Design and validation of magnetic particle spectrometer for characterization of magnetic nanoparticle relaxation dynamics
Source abstract
Abstract
The design and validation of a magnetic particle spectrometer (MPS) system used to study the linear and nonlinear behavior of magnetic nanoparticle suspensions is presented. The MPS characterizes the suspension dynamic response, both due to relaxation and saturation effects, which depends on the magnetic particles and their environment. The system applies sinusoidal excitation magnetic fields varying in amplitude and frequency and can be configured for linear measurements (1 mT at up to 120 kHz) and nonlinear measurements (50 mT at up to 24 kHz). Time-resolved data acquisition at up to 4 MS/s combined with hardware and software-based signal processing allows for wide-band measurements up to 50 harmonics in nonlinear mode. By cross-calibrating the instrument with a known sample, the instantaneous sample magnetization can be quantitatively reconstructed. Validation of the two MPS modes are performed for iron oxide and cobalt ferrite suspensions, exhibiting Néel and Brownian relaxation, respectively.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Characterization and Applications of Magnetic Nanoparticles · Geomagnetism and Paleomagnetism Studies · Minerals Flotation and Separation Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
University of Florida 1 Department of Electrical and Computer Engineering, , Gainesville, Florida 32608, USA
University of Florida 2 Department of Chemical Engineering, , Gainesville, Florida 32608, USA
University of Florida 3 J. Crayton Pruitt Family Department of Biomedical Engineering, , Gainesville, Florida 32608, USA
Funding
National Institutes of Health · University of Florida
References (13)
Show source-provided references
1. Nature · 2005 · 435 · 121410.1038/nature03808 ↗
2. Rev. of Sci. Instrum. · 2012 · 83 · 03370810.1063/1.3694534 ↗
3. Phys. Med. Biol. · 2009 · 54 · L110.1088/0031-9155/54/5/l01 ↗
4. Theranostics · 2016 · 6 · 29110.7150/thno.13728 ↗
5. Nature Scientific Reports · 2016 · 6 · 3418010.1038/srep34180 ↗
6. J. Magn. Magn. Mater. · 2016 · 419 · 26710.1016/j.jmmm.2016.06.038 ↗
7. J. Phys. D-Applied Physics · 2009 · 42 · 231310.1088/0022-3727/42/20/205007 ↗
8. J. Phys. D-Applied Physics · 2014 · 47 · 15500110.1088/0022-3727/47/15/155001 ↗
9. Med. Phys. · 2008 · 35 · 198810.1118/1.2961396 ↗
10. Radiol. Phys. Technol. · 2013 · 6 · 39910.1007/s12194-013-0213-6 ↗
11. Nature Mater. · 2004 · 3 · 89110.1038/nmat1251 ↗
12. J. Colloid and Interface Science · 2012 · 377 · 4010.1016/j.jcis.2012.03.050 ↗
13. J. Magn. Magn. Mater. · 2013 · 328 · 4110.1016/j.jmmm.2012.09.069 ↗