Abstract

Abstract Objective: The objective of this study was to compare the biomechanical strength of two barbed suture devices: V‐Loc™ 180 Wound Closure Device and Quill™ PDO Knotless Tissue‐Closure Device following primary cosmetic skin closures in a porcine dermal model. Method: This prospective randomized, controlled in vivo trial compared size 3/0 V‐Loc™ 180 device to size 2/0 Quill™ PDO device. Both products were tested for dermal closure in adult porcine models and evaluated at five timepoints. At postoperative days 0, 3, 7, 14, and 28 sutured tissue regions were excised post mortem and tested for intradermal wound holding strength. Results: Wounds closed with V‐Loc™ 180 device were stronger than Quill™ PDO device at days 0, 3, 7, and 14 with these differences being significant (p < 0.05) at days 3 and 7. At day 3, the average maximum load of V‐Loc™ 180 was 13.53 kgf and Quill™ PDO was 10.38 kgf (p = 0.002). At day 7, the average maximum load of V‐Loc™ 180 was 10.4 kgf and Quill™ PDO was 7.56 kgf (p = 0.001). Throughout the duration of the study, there was no suture extrusion or tissue distortion and all wounds healed with no major complications. Conclusions: In this study, V‐Loc™ 180 device was significantly stronger than Quill™ PDO device during the critical phases of wound healing in skin. © 2012 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2012.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Surgical Sutures and Adhesives · Wound Healing and Treatments · Reconstructive Facial Surgery Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (19)

Show source-provided references
1. AlcamoJH. Surgical Suture. U.S. Patent 3 123 077 (1964).
2. 10.1007/s10856-007-3114-610.1007/s10856-007-3114-6 ↗
3. 10.1111/j.1445-2197.2007.04394.x10.1111/j.1445-2197.2007.04394.x ↗
4. 10.1016/j.jcws.2009.01.00210.1016/j.jcws.2009.01.002 ↗
5. 10.1016/j.asj.2006.08.01110.1016/j.asj.2006.08.011 ↗
6. 10.1097/01.prs.0000299452.24743.6510.1097/01.prs.0000299452.24743.65 ↗
7. 10.1097/PRS.0b013e318177398210.1097/PRS.0b013e3181773982 ↗
8. 10.1097/PRS.0b013e3181cb64f710.1097/PRS.0b013e3181cb64f7 ↗
9. 10.1097/01.prs.0000209971.08264.b010.1097/01.prs.0000209971.08264.b0 ↗
10. 10.1177/1090820X1039501010.1177/1090820X10395010 ↗
11. 1996
12. Quill™ SRS Techniques and Procedures A Novel Approach to Soft Tissue Approximation.2008:8.
13. V‐Loc™ 180 Absorbable Wound Closure Device Step by Step Guide to Wound Closure Applications.
14. LeungJC GLRuff MegaroMA. Barbed bi‐directional medical sutures: Biomechanical properties and wound closure efficacy study. Society for Biomaterials 28th Meeting Transactions; Tampa FL: Paril 24–27 2002. p724.
15. RodeheaverGT Pineros‐FernandezA SalopekLS Rodeheaver PA Leung JC Ruff GL Batchelor SD. Barbed sutures for wound closure: In vivo wound society tissue compatibility and cosmesis measurements. Society for Biomaterials 30th Annual Meeting Transactions; Memphis TN: April 27–30 2005. p232.
16. 10.1001/archderm.143.7.86910.1001/archderm.143.7.869 ↗
17. LeungJC RuffGL BarchelorSD. Performance enhancements of a knotless suture via barb geometry modifications. 7th World Biomaterials Congress; Sydney AUS: May 17–21 2004. p1587.
18. 10.1016/j.jbiomech.2009.08.03310.1016/j.jbiomech.2009.08.033 ↗
19. 10.1016/j.jbiomech.2009.11.01210.1016/j.jbiomech.2009.11.012 ↗