biol. Med. 29, 691–698 (2016).10.1007/s10334-016-0525-8 ↗
9. Obungoloch, J. et al. Design of a sustainable prepolarizing magnetic resonance imaging system for infant hydrocephalus. Magnetic Reson. Mater. Phys. Biol. Med. 31, 665–676 (2018).10.1007/s10334-018-0683-y ↗
10. Harper, J. R. et al. An unmatched radio frequency chain for low-field magnetic resonance imaging. Front. Phys. https://doi.org/10.3389/fphy.2021.727536 (2022).10.3389/fphy.2021.727536 ↗
11. Sarracanie, M. et al. Low-cost high-performance MRI. Sci. Rep. 5, 15177 (2015).10.1038/srep15177 ↗
12. He, Y. et al. Use of 2.1 MHz MRI scanner for brain imaging and its preliminary results in stroke. J. Magn. Reson. 319, 106829 (2020).10.1016/j.jmr.2020.106829 ↗
13. Sheth, K. N. et al. Assessment of brain injury using portable, low-field magnetic resonance imaging at the bedside of critically ill patients. JAMA Neurol. 78, 41–47 (2020).10.1001/jamaneurol.2020.3263 ↗
14. Liu, Y. et al. A low-cost and shielding-free ultra-low-field brain MRI scanner. Nat. Commun. 12, 7238 (2021).
15. Halbach, K. Design of permanent multipole magnets with oriented rare earth cobalt material. Nucl. Instrum. Methods 169, 1–10 (1980).10.1016/0029-554X(80)90094-4 ↗
16. Blümler, P. & Casanova, F. In Mobile NMR and MRI: Developments and Applications 133–157 (The Royal Society of Chemistry, 2016).10.1039/9781782628095-00133 ↗
17. Soltner, H. & Blümler, P. Dipolar Halbach magnet stacks made from identically shaped permanent magnets for magnetic resonance. Concepts Magn. Reson. Part A 36, 211–222 (2010).10.1002/cmr.a.20165 ↗
18. Cooley, C. Z. et al. Two-dimensional imaging in a lightweight portable MRI scanner without gradient coils. Magn. Reson. Med. 73, 872–883 (2015).10.1002/mrm.25147 ↗
19. Stockmann, J. P., Cooley, C. Z., Guerin, B., Rosen, M. S. & Wald, L. L. Transmit array spatial encoding (TRASE) using broadband WURST pulses for RF spatial encoding in inhomogeneous B0 fields. J. Magn. Reson. 268, 36–48 (2016).10.1016/j.jmr.2016.04.005 ↗
20. Cooley, C. Z. et al. A portable scanner for magnetic resonance imaging of the brain. Nat. Biomed. Eng. 5, 229–239 (2021).10.1038/s41551-020-00641-5 ↗
21. Cooley, C. Z. et al. Design of sparse Halbach magnet arrays for portable MRI using a genetic algorithm. IEEE Trans. Magn. https://doi.org/10.1109/tmag.2017.2751001 (2018).10.1109/tmag.2017.2751001 ↗
22. Martens, M. A. et al. Insertable biplanar gradient coils for magnetic resonance imaging. Rev. Sci. Instrum. 62, 2639–2645 (1991).10.1063/1.1142245 ↗
23. While, P. T., Forbes, L. K. & Crozier, S. 3D gradient coil design for open MRI systems. J. Magn. Reson. 207, 124–133 (2010).10.1016/j.jmr.2010.08.017 ↗
24. Zhang, R. et al. An optimized target-field method for MRI transverse biplanar gradient coil design. Meas. Sci. Technol. 22, 125505 (2011).10.1088/0957-0233/22/12/125505 ↗
25. Matsuzawa, K., Abe, M., Kose, K. & Terada, Y. Oval gradient coils for an open magnetic resonance imaging system with a vertical magnetic field. J. Magn. Reson. 278, 51–59 (2017).10.1016/j.jmr.2017.03.009 ↗
26. Shen, S., Koonjoo, N., Kong, X., Rosen, M. S. & Xu, Z. Gradient coil design and optimization for an ultra-low-field MRI system. Appl. Magn. Reson. 53, 895–914 (2022).10.1007/s00723-022-01470-2 ↗
