Effects of cyclic flexural fatigue on porcine bioprosthetic heart valve heterograft biomaterials

Ali Mirnajafi, Brett Zubiate, Michael S. Sacks

Publisher page

Abstract

Abstract Although bioprosthetic heart valves (BHV) remain the primary treatment modality for adult heart valve replacement, continued problems with durability remain. Several studies have implicated flexure as a major damage mode in porcine‐derived heterograft biomaterials used in BHV fabrication. Although conventional accelerated wear testing can provide valuable insights into BHV damage phenomena, the constituent tissues are subjected to complex, time‐varying deformation modes (i.e., tension and flexure) that do not allow for the control of the amount, direction, and location of flexure. Thus, in this study, customized fatigue testing devices were developed to subject circumferentially oriented porcine BHV tissue strips to controlled cyclic flexural loading. By using this approach, we were able to study layer‐specific structural damage induced by cyclic flexural tensile and compressive stresses alone. Cycle levels of 10 × 106, 25 × 106, and 50 × 106 were used, with resulting changes in flexural stiffness and collagen structure assessed. Results indicated that flexural rigidity was markedly reduced after only 10 × 106 cycles, and progressively decayed at a lower rate with cycle number thereafter. Moreover, the against‐curvature fatigue direction induced the most damage, suggesting that the ventricularis and fibrosa layers have low resistance to cyclic flexural compressive and tensile loads, respectively. The histological analyses indicated progressive collagen fiber delamination as early as 10 × 106 cycles but otherwise no change in gross collagen orientation. Our results underscore that porcine‐derived heterograft biomaterials are very sensitive to flexural fatigue, with delamination of the tissue layers the primary underlying mechanism. This appears to be in contrast to pericardial BHV, wherein high tensile stresses are considered to be the major cause of structural failure. These findings point toward the need for the development of chemical fixation technologies that minimize flexure‐induced damage to extend porcine heterograft biomaterial durability. © 2010 Wiley Periodicals, Inc. J Biomed Mater Res, 2010
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Cardiac Valve Diseases and Treatments · Electrospun Nanofibers in Biomedical Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (48)

Show source-provided references
1. 10.1586/17434440.3.6.81710.1586/17434440.3.6.817 ↗
2. Schoen FJ · 1991 · 125
3. 10.1002/(SICI)1097-4636(19991215)47:4<439::AID-JBM1>3.0.CO;2-O10.1002/(SICI)1097-4636(19991215)47:4<439::AID-JBM1>3.0.CO;2-O ↗
4. 10.1111/j.1540-8191.1994.tb00932.x10.1111/j.1540-8191.1994.tb00932.x ↗
5. 10.1097/00002480-199901000-0001410.1097/00002480-199901000-00014 ↗
6. Sacks MS · 1999
7. 10.1002/(SICI)1097-4636(199907)46:1<44::AID-JBM5>3.0.CO;2-D10.1002/(SICI)1097-4636(199907)46:1<44::AID-JBM5>3.0.CO;2-D ↗
8. 10.1016/j.biomaterials.2004.06.04610.1016/j.biomaterials.2004.06.046 ↗
9. 10.1002/jbm.1029310.1002/jbm.10293 ↗
10. 10.1016/0002-9149(78)90873-110.1016/0002-9149(78)90873-1 ↗
11. SacksMS LiaoJ.Collagen fiber moduli of the aortic and mitral heart valve leaflets. Submitted for Publication.
12. 10.1615/JLongTermEffMedImplants.v11.i34.10010.1615/JLongTermEffMedImplants.v11.i34.100 ↗
13. 10.1016/S0002-9440(10)62227-410.1016/S0002-9440(10)62227-4 ↗
14. 10.1016/j.biomaterials.2005.08.00310.1016/j.biomaterials.2005.08.003 ↗
15. Prosthetic Devices Branch, D.o.C. Respiratory and Neurological Devices · 1993
16. Tissue damage and calcification may be independent mechanisms of bioprosthetic heart valve failure
17. 10.1016/0021-9290(94)90027-210.1016/0021-9290(94)90027-2 ↗
18. Mechanical design of elastic biopolymers
19. Fatigue‐induced damage in gluteraldehyde‐preserved heart valve tissue10.1016/S0022-5223(19)41138-0 ↗
20. 10.1016/0021-9290(77)90086-010.1016/0021-9290(77)90086-0 ↗
21. 10.1016/0142-9612(80)90049-610.1016/0142-9612(80)90049-6 ↗
22. 10.1016/S0142-9612(03)00051-610.1016/S0142-9612(03)00051-6 ↗
23. Focus on materials with scattered light
24. 10.1007/BF0268484510.1007/BF02684845 ↗
25. Thubrikar M · 1990
26. 10.1098/rstb.2007.212210.1098/rstb.2007.2122 ↗
27. 10.1115/1.276811110.1115/1.2768111 ↗
28. 10.1002/jbm.a.3087110.1002/jbm.a.30871 ↗
29. 10.1002/jbm.a.3003110.1002/jbm.a.30031 ↗
30. The effects of leaflet stiffness on the dynamic geometry of the bioprosthetic aortic heart valve
31. Frisch‐Fay R · 1962
32. 10.1097/00002480-199407000-0007510.1097/00002480-199407000-00075 ↗
33. 10.1016/j.carpath.2006.05.00110.1016/j.carpath.2006.05.001 ↗
34. 10.3109/0300820780915228510.3109/03008207809152285 ↗
35. Fatigue‐induced damage in glutaraldehyde preserved heart valve tissue
36. 10.1016/j.athoracsur.2004.06.03310.1016/j.athoracsur.2004.06.033 ↗
37. A new in vitro test method for calcification of bioprosthetic heart valves10.1177/039139889702000506 ↗
38. In vitro evaluation of the calcification behavior of polyurethane biomaterials for cardiovascular applications
39. 10.1016/j.jbiomech.2009.05.01510.1016/j.jbiomech.2009.05.015 ↗
40. Aortic valve structure‐function correlations: Role of elastic fibers no longer a stretch of the imagination
41. 10.1007/978-1-4757-2257-410.1007/978-1-4757-2257-4 ↗
42. 10.1002/1097-4636(20010915)56:4<478::AID-JBM1119>3.0.CO;2-C10.1002/1097-4636(20010915)56:4<478::AID-JBM1119>3.0.CO;2-C ↗
43. BilliarK SacksM.Long‐term mechanical fatigue response of porcine bioprosthetic heart valves. International Mechanical Engineering Congress and Exposition Anaheim CA November 15–20 1998 Advances in Bioengineering.10.1115/IMECE1998-0165 ↗
44. Fatigue‐induced changes to the biaxial mechanical properties of glutaraldehyde‐fixed porcine aortic valve leaflets
45. Smith DB · 1996 · 78
46. 10.1016/j.biomaterials.2007.08.02110.1016/j.biomaterials.2007.08.021 ↗
47. 10.1016/j.biomaterials.2006.09.02910.1016/j.biomaterials.2006.09.029 ↗
48. 10.1016/j.jorganchem.2004.10.01110.1016/j.jorganchem.2004.10.011 ↗