Quantitative X-ray fluorescence computed tomography for low-Z samples using an iterative absorption correction algorithm
Source abstract
Abstract
X-ray fluorescence computed tomography is often used to measure trace element distributions within low-Z samples, using algorithms capable of X-ray absorption correction when sample self-absorption is not negligible. Its reconstruction is more complicated compared to transmission tomography, and therefore not widely used. We describe in this paper a very practical iterative method that uses widely available transmission tomography reconstruction software for fluorescence tomography. With this method, sample self-absorption can be corrected not only for the absorption within the measured layer but also for the absorption by material beyond that layer. By combining tomography with analysis for scanning X-ray fluorescence microscopy, absolute concentrations of trace elements can be obtained. By using widely shared software, we not only minimized the coding, took advantage of computing efficiency of fast Fourier transform in transmission tomography software, but also thereby accessed well-developed data processing tools coming with well-known and reliable software packages. The convergence of the iterations was also carefully studied for fluorescence of different attenuation lengths. As an example, fish eye lenses could provide valuable information about fish life-history and endured environmental conditions. Given the lens’s spherical shape and sometimes the short distance from sample to detector for detecting low concentration trace elements, its tomography data are affected by absorption related to material beyond the measured layer but can be reconstructed well with our method. Fish eye lens tomography results are compared with sliced lens 2D fluorescence mapping with good agreement, and with tomography providing better spatial resolution.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Advanced X-ray and CT Imaging · Medical Imaging Techniques and Applications · Radiation Dose and Imaging
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Cornell University 1 Cornell High Energy Synchrotron Source, , Ithaca, New York 14853, USA
SUNY College of Environmental Science and Forestry 2 Environmental and Forest Biology, , Syracuse, New York 13210, USA
University of Tartu 3 Estonian Marine Institute, , Tartu, Tartumaa 51014, Estonia
Funding
National Science Foundation
References (45)
Show source-provided references
1. Rev. Sci. Instrum. · 1989 · 60 · 245810.1063/1.1140698 ↗
2. Proc. SPIE · 1999 · 3772 · 30410.1117/12.363733 ↗
3. 2000
4. Proc. SPIE · 2000 · 4142 · 28710.1117/12.410573 ↗
5. AIP Conference Proceedings · 2007 · 879 · 133710.1063/1.2436311 ↗
6. Nucl. Instrum. Methods Phys. Res. B · 2013 · 305 · 510.1016/j.nimb.2013.04.045 ↗
7. J. Synchrotron Rad. · 2014 · 21 · 103110.1107/s160057751401621x ↗
8. Environ. Sci. Technol. · 2001 · 35 · 386310.1021/es0105459 ↗
9. Proc. SPIE · 2002 · 4503 · 22210.1117/12.452848 ↗
10. Proc. SPIE · 2002 · 4503 · 24010.1117/12.452865 ↗
11. Environ Sci Technol. · 2004 · 38 · 607410.1021/es049448g ↗
12. Science · 2006 · 314 · 129510.1126/science.1132563 ↗
13. J. Theor. Biol. · 1972 · 36 · 10510.1016/0022-5193(72)90180-4 ↗
14. 1988
15. IEEE Trans. Nucl. Sci. · 1998 · 45 · 7510.1109/23.659557 ↗
16. X-Ray Spectrom. · 2002 · 31 · 44810.1002/xrs.590 ↗
17. J. Phys. IV France · 2003 · 104 · 62710.1051/jp4:20030158 ↗
18. J. Appl. Phys. · 2003 · 94 · 14510.1063/1.1578176 ↗
19. IEEE Trans. Med. Imaging · 1982 · 2 · 113
20. IEEE Trans. Med. Imaging · 1994 · 13 · 60110.1109/42.363108 ↗
21. J. Synchrotron Rad. · 2011 · 19 · 201
22. IEEE Trans. Nucl. Sci. · 1997 · 40 · 54
23. IEEE Trans. Nucl. Sci. · 1991 · 38 · 172110.1109/23.124168 ↗
24. Comput. Phys. Commun. · 2001 · 141 · 41210.1016/s0010-4655(01)00419-2 ↗
25. Nucl. Instr. Meth. A. · 2007 · 579 · 32210.1016/j.nima.2007.04.069 ↗
26. IEEE Trans. Med. Imaging · 2006 · 25 · 111710.1109/tmi.2006.877441 ↗
27. Appl. Phys. Lett. · 2001 · 79 · 191210.1063/1.1402643 ↗
28. IEEE Trans. Med. Imaging · 2011 · 30 · 43810.1109/tmi.2010.2085011 ↗
29. Environ. Sci. Technol. · 2005 · 39 · 221010.1021/es0492034 ↗
30. Nature Communications · 2016 · 7 · 1263410.1038/ncomms12634 ↗
31. Powder Diffr. · 2010 · 25 · 16910.1154/1.3397114 ↗
32. Science · 2006 · 314 · 129510.1126/science.1132563 ↗
33. Plant Physiology · 2010 · 152 · 30910.1104/pp.109.146126 ↗
34. Environ. Sci. Technol. · 2010 · 44 · 810810.1021/es101139z ↗
35. PLoS One · 2011 · 6 · e2062610.1371/journal.pone.0020626 ↗
36. Metallomics · 2014 · 6 · 164810.1039/c4mt00121d ↗
37. Spectrochimica Acta B · 2007 · 62 · 6310.1016/j.sab.2006.12.002 ↗
38. J. Marine Syst. · 2015 · 141 · 16710.1016/j.jmarsys.2014.02.014 ↗
39. Marine Biology · 1998 · 130 · 37710.1007/s002270050258 ↗
40. Fishery Bulletin · 2001 · 99 · 410
41. Proc. SPIE · 2004 · 5535 · 23210.1117/12.558948 ↗
42. J. Synchrotron Rad. · 2014 · 21 · 118810.1107/s1600577514013939 ↗
43. J. Synchrotron Rad. · 2006 · 13 · 7410.1107/s0909049505038562 ↗
44. IEEE Trans. Nucl. Sci. · 1974 · 21 · 7810.1109/TNS.1974.6499238 ↗
45. Applied Optics · 1993 · 32 · 373610.1364/ao.32.003736 ↗