27. de Vos, B., Fuchs, P., O’Reilly, T., Webb, A. & Remis, R. Gradient coil design and realization for a halbach-based MRI system. IEEE Trans. Magn. 56, 1–8 (2020).10.1109/TMAG.2019.2958561 ↗
28. Přibil, J., Přibilová, A. & Frollo, I. Vibration and noise in magnetic resonance imaging of the vocal tract: differences between whole-body and open-air devices. Sensors https://doi.org/10.3390/s18041112 (2018).10.3390/s18041112 ↗
29. Sharp, J. C. & King, S. B. MRI using radiofrequency magnetic field phase gradients. Magn. Reson. Med. 63, 151–161 (2010).10.1002/mrm.22188 ↗
30. Wang, P. et al. Correcting image distortions from a nonlinear B1+-gradient field in frequency-modulated Rabi-encoded echoes. Magn. Reson. Med. https://doi.org/10.1002/mrm.29549 (2022).10.1002/mrm.29549 ↗
31. Torres, E. et al. B(1)-gradient-based MRI using frequency-modulated Rabi-encoded echoes. Magn. Reson. Med. 87, 674–685 (2022).10.1002/mrm.29002 ↗
32. Gruber, B., Froeling, M., Leiner, T. & Klomp, D. W. J. RF coils: a practical guide for nonphysicists. J. Magn. Reson. Imaging 48, 590–604 (2018).10.1002/jmri.26187 ↗
33. Shen, S. et al. An optimized quadrature RF receive coil for very-low-field (50.4 mT) magnetic resonance brain imaging. J. Magn. Reson. 342, 107269 (2022).10.1016/j.jmr.2022.107269 ↗
34. O’Reilly, T., Teeuwisse, W. M., de Gans, D., Koolstra, K. & Webb, A. G. In vivo 3D brain and extremity MRI at 50 mT using a permanent magnet Halbach array. Magn. Reson. Med. 85, 495–505 (2021).10.1002/mrm.28396 ↗
35. Shen, S., Xu, S., Koonjoo, N. & Rosen, M. S. Optimization of a close-fitting volume RF coil for brain imaging at 6.5 mT using linear programming. IEEE Trans. Biomed. Eng. 68, 1106–1114 (2021).10.1109/TBME.2020.3002077 ↗
36. de Vos, B. et al. Design, characterisation and performance of an improved portable and sustainable low-field MRI system. Front. Phys. https://doi.org/10.3389/fphy.2021.701157 (2021).10.3389/fphy.2021.701157 ↗
37. Tewari, S., Yousefi, S. & Webb, A. Deep neural-network based optimization for the design of a multi-element surface magnet for MRI applications. Inverse Probl. https://doi.org/10.1088/1361-6420/ac492a (2022).10.1088/1361-6420/ac492a ↗
38. Giovannetti, G. & Menichetti, L. Litz wire RF coils for low frequency NMR applications. Measurement 110, 116–120 (2017).10.1016/j.measurement.2017.06.031 ↗
39. Miltner, O. et al. Portable NMR-MOUSE (R): a new method and its evaluation of the Achilles tendon. Z. Orthop. Grenzgeb. 141, 148–152 (2003).10.1055/s-2003-38657 ↗
40. Perlo, J., Casanova, F. & Blumich, B. 3D imaging with a single-sided sensor: an open tomograph. J. Magn. Reson. 166, 228–235 (2004).10.1016/j.jmr.2003.10.018 ↗
41. Perlo, J. et al. Desktop MRI as a promising tool for mapping intra-aneurismal flow. Magn. Reson. Imaging 33, 328–335 (2015).10.1016/j.mri.2014.12.005 ↗
42. Srinivas, S. A. et al. External dynamic interference estimation and removal (EDITER) for low field MRI. Magn. Reson. Med. https://doi.org/10.1002/mrm.28992 (2021).10.1002/mrm.28992 ↗
43. Srinivas, S. A., Cooley, C. Z., Stockmann, J. P., McDaniel, P. C. & Wald, L. L. In Proceedings of International Society of Magnetic Resonance in Medicine 1269 (2020).
44. Rearick, T., Charvat, G. L., Rosen, M. S. & Rothberg, J. M. Noise Supression Methods and Apparatus. United States of America Patent US9797971B2 (2017).
