The effect of the inflammatory species hypochlorous acid on the corrosion and surface damage of Ti‐6Al‐4V and CoCrMo alloys
Source abstract
Abstract
Abstract Performance of metallic orthopedic alloys is dependent on the local biological conditions within the joint. Inflammation within a joint can lead to severe conditions at the surface of the implant and may represent a driving force of corrosion which is not well understood. In this study, the corrosion response of CoCrMo and Ti‐6Al‐4V alloys was studied in hypochlorous acid (HOCl) saline solutions, which served as a simulated inflammatory environment. Anodic polarization and electrochemical impedance spectroscopy were used to evaluate the electrochemical behavior of these alloys in a range of hypochlorous acid solutions. The open circuit potential of CoCrMo rose to over 600 mV (vs. Ag/AgCl) in 30 mM HOCl solutions while the corresponding corrosion current increased three orders of magnitude over controls. Visual analysis of CoCrMo disks after immersion in 50 mM HOCl for 5 days revealed extensive surface damage. This study shows that HOCl can polarize CoCrMo surfaces above the transpassive potential without an external power source. Ultimately, long‐term exposure to HOCl within an inflamed joint may significantly affect patient outcomes through accelerated generation of corrosion products. © 2018 Wiley Periodicals, Inc. J Biomed Mater Res Part A: 106A: 3185–3194, 2018.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Nanoparticles: synthesis and applications · Trace Elements in Health
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Bioengineering Clemson University Clemson South Carolina
Clemson‐Medical University of South Carolina Bioengineering Program Syracuse Biomaterials Institute Syracuse New York
Department of Biomedical and Chemical Engineering Syracuse University Syracuse New York
Funding
DePuy Synthes · South Carolina SmartState Program · Hansörg Wyss Endowment
References (29)
Show source-provided references
1. Ratner BD · 2013
2. 10.1016/B978-0-08-055294-1.00013-110.1016/B978-0-08-055294-1.00013-1 ↗
3. 10.1016/B978-0-12-803269-5.00003-610.1016/B978-0-12-803269-5.00003-6 ↗
4. Gilbert J · 2013 · 1
5. Biologic responses to orthopedic implants: Innate and adaptive immune responses to implant debris
6. 10.1016/j.biomaterials.2011.05.07810.1016/j.biomaterials.2011.05.078 ↗
7. 10.1189/jlb.101254410.1189/jlb.1012544 ↗
8. 10.1056/NEJM19890209320060610.1056/NEJM198902093200606 ↗
9. 10.2106/00004623-199401000-0001410.2106/00004623-199401000-00014 ↗
10. 10.5006/256310.5006/2563 ↗
11. 10.1002/jbm.b.3397210.1002/jbm.b.33972 ↗
12. 10.1002/jbm.b.3306710.1002/jbm.b.33067 ↗
13. 10.1016/j.electacta.2014.01.14310.1016/j.electacta.2014.01.143 ↗
14. Jones DA · 1996
15. 10.2106/JBJS.L.0104210.2106/JBJS.L.01042 ↗
16. 10.1007/s11999-014-3902-510.1007/s11999-014-3902-5 ↗
17. 10.1002/jbm.b.3383010.1002/jbm.b.33830 ↗
18. 10.1002/jbm.b.3312310.1002/jbm.b.33123 ↗
19. 10.1134/S000629791313007510.1134/S0006297913130075 ↗
20. BaiZhijun GilbertJeremy ChandrasekaranNithya inventors; Syracuse University assignee. Method for preparing biomedical surfaces. patent US2012022650A1. Jan 26 2012.
21. The oxidation of metals at high temperature
22. 10.1016/S0010-938X(97)89244-X10.1016/S0010-938X(97)89244-X ↗
23. 10.1016/j.electacta.2003.12.04310.1016/j.electacta.2003.12.043 ↗
24. 10.2174/092986705376463510.2174/0929867053764635 ↗
25. 10.1016/j.freeradbiomed.2012.06.00210.1016/j.freeradbiomed.2012.06.002 ↗
26. Pourbaix M. · 1966
27. 10.2106/00004623-199802000-0001510.2106/00004623-199802000-00015 ↗
28. 10.1302/2046-3758.14.200004710.1302/2046-3758.14.2000047 ↗
29. 10.1002/jor.2243110.1002/jor.22431 ↗