The Effect of an Implanted Filter on Valsalva-Compression and Respiratory-Compression of the Inferior Vena Cava
Source abstract
Abstract
We carry out computational simulations of the Valsalva compression of an inferior vena cava (IVC) with a filter implanted in it. We find that when we treat the IVC wall as simply a boundary between 2 fluids and apply an external pressure on the IVC, the deformation magnitudes and patterns do not agree with data in the literature. We conclude that IVC compression is mainly caused by solid bodies (i.e. tissue and organs) compressing the IVC, and develop a model to simulate this phenomenon. We calibrate our model to data in the literature for Valsalva compression with and without an implanted filter. We then use our approach to predict the area reduction of the IVC during breathing when 2 different types of filters are implanted. Not surprisingly, we find that a compliant filter is less able to resist the compression of the IVC during respiration than a stiffer one, and we quantify the difference. We anticipate that, with further development, our model can be used to make assessments of design and testing parameters that can help to avoid fatigue failure of a filter that is subjected to millions of compressions due to breathing.
Source: expertdata · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Cardiac Arrhythmias and Treatments · Cardiovascular Function and Risk Factors · Cardiovascular Syncope and Autonomic Disorders
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (55)
Show source-provided references
1. Netter, F.H.: Atlas of Human Anatomy. Elsevier, Amsterdam (2019)
2. Greenfield, L.J., Zocco, J., Wilk, J., Schroeder, T.M., Elkins, R.C.: Clinical experience with the Kim-Ray Greenfield vena caval filter. Ann. Surg. 185, 692–698 (1977). https://doi.org/10.1097/00000658-197706000-0001210.1097/00000658-197706000-00012 ↗
3. Grassi, C.J.: Inferior vena caval filters: analysis of five currently available devices. Am. J. Roentgenol. 156, 813–821 (1991)10.2214/ajr.156.4.2003449 ↗
4. Neuerburg, J., Günther, R.W., Rassmussen, E., Vorwerk, D., Tonn, K., Handt, S., Küpper, W., Hansen, J.V.: New retrievable percutaneous vena cava filter: experimental in vitro and in vivo evaluation. Cardiovasc. Interv. Radiol. 16, 224–229 (1993). https://doi.org/10.1007/BF0260296510.1007/BF02602965 ↗
5. Deso, S.E., Idakoji, I.A., Kuo, W.T.: Evidence-based evaluation of inferior vena cava filter complications based on filter type. Semin. Intervent. Radiol. 33, 93–100 (2016). https://doi.org/10.1055/s-0036-158320810.1055/s-0036-1583208 ↗
6. Grant, E., Rendano, F., Sevinc, E., Gammelgaard, J., Holm, H., Grovall, S.: Normal inferior vena cava: Caliber changes observed by dynamic ultrasound. Am. J. Roentgenol. 135, 335–338 (1980)10.2214/ajr.135.2.335 ↗
7. Murphy, E.H., Johnson, E.D., Arko, F.R.: Evaluation of wall motion and dynamic geometry of the inferior vena cava using intravascular ultrasound: implications for future device design. J. Endovasc. Ther. 15, 349–355 (2008). https://doi.org/10.1583/08-2424.110.1583/08-2424.1 ↗
8. Klabunde, R.E.: Cardiovascular Physiology Concepts, 2nd edn. LWW, Philadelphia (2011)
