journal article·2006-06

Osteoblast response to plasma‐spray porous Ti6Al4V coating on substrates of identical alloy

L. Saldaña, J.L. González‐Carrasco, M. Rodríguez, L. Munuera, N. Vilaboa

Publisher page

Abstract

Abstract We have evaluated the in‐vitro biocompatibility of Ti6Al4V alloy coated by plasma spraying with an identical alloy. These surfaces are widely used in cementless prosthetic components, although osteoblasts behavior on this treated alloy has not been evaluated to date. Cross sectional examination revealed a thick and rough coating of identical composition without sign of delamination. Within the coating, small discontinuities and nonconnected pores were observed. Osteoblast response was evaluated by assessing cell adhesion, proliferation, and differentiation of primary cultures of human osteoblastic cells. Compared to the polished alloy, osteoblast adhesion measured as cell attachment and actin network reorganization was delayed on the plasma‐sprayed surface. Cell proliferation and viability were also impaired on the rough surface. Several informative markers of osteoblastic differentiation such as procollagen I peptide, alkaline phosphatase, osteocalcin, osteoprotegerin, and mineralized nodule formation were evaluated and indicated that the plasma‐sprayed alloy favored a more differentiated phenotype than polished alloy. Taken together, our in vitro results indicate that successful osseointegration of plasma spraying of Ti6Al4V with an identical alloy is mediated by modulation of osteoblastic differentiation and mineralization. © 2006 Wiley Periodicals, Inc. J Biomed Mater Res, 2006
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Bone Tissue Engineering Materials · Orthopaedic implants and arthroplasty · Titanium Alloys Microstructure and Properties

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (44)

Show source-provided references
1. Chemical composition and morphology of titanium surface oxides10.1557/PROC-55-351 ↗
2. 10.1016/0142-9612(96)88711-410.1016/0142-9612(96)88711-4 ↗
3. 10.1002/(SICI)1097-4636(19970305)34:3<279::AID-JBM2>3.0.CO;2-H10.1002/(SICI)1097-4636(19970305)34:3<279::AID-JBM2>3.0.CO;2-H ↗
4. 10.1115/1.279804210.1115/1.2798042 ↗
5. 10.1002/jbm.1006310.1002/jbm.10063 ↗
6. 10.1002/jbm.82029031410.1002/jbm.820290314 ↗
7. 10.1002/(SICI)1097-4636(199801)39:1<77::AID-JBM10>3.0.CO;2-L10.1002/(SICI)1097-4636(199801)39:1<77::AID-JBM10>3.0.CO;2-L ↗
8. 10.1016/S0890-6955(97)00114-410.1016/S0890-6955(97)00114-4 ↗
9. 10.1002/jbm.82025070810.1002/jbm.820250708 ↗
10. 10.1034/j.1600-0501.1997.080601.x10.1034/j.1600-0501.1997.080601.x ↗
11. 10.1016/S1350-4533(01)00050-910.1016/S1350-4533(01)00050-9 ↗
12. 10.1023/A:101283790591010.1023/A:1012837905910 ↗
13. 10.1023/A:101856052862510.1023/A:1018560528625 ↗
14. 10.1016/S0142-9612(01)00271-X10.1016/S0142-9612(01)00271-X ↗
15. 10.4028/www.scientific.net/KEM.198-199.17910.4028/www.scientific.net/KEM.198-199.179 ↗
16. Powder metal‐made orthopedic implants with porous surface for fixation by tissue ingrowth10.1097/00003086-198306000-00007 ↗
17. The optimum pore size for the fixation of porous‐surface metal implants by the ingrowth of bone10.1097/00003086-198007000-00045 ↗
18. 10.1016/S0883-5403(87)80024-410.1016/S0883-5403(87)80024-4 ↗
19. 10.1002/jbm.82027051110.1002/jbm.820270511 ↗
20. 10.1016/S0267-6605(86)80062-210.1016/S0267-6605(86)80062-2 ↗
21. 10.2106/00004623-198668040-0000910.2106/00004623-198668040-00009 ↗
22. 10.1034/j.1600-0501.1998.090407.x10.1034/j.1600-0501.1998.090407.x ↗
23. 10.1007/BF0322256010.1007/BF03222560 ↗
24. 10.2106/00004623-199807000-0000310.2106/00004623-199807000-00003 ↗
25. 10.1302/0301-620X.79B6.748210.1302/0301-620X.79B6.7482 ↗
26. 10.1097/00003086-199905000-0002210.1097/00003086-199905000-00022 ↗
27. Comparison of total hip arthroplasty performed with and without cement: A randomized trial10.2106/00004623-200210000-00013 ↗
28. 10.1097/00003086-200112000-0001310.1097/00003086-200112000-00013 ↗
29. 10.1016/j.arth.2004.01.00610.1016/j.arth.2004.01.006 ↗
30. Brochure of Industrias Quirúrgicas de Levante SA Recubrimiento poroso de Ti6Al4V por Plasma Spray Tecnologia de recubrimiento Poroso R.E. Luedemann.
31. 10.1359/jbmr.1997.12.5.77810.1359/jbmr.1997.12.5.778 ↗
32. 10.1002/jbm.a.3026410.1002/jbm.a.30264 ↗
33. Concentration dependent effect of titanium and aluminium ions released from thermally oxidized Ti6Al4V alloy on human osteoblasts
34. 10.1016/S0142-9612(00)00042-910.1016/S0142-9612(00)00042-9 ↗
35. 10.1016/S0142-9612(98)00144-610.1016/S0142-9612(98)00144-6 ↗
36. 10.1359/jbmr.2000.15.6.116910.1359/jbmr.2000.15.6.1169 ↗
37. 10.1016/j.biomaterials.2004.09.00310.1016/j.biomaterials.2004.09.003 ↗
38. 10.1002/jbm.1029010.1002/jbm.10290 ↗
39. 10.1002/(SICI)1097-4636(19980905)41:3<489::AID-JBM20>3.0.CO;2-C10.1002/(SICI)1097-4636(19980905)41:3<489::AID-JBM20>3.0.CO;2-C ↗
40. 10.1002/1097-4636(20010905)56:3<417::AID-JBM1111>3.0.CO;2-K10.1002/1097-4636(20010905)56:3<417::AID-JBM1111>3.0.CO;2-K ↗
41. 10.1177/15440591030820090710.1177/154405910308200907 ↗
42. Development of the osteoblast phenotype: Molecular mechanisms mediating osteoblast growth and differentiation
43. 10.1007/s00223-001-1114-y10.1007/s00223-001-1114-y ↗
44. 10.1002/jbm.a.3002510.1002/jbm.a.30025 ↗