45. Yuen, M. M. et al. Portable, low-field magnetic resonance imaging enables highly accessible and dynamic bedside evaluation of ischemic stroke. Sci. Adv. 8, eabm3952 (2022).10.1126/sciadv.abm3952 ↗
46. Dyvorne, H. et al. Freeing MRI from its Faraday cage with Interference Rejection. Proceedings of the International Society for Magnetic Resonance in Medicine 749 (2021).
47. Zhao, Y., Xiao, L., Liu, Y., Leong, A. T. L. & Wu, E. X. Electromagnetic interference elimination via active sensing and deep learning prediction for radiofrequency shielding-free MRI. NMR Biomed. 3864 (2023).10.1002/nbm.4956 ↗
48. Kibret, B., Teshome, A. K. & Lai, D. T. H. Analysis of the human body as an antenna for wireless implant communication. IEEE Trans. Antennas Propag. 64, 1466–1476 (2016).10.1109/TAP.2016.2526070 ↗
49. Sen, S., Maity, S. & Das, D. The body is the network: to safeguard sensitive data, turn flesh and tissue into a secure wireless channel. IEEE Spectr. 57, 44–49 (2020).10.1109/MSPEC.2020.9271808 ↗
50. Ingle, V., Kogon, S. & Manolakis, D. Statistical and Adaptive Signal Processing (Artech, 2005).
51. Stearns, S. D. In Advanced Topics in Signal Processing 246–288 (Prentice-Hall, Inc., 1987).
52. Lu, H. et al. Routine clinical brain MRI sequences for use at 3.0 Tesla. J. Magn. Reson. Imaging 22, 13–22 (2005).10.1002/jmri.20356 ↗
53. O’Reilly, T. & Webb, A. G. In vivo T1 and T2 relaxation time maps of brain tissue, skeletal muscle, and lipid measured in healthy volunteers at 50 mT. Magn. Reson. Med. https://doi.org/10.1002/mrm.29009 (2021).10.1002/mrm.29009 ↗
54. Ma, D. et al. Magnetic resonance fingerprinting. Nature 495, 187–192 (2013).10.1038/nature11971 ↗
55. Zhu, B., Liu, J. Z., Cauley, S. F., Rosen, B. R. & Rosen, M. S. Image reconstruction by domain-transform manifold learning. Nature 555, 487–492 (2018).10.1038/nature25988 ↗
56. Lin, D. J., Johnson, P. M., Knoll, F. & Lui, Y. W. Artificial intelligence for MR image reconstruction: an overview for clinicians. J. Magn. Reson. Imaging 53, 1015–1028 (2021).10.1002/jmri.27078 ↗
57. Xiao, L. et al. Partial Fourier reconstruction of complex MR images using complex-valued convolutional neural networks. Magn. Reson. Med. 87, 999–1014 (2022).10.1002/mrm.29033 ↗
58. Koonjoo, N., Zhu, B., Cody Bagnall, G., Bhutto, D. & Rosen, M. S. Boosting the signal-to-noise of low-field MRI with deep learning image reconstruction. Sci. Rep. 11, 8248 (2021).10.1038/s41598-021-87482-7 ↗
59. Waddington, D. E. J. et al. Real-time radial reconstruction with domain transform manifold learning for MRI-guided radiotherapy. Med Phys. 50, 1962–1974 (2023).10.1002/mp.16224 ↗
60. Lecun, Y., Bottou, L., Bengio, Y. & Haffner, P. Gradient-based learning applied to document recognition. Proc. IEEE 86, 2278–2324 (1998).10.1109/5.726791 ↗
61. Wang, Z., Chen, J. & Hoi, S. C. H. Deep learning for image super-resolution: a survey. IEEE Trans. Pattern Anal. Mach. Intell. 43, 3365–3387 (2021).10.1109/TPAMI.2020.2982166 ↗
62. Yang, W. et al. Deep learning for single image super-resolution: a brief review. IEEE Trans. Multimed. 21, 3106–3121 (2019).10.1109/TMM.2019.2919431 ↗
63. Cherukuri, V., Guo, T., Schiff, S. J. & Monga, V. Deep MR brain image super-resolution using spatio-structural priors. IEEE Trans. Image Process. https://doi.org/10.1109/tip.2019.2942510 (2019).10.1109/tip.2019.2942510 ↗
64. Goodfellow, I. et al. Generative adversarial networks. Adv. Neural Inf. Process. Syst. https://doi.org/10.1145/3422622 (2014).10.1145/3422622 ↗