9. Tibbs, D.J.: Varicose Veins and Related Disorders. Butterworth-Heinemann, Stoneham (1991)
10. Uva, B., Aliverti, A., Bovio, D., Kayser, B.: The “abdominal circulatory pump”: an auxiliary heart during exercise? Front. Physiol. 6, 1–12 (2016). https://doi.org/10.3389/fphys.2015.0041110.3389/fphys.2015.00411 ↗
11. LoMauro, A., Aliverti, A.: Blood shift during cough: negligible or significant? Front. Physiol. 9, 1–7 (2018). https://doi.org/10.3389/fphys.2018.0050110.3389/fphys.2018.00501 ↗
12. De Gregorio, M.A., Gamboa, P., Bonilla, D.L., Sanchez, M., Higuera, M.T., Medrano, J., Mainar, A., Lostalé, F., Laborda, A.: Retrieval of Günther Tulip optional vena cava filters 30 days after implantation: a prospective clinical study. J. Vasc. Interv. Radiol. 17, 1781–1789 (2006). https://doi.org/10.1097/01.RVI.0000244837.46324.7210.1097/01.RVI.0000244837.46324.72 ↗
13. Blehar, D.J., Resop, D., Chin, B., Dayno, M., Gaspari, R.: Inferior vena cava displacement during respirophasic ultrasound imaging. Crit. Ultrasound J. 4, 1–5 (2012). https://doi.org/10.1186/2036-7902-4-1810.1186/2036-7902-4-18 ↗
14. Pstras, L., Thomaseth, K., Waniewski, J., Balzani, I., Bellavere, F.: The Valsalva manoeuvre: physiology and clinical examples. Acta Physiol. 217, 103–119 (2016). https://doi.org/10.1111/apha.1263910.1111/apha.12639 ↗
15. Laborda, A., Sierra, S., Malve, M., De Blas, I., Ioakeim, I., Kuo, W.T., De Gregorio, M.A.: Influence of breathing movements and Valsalva maneuver on vena caval dynamics. World J. Radiol. 6, 833–839 (2014). https://doi.org/10.4329/wjr.v6.i10.83310.4329/wjr.v6.i10.833 ↗
16. Laborda, A., Kuo, W.T., Ioakeim, I., De Blas, I., Malvè, M., Lahuerta, C., De Gregorio, M.A.: Respiratory-induced haemodynamic changes: a contributing factor to IVC filter penetration. Cardiovasc. Interv. Radiol. 38, 1192–1197 (2015). https://doi.org/10.1007/s00270-015-1077-410.1007/s00270-015-1077-4 ↗
17. Joseph, A.A., Voit, D., Frahm, J.: Inferior vena cava revisited – real-time flow MRI of respiratory maneuvers. NMR Biomed. 33, 1–8 (2020). https://doi.org/10.1002/nbm.423210.1002/nbm.4232 ↗
18. Kalva, S.P., Wicky, S., Waltman, A.C., Athanasoulis, C.A.: TrapEase vena cava filter: experience in 751 patients. J. Endovasc. Ther. 13, 365–372 (2006). https://doi.org/10.1583/05-1741.110.1583/05-1741.1 ↗
19. Johnson, M.S., Nemcek, A.A. Jr., Benenati, J.F., Baumann, D.S., Dolmatch, B.L., Kaufman, J.A., Garcia, M.J., Stecker, M.S., Venbrux, A.C., Haskal, Z.J., Avelar, R.L.: The safety and effectiveness of the retrievable Option inferior vena cava filter: a United States prospective multicenter clinical study. J. Vasc. Interv. Radiol. 21, 1173–1184 (2010). https://doi.org/10.1016/j.jvir.2010.04.00410.1016/j.jvir.2010.04.004 ↗
20. Stavropoulos, S.W., Sing, R.F., Elmasri, F., Silver, M.J., Powell, A., Lynch, F.C., Aal, A.K.A., Lansky, A.J., Settlage, R.A., Muhs, B.E.: The Denali trial: an interim analysis of a prospective, multicenter study of the Denali retrievable inferior vena cava filter. J. Vasc. Interv. Radiol. 25, 1497–1505.e1 (2014). https://doi.org/10.1016/j.jvir.2014.07.00110.1016/j.jvir.2014.07.001 ↗
