Thrombogenicity assessment of Pipeline Flex, Pipeline Shield, and FRED flow diverters in an in vitro human blood physiological flow loop model

Gaurav Girdhar, Arielle Andersen, Elizabeth Pangerl, Reza Jahanbekam, Samantha Ubl, Kevin Nguyen, John Wainwright, Michael F. Wolf

Abstract

Abstract Endovascular treatment of intracranial aneurysms with endoluminal flow diverters (single or multiple) has proven to be clinically safe and effective, but is associated with a risk of thromboembolic complications. Recently, a novel biomimetic surface modification with covalently bound phosphorylcholine (Shield Technology™) has shown to reduce the material thrombogenicity of the Pipeline flow diverter. Thrombogenicity of Pipeline Flex, Pipeline Shield, and Flow Redirection Endoluminal Device (FRED) in the presence of human blood under physiological flow conditions—in addition to relative increase in thrombogenicity with multiple devices—remains unknown and was investigated here. Thrombin generation (mean ± SD; μg/mL; thrombin–antithrombin complex or TAT) was measured as FRED (30.3 ± 2.9), Pipeline (13.9 ± 4.4), Pipeline Shield (0.4 ± 0.3), and negative control (no device; 0.1 ± 0.0). Platelet activation (mean ± SD; IU/μL; beta‐thromboglobulin or βTG) was measured as FRED (148 ± 45), Pipeline (92.8 ± 41), Pipeline Shield (16.2 ± 3.5), and negative control (2.70 ± 0.16). FRED was significantly more thrombogenic than Pipeline and Pipeline Shield (p < 0.05) for TAT. Additionally, Pipeline Shield had significantly lower TAT and βTG than the other devices tested (p < 0.05) and these were comparable to the negative control (p > 0.05). TAT and βTG scaled proportionately with multiple Pipeline devices (N = 6) but was unaffected by multiple Pipeline Shield (N = 6) devices—the latter being statistically similar to negative control (p > 0.05). © 2018 The Authors. Journal Of Biomedical Materials Research Part A Published By Wiley Periodicals, Inc. J Biomed Mater Res Part A: 106A: 3195–3202, 2018.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Intracranial Aneurysms: Treatment and Complications · Acute Ischemic Stroke Management · Vascular Malformations Diagnosis and Treatment

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Medtronic Neurovascular Irvine California

Medtronic Core Technologies Minneapolis Minnesota

References (27)

Show source-provided references
1. 10.1179/1743132814Y.000000031810.1179/1743132814Y.0000000318 ↗
2. Flow diverters for intracranial aneurysms
3. 10.1136/neurintsurg-2014-01147510.1136/neurintsurg-2014-011475 ↗
4. 10.3174/ajnr.A242110.3174/ajnr.A2421 ↗
5. 10.1179/1743132814Y.000000045010.1179/1743132814Y.0000000450 ↗
6. 10.1148/radiol.1312009910.1148/radiol.13120099 ↗
7. The Pipeline flow‐diverting stent for exclusion of ruptured intracranial aneurysms with difficult morphologies
8. 10.1136/neurintsurg-2016-01296910.1136/neurintsurg-2016-012969 ↗
9. 10.1007/s11239-015-1228-010.1007/s11239-015-1228-0 ↗
10. 10.1136/neurintsurg-2016-01261210.1136/neurintsurg-2016-012612 ↗
11. 10.1136/neurintsurg-2017-01317510.1136/neurintsurg-2017-013175 ↗
12. 10.1136/neurintsurg-2016-01289610.1136/neurintsurg-2016-012896 ↗
13. 10.1136/neurintsurg-2017-01355910.1136/neurintsurg-2017-013559 ↗
14. Treatment of intra‐ and extracranial aneurysms using the flow‐redirection endoluminal device: Multicenter experience and follow‐up results
15. Comparison of modified chandler, roller pump, and ball valve circulation models for in vitro testing in high blood flow conditions: Application in thrombogenicity testing of different materials for vascular applications
16. Use of simple and complex in vitro models for multiparameter characterization of human blood‐material/device interactions
17. An examination of the fluid mechanics and thrombus formation time parameters in a Chandler rotating loop system
18. 10.1007/s10856-011-4335-210.1007/s10856-011-4335-2 ↗
19. 10.1016/S1051-0443(07)61868-510.1016/S1051-0443(07)61868-5 ↗
20. Blood‐flow volume quantification in internal carotid and vertebral arteries: Comparison of 3 different ultrasound techniques with phase‐contrast MR imaging
21. 10.1160/TH08-06-039810.1160/TH08-06-0398 ↗
22. 10.1177/107602961667373210.1177/1076029616673732 ↗
23. 10.1016/0049-3848(87)90274-X10.1016/0049-3848(87)90274-X ↗
24. 10.1111/j.1538-7836.2005.01249.x10.1111/j.1538-7836.2005.01249.x ↗
25. 10.1160/TH08-02-010310.1160/TH08-02-0103 ↗
26. 10.1136/bcr-2017-21940610.1136/bcr-2017-219406 ↗
27. 10.1136/neurintsurg-2016-012657.rep10.1136/neurintsurg-2016-012657.rep ↗