65. Song, T. A., Chowdhury, S. R., Yang, F. & Dutta, J. PET image super-resolution using generative adversarial networks. Neural Netw. 125, 83–91 (2020).10.1016/j.neunet.2020.01.029 ↗
66. Sotiras, A., Davatzikos, C. & Paragios, N. Deformable medical image registration: a survey. IEEE Trans. Med. Imaging 32, 1153–1190 (2013).10.1109/TMI.2013.2265603 ↗
67. Kondrateva, E. et al. Domain Sshift in Computer Vision Models for MRI Data Analysis: An Overview. Vol. 11605 (SPIE, 2021).10.1117/12.2587872 ↗
68. Wang, M. & Deng, W. Deep visual domain adaptation: a survey. Neurocomputing 312, 135–153 (2018).10.1016/j.neucom.2018.05.083 ↗
69. Iglesias, J. et al. Quantitative Brain Morphometry of Portable Low-Field-Strength MRI Using Super-Resolution Machine Learning. Radiology 306, 3 (2023).10.1148/radiol.220522 ↗
70. Tobin, J. et al. Domain randomization for transferring deep neural networks from simulation to the real world. IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS), 23–30 (IEEE, 2017).10.1109/IROS.2017.8202133 ↗
71. Cohen, J., Luck, M. & Honari, S. In Medical Image Computing and Computer Assisted Intervention – MICCAI 2018 (eds Frangi, A. et al.), 529–536, (Springer, 2018).10.1007/978-3-030-00928-1_60 ↗
72. Malinin, A. & Gales, M. In Proceedings of the 32nd International Conference on Neural Information Processing Systems 7047–7058 (Curran Associates Inc., 2018).
73. Kendall, A. & Gal, Y. What uncertainties do we need in Bayesian deep learning for computer vision? 31st Conference on Neural Information Processing Systems vol. 30 (NIPS, 2017).
74. Gal, Y. & Ghahramani, Z. Dropout as a Bayesian approximation: representing model uncertainty in deep learning. PMLR 48, 1050–1059 (2016).
75. Lakshminarayanan, B., Pritzel, A. & Blundell, C. Simple and scalable predictive uncertainty estimation using deep ensembles. 31st Conference on Neural Information Processing Systems 6405–6416 (2017).
76. Sensoy, M., Kaplan, L. & Kandemir, M. In Proceedings of the 32nd International Conference on Neural Information Processing Systems 3183–3193 (Curran Associates Inc., 2018).
77. Mehrabi, N., Morstatter, F., Saxena, N., Lerman, K. & Galstyan, A. A survey on bias and fairness in machine learning. ACM Comput. Surv. 54, 1–35 (2021).10.1145/3457607 ↗
78. Petersen, E. et al. Feature robustness and sex differences in medical imaging: a case study in MRI-based Alzheimer’s disease detection. In Medical Image Computing and Computer Assisted Intervention – MICCAI 2022: 25th International Conference Proceedings 88–98 (2022).10.1007/978-3-031-16431-6_9 ↗
79. Dressel, J. & Farid, H. The accuracy, fairness, and limits of predicting recidivism. Sci. Adv. 4, eaao5580 (2018).10.1126/sciadv.aao5580 ↗
80. Larson, J., Mattu, S., Kirchner, L. & Angwin, J. Machine Bias https://www.propublica.org/article/machine-bias-risk-assessments-in-criminal-sentencingProPublica (2016).
81. Chouldechova, A., Benavides-Prado, D., Fialko, O. & Vaithianathan, R. In Proceedings of the 1st Conference on Fairness, Accountability and Transparency Vol. 81 (eds Sorelle, A. F. & Christo, W.) 134–148 (Proceedings of Machine Learning Research, 2018).
82. Puyol Anton, E. et al. Fairness in AI: are deep learning-based CMR segmentation algorithms biased. Eur. Heart J. https://doi.org/10.1093/eurheartj/ehab724.3055 (2021).10.1093/eurheartj/ehab724.3055 ↗
83. Ruijsink, B. et al. Fully automated, quality-controlled cardiac analysis from CMR: validation and large-scale application to characterize cardiac function. JACC: Cardiovasc. Imaging 13, 684–695 (2019).