21. Smouse, H.B., Rosenthal, D., Van Ha, T., Knox, M.F., Dixon, R.G., Voorhees, W.D., McCann-Brown, J.A.: Long-term retrieval success rate profile for the Günther Tulip vena cava filter. J. Vasc. Interv. Radiol. 20, 871–877 (2009). https://doi.org/10.1016/j.jvir.2009.03.03310.1016/j.jvir.2009.03.033 ↗
22. Le Blanche, A.F., Benazzouz, A., Reynaud, P., Fernandez, R.O., Emanuelli, G., Ricco, J.B., Delcour, C.: The VenaTech LP permanent caval filter: effectiveness and safety in the prevention of pulmonary embolism-a European multicenter study. J. Vasc. Interv. Radiol. 19, 509–515 (2008). https://doi.org/10.1016/j.jvir.2007.11.02410.1016/j.jvir.2007.11.024 ↗
23. Lyon, S.M., Riojas, G.E., Uberoi, R., Patel, J., Lipp, M.E.B., Plant, G.R., De Gregorio, M.A., Günther, R.W., Voorhees, W.D., McCann-Brown, J.A.: Short- and long-term retrievability of the Celect vena cava filter: results from a multi-institutional registry. J. Vasc. Interv. Radiol. 20, 1441–1448 (2009). https://doi.org/10.1016/j.jvir.2009.07.03810.1016/j.jvir.2009.07.038 ↗
24. Roehm, J.O.F.: The bird’s nest filter: a new percutaneous transcatheter inferior vena cava filter. J. Vasc. Surg. 1, 498–501 (1984). https://doi.org/10.1016/0741-5214(84)90092-210.1016/0741-5214(84)90092-2 ↗
25. Robertson, S.W., Pelton, A.R., Ritchie, R.O.: Mechanical fatigue and fracture of Nitinol. Int. Mater. Rev. 57, 1–36 (2012). https://doi.org/10.1179/1743280411Y.000000000910.1179/1743280411Y.0000000009 ↗
26. Schmidt, P.A., Blair, R.R., Earthman, J.C.: Development of a novel specimen geometry for fatigue testing of fine wire. J. Test. Eval. 23, 73–79 (1995). https://doi.org/10.1520/jte10897j10.1520/jte10897j ↗
27. Kobayashi, E., Mochizuki, H., Doi, H., Yoneyama, T., Hanawa, T.: Fatigue life prediction of biomedical titanium alloys under tensile/torsional stress. Mater. Trans. 47, 1826–1831 (2006). https://doi.org/10.2320/matertrans.47.182610.2320/matertrans.47.1826 ↗
28. Huang, J.Y., Yen, J.J., Jeng, S.L., Chen, C.Y., Kuo, R.C.: High-cycle fatigue behavior of type 316L stainless steel. Mater. Trans. 47, 409–417 (2006). https://doi.org/10.2320/matertrans.47.40910.2320/matertrans.47.409 ↗
29. Hull, J.E., Robertson, S.W.: Bard Recovery filter: evaluation and management of vena cava limb perforation, fracture, and migration. J. Vasc. Interv. Radiol. 20, 52–60 (2009). https://doi.org/10.1016/j.jvir.2008.09.03210.1016/j.jvir.2008.09.032 ↗
30. Kuo, W.T., Robertson, S.W., Odegaard, J.I., Hofmann, L.V.: Complex retrieval of fractured, embedded, and penetrating inferior vena cava filters: a prospective study with histologic and electron microscopic analysis. J. Vasc. Interv. Radiol. 24, 622–630.e1 (2013). https://doi.org/10.1016/j.jvir.2013.01.00810.1016/j.jvir.2013.01.008 ↗
31. Birring, S.S., Matos, S., Patel, R.B., Prudon, B., Evans, D.H., Pavord, I.D.: Cough frequency, cough sensitivity and health status in patients with chronic cough. Respir. Med. 100, 1105–1109 (2006). https://doi.org/10.1016/j.rmed.2005.09.02310.1016/j.rmed.2005.09.023 ↗
32. McGrath, D.J., Thiebes, A.L., Cornelissen, C.G., O’Brien, B., Jockenhoevel, S., Bruzzi, M., McHugh, P.E.: Evaluating the interaction of a tracheobronchial stent in an ovine in-vivo model. Biomech. Model. Mechanobiol. 17