The effects of degree of crosslinking on the fatigue crack initiation and propagation resistance of orthopedic‐grade polyethylene
Source abstract
Abstract
Abstract Crosslinked ultrahigh molecular weight polyethylene (UHMWPE) has been recently approved by the Food and Drug Administration for use in orthopedic implants. The majority of commercially available UHMWPE orthopedic components are crosslinked using e‐beam or gamma radiation. The level of crosslinking is controlled with radiation dose and free radicals are eliminated through heat treatments to prevent long‐term degradation associated with chain scission or oxidation mechanisms. Laboratory studies have demonstrated a substantial improvement in the wear resistance of crosslinked UHMWPE. However, a concern about the resistance to fatigue damage remains in the clinical community, especially for tibial components that sustain high cyclic contact stresses. The objective of this study was to investigate both the initiation and propagation aspects of fatigue cracks in radiation crosslinked medical‐grade UHMWPE. This work evaluated three levels of radiation, which induced three crosslink densities, on the fatigue crack propagation and total fatigue life behavior. Both as‐received UHMWPE, as well as those that underwent an identical thermal history as the crosslinked UHMWPE were used as controls. Fractured crack propagation specimens were examined using scanning electron microscopy to elucidate fatigue fracture mechanisms. The results of this work indicated that a low crosslink density may optimize the fatigue resistance from both a crack initiation and propagation standpoint. © 2003 Wiley Periodicals, Inc. J Biomed Mater Res 66A: 146–154, 2003
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Titanium Alloys Microstructure and Properties
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (24)
Show source-provided references
1. JastyM BragdonCR O'ConnorDO MuratogluOK PremnathV MerrillEW HarrisWH.Marked improvement in the wear resistance of a new form of UHMWPE in a physiologic hip simulator. In: Transactions of the 43rd Annual Meeting of the Orthopedic Research Society San Francisco;1997. p785.
2. 10.1016/S0142-9612(99)00039-310.1016/S0142-9612(99)00039-3 ↗
3. 10.1002/jor.110017020310.1002/jor.1100170203 ↗
4. 10.1016/S0883-5403(99)90086-410.1016/S0883-5403(99)90086-4 ↗
5. Improvement of polyethylene by irradiation in artificial joints
6. SunDC WangA StarkC DumbletonJH.The concept of stabilization in UHMWPE. In: Transactions of the 5th World Biomaterials Congress;1996. p195.
7. 10.1016/S0043-1648(96)07362-010.1016/S0043-1648(96)07362-0 ↗
8. 10.1016/S0142-9612(99)00038-110.1016/S0142-9612(99)00038-1 ↗
9. GomollA WanichT BellareA.Fracture toughness of ultrahigh molecular weight polyethylene used in orthopedic implants. In: Transactions of 11th International Conference on Deformation Yield and Fracture of Polymers Cambridge UK;2000. p254–257.
10. ShenF‐W McKellopH.Long term stability of chemically crosslinked UHMW polyethylene acetabular cups. In: Transactions of 27th Annual Society for Biomaterials Conference Minneapolis;2001.
11. MuratogluOK BiggsSA BragdonCR O'ConnorDO MerrillEW PremnathV JastyM HarrisWH.Long term stability of radiation and peroxide crosslinked UHMWPE. In: Transactions of 23rd Annual Society for Biomaterials Conference;1997. p776.
12. 10.1002/(SICI)1097-4636(19990915)46:4<573::AID-JBM16>3.0.CO;2-A10.1002/(SICI)1097-4636(19990915)46:4<573::AID-JBM16>3.0.CO;2-A ↗
13. GreenwaldAS BauerTW RiesMD.New polys for old: Contribution or caveat. In: Transactions of 68th Annual Meeting of the American Academy of Orthopaedic Surgeons San Francisco;2001.
14. Wright T · 2002
15. 10.2106/00004623-198668070-0001010.2106/00004623-198668070-00010 ↗
16. O'ConnorDO MuratogluOK BragdonCR LowensteinJ JastyM HarrisWH.Wear and high cycle fatigue of highly crosslinked UHMWPE. In: Transactions of 44th Annual Meeting of the Orthopaedic Research Society Anaheim CA;1999. p816.
17. KrzypowDJ BensusanJ SevoK HaggardW ParrJ GoldbergV RimnacC.The fatigue crack propagation resistance of gamma radiation or peroxide crosslinked UHMW polyethylene. In: Transactions of 6th World Biomaterials Congress Hawaii;2000. p382.
18. 10.1016/S0032-3861(97)00448-510.1016/S0032-3861(97)00448-5 ↗
19. 10.1016/S0032-3861(99)00199-810.1016/S0032-3861(99)00199-8 ↗
20. Hertzberg RW · 1980
21. 10.1016/S0142-9612(98)00112-410.1016/S0142-9612(98)00112-4 ↗
22. 10.1002/jor.110002020210.1002/jor.1100020202 ↗
23. 10.1023/A:101096902990410.1023/A:1010969029904 ↗
24. 10.1080/0022234990824814010.1080/00222349908248140 ↗