84. Bellamy, R. K. E. et al. AI Fairness 360: An extensible toolkit for detecting and mitigating algorithmic bias. IBM J. Res. Dev. https://doi.org/10.1147/JRD.2019.2942287 (2019).10.1147/JRD.2019.2942287 ↗
85. Chalela, J. A. et al. Magnetic resonance imaging and computed tomography in emergency assessment of patients with suspected acute stroke: a prospective comparison. Lancet 369, 293–298 (2007).10.1016/S0140-6736(07)60151-2 ↗
86. Campbell, B. C. & Parsons, M. W. Imaging selection for acute stroke intervention. Int. J. Stroke 13, 554–567 (2018).10.1177/1747493018765235 ↗
87. Shah, S. et al. Screening with MRI for accurate and rapid stroke treatment: SMART. Neurology 84, 2438–2444 (2015).10.1212/WNL.0000000000001678 ↗
88. Sorensen, A. G. et al. Hyperacute stroke: evaluation with combined multisection diffusion-weighted and hemodynamically weighted echo-planar MR imaging. Radiology 199, 391–401 (1996).10.1148/radiology.199.2.8668784 ↗
89. Fiebach, J. B. et al. CT and diffusion-weighted MR imaging in randomized order: diffusion-weighted imaging results in higher accuracy and lower interrater variability in the diagnosis of hyperacute ischemic stroke. Stroke 33, 2206–2210 (2002).10.1161/01.STR.0000026864.20339.CB ↗
90. Thomalla, G. et al. Intravenous alteplase for stroke with unknown time of onset guided by advanced imaging: systematic review and meta-analysis of individual patient data. Lancet 396, 1574–1584 (2020).10.1016/S0140-6736(20)32163-2 ↗
91. Powers, W. J. et al. Guidelines for the early management of patients with acute ischemic stroke: 2019 update to the 2018 guidelines for the early management of acute ischemic stroke: a guideline for healthcare professionals from the American Heart Association/American Stroke Association. Stroke 50, e344–e418 (2019).10.1161/STROKEAHA.118.022606 ↗
92. Thomalla, G. et al. DWI-FLAIR mismatch for the identification of patients with acute ischaemic stroke within 4.5 h of symptom onset (PRE-FLAIR): a multicentre observational study. Lancet Neurol. 10, 978–986 (2011).10.1016/S1474-4422(11)70192-2 ↗
93. Kidwell, C. S. et al. Comparison of MRI and CT for detection of acute intracerebral hemorrhage. J. Am. Med. Assoc. 292, 1823–1830 (2004).10.1001/jama.292.15.1823 ↗
94. Yuen, M. M. et al. Qualitative description of ischemic stroke appearance on low-field, point-of-care magnetic resonance imaging. Stroke 52, A33–A33 (2021).
95. Mazurek, M. H. et al. Portable, bedside, low-field magnetic resonance imaging for evaluation of intracerebral hemorrhage. Nat. Commun. 12, 5119 (2021).10.1038/s41467-021-25441-6 ↗
96. Grotta, J. C. et al. Prospective, multicenter, controlled trial of mobile stroke units. N. Engl. J. Med. 385, 971–981 (2021).10.1056/NEJMoa2103879 ↗
97. Lahner, D. et al. Incidence of complications in intrahospital transport of critically ill patients–experience in an Austrian university hospital. Wien. Klin. Wochenschr. 119, 412–416 (2007).10.1007/s00508-007-0813-4 ↗
98. Papson, J. P., Russell, K. L. & Taylor, D. M. Unexpected events during the intrahospital transport of critically ill patients. Acad. Emerg. Med. 14, 574–577 (2007).10.1197/j.aem.2007.02.034 ↗
99. Venkategowda, P. M., Rao, S. M., Mutkule, D. P. & Taggu, A. N. Unexpected events occurring during the intra-hospital transport of critically ill ICU patients. Indian J. Crit. Care Med. 18, 354–357 (2014).10.4103/0972-5229.133880 ↗
100. Knight, P. H. et al. Complications during intrahospital transport of critically ill patients: focus on risk identification and prevention. Int. J. Crit. Illn. Inj. Sci. 5, 256–264 (2015).10.4103/2229-5151.170840 ↗