journal article·2014-08

Biodegradable polymeric coatings on cochlear implant surfaces and their influence on spiral ganglion cell survival

P. Ceschi, A. Bohl, K. Sternberg, A. Neumeister, V. Senz, K.P. Schmitz, M. Kietzmann, V. Scheper, T. Lenarz, T. Stöver, G. Paasche

Publisher page

Abstract

Abstract To improve the electrode–nerve interface of cochlear implants (CI), the role of poly(L‐lactide) (PLLA) and poly(4‐hydroxybutyrate) (P(4HB)) as potential coating matrices for CI was assessed both in vitro and in vivo in terms of degradation behavior and effects on spiral ganglion neurons, the main target of the electrical stimulation with a CI. Growth rates of fibroblasts on the polymers were investigated and a direct‐contact test with freshly isolated spiral ganglion cells (SGC) was performed. In addition, the effects of the polymer degradation inside the inner ear were evaluated in vivo. The polymer degradation was assessed by use of scanning electron microscopy in combination with an energy‐dispersive X‐ray analysis. In vitro, no influence of the polymers was detected on fibroblasts' viability and on SGC survival rate. In vivo, SGC density was decreased only 6 months after implantation in the basal and middle turns of the cochlea in comparison to normal‐hearing animals but not between implanted groups (coated or uncoated). The analysis of the electrode models showed that in vivo P(4HB) is characterized by a gradual degradation completed after 6 months; whereas, the PLLA coatings burst along their longitudinal axis but showed only little degradation within the same time frame. In conclusion, both polymers seem to justify further evaluation as possible coating for CI electrodes. Of the two options, due to its excellent coating adhesion/stability and optimal degradation behavior, P(4HB) may prove to be the more promising biodegradable polymer for designing a drug delivery system from the surface of CI electrodes. © 2014 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 102B: 1255–1267, 2014.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Hearing Loss and Rehabilitation · Conducting polymers and applications · Hearing, Cochlea, Tinnitus, Genetics

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Hannover Medical School Department of Otolaryngology Hannover Germany

School of Veterinary Medicine Hannover Foundation Department of Pharmacology Toxicology and Pharmacy Hannover Germany

University of Rostock, Institute for Biomedical Engineering Rostock Germany

Laser Zentrum Hannover e.V. Hannover Germany

KGU Department of Otolaryngology Frankfurt Germany

Funding

Deutsche Forschungsgemeinschaft

References (54)

Show source-provided references
1. 10.1016/S0378-5955(96)00193-110.1016/S0378-5955(96)00193-1 ↗
2. 10.1159/00010320810.1159/000103208 ↗
3. 10.1177/00034894981070110210.1177/000348949810701102 ↗
4. PfingstBE ColesaDJ SuGL TayehN.Effects of deafening and implantation on electrical detection thresholds. ARO Abstract #899 2004.
5. 10.1159/00011178010.1159/000111780 ↗
6. 10.1097/MLG.0b013e31818173ec10.1097/MLG.0b013e31818173ec ↗
7. 10.1097/01.aud.0000261655.30652.6210.1097/01.aud.0000261655.30652.62 ↗
8. 10.1097/MAO.0b013e3181ab8fba10.1097/MAO.0b013e3181ab8fba ↗
9. New developments in cochlear implants
10. 10.1002/jbm.b.3108410.1002/jbm.b.31084 ↗
11. 10.1517/17425247.5.10.105910.1517/17425247.5.10.1059 ↗
12. 10.1159/00024189710.1159/000241897 ↗
13. 10.1097/MLG.0b013e31815f8e4110.1097/MLG.0b013e31815f8e41 ↗
14. 10.1159/00024189610.1159/000241896 ↗
15. 10.1159/00024189210.1159/000241892 ↗
16. 10.1016/S0378-5955(99)00193-810.1016/S0378-5955(99)00193-8 ↗
17. 10.1097/00129492-200303000-0001610.1097/00129492-200303000-00016 ↗
18. 10.1016/j.otohns.2006.11.00610.1016/j.otohns.2006.11.006 ↗
19. Electrode features for hearing preservation and drug delivery strategies
20. 10.1002/jbm.b.3300410.1002/jbm.b.33004 ↗
21. 10.1179/1754762811Y.000000002410.1179/1754762811Y.0000000024 ↗
22. 10.1111/jphp.1208610.1111/jphp.12086 ↗
23. 10.1111/j.1540-8183.1995.tb00565.x10.1111/j.1540-8183.1995.tb00565.x ↗
24. 10.1016/j.jcin.2011.11.01310.1016/j.jcin.2011.11.013 ↗
25. 10.1093/eurheartj/ehs08610.1093/eurheartj/ehs086 ↗
26. 10.1586/erc.12.2910.1586/erc.12.29 ↗
27. 10.1583/1545-1550(2007)14[725:ABSBOP]2.0.CO;210.1583/1545-1550(2007)14[725:ABSBOP]2.0.CO;2 ↗
28. 10.1016/S0142-9612(03)00161-310.1016/S0142-9612(03)00161-3 ↗
29. 10.1093/bmb/59.1.22710.1093/bmb/59.1.227 ↗
30. 10.1007/s10856-007-0148-810.1007/s10856-007-0148-8 ↗
31. 10.1007/s10439-007-9376-910.1007/s10439-007-9376-9 ↗
32. 10.1002/jbm.b.3182610.1002/jbm.b.31826 ↗
33. 10.1002/jbm.a.3166610.1002/jbm.a.31666 ↗
34. 10.1007/s10856-012-4698-z10.1007/s10856-012-4698-z ↗
35. 10.1016/0378-5955(88)90111-610.1016/0378-5955(88)90111-6 ↗
36. 10.1097/00001756-200512190-0000810.1097/00001756-200512190-00008 ↗
37. 10.1152/jn.01112.200510.1152/jn.01112.2005 ↗
38. 10.1115/1.183534910.1115/1.1835349 ↗
39. 10.1159/00024189210.1159/000241892 ↗
40. 10.1016/j.biomaterials.2008.04.02310.1016/j.biomaterials.2008.04.023 ↗
41. 10.1016/S1071-0949(03)90012-610.1016/S1071-0949(03)90012-6 ↗
42. 10.1016/j.jacc.2011.05.05010.1016/j.jacc.2011.05.050 ↗
43. 10.1016/S0142-9612(03)00451-410.1016/S0142-9612(03)00451-4 ↗
44. 10.1113/jphysiol.2006.11288810.1113/jphysiol.2006.112888 ↗
45. 10.1016/S1369-703X(03)00040-810.1016/S1369-703X(03)00040-8 ↗
46. 10.1159/00016591510.1159/000165915 ↗
47. 10.1016/j.expneurol.2008.08.00210.1016/j.expneurol.2008.08.002 ↗
48. 10.1002/cii.1710.1002/cii.17 ↗
49. 10.1159/00009561410.1159/000095614 ↗
50. 10.1007/s10162-011-0283-210.1007/s10162-011-0283-2 ↗
51. 10.1016/j.neuroscience.2012.03.03310.1016/j.neuroscience.2012.03.033 ↗
52. Some morphological features of neurons in the rat spinal ganglion10.1007/BF00453378 ↗
53. 10.1016/j.nbd.2007.07.02610.1016/j.nbd.2007.07.026 ↗
54. 10.1097/01.mao.0000179527.17285.8510.1097/01.mao.0000179527.17285.85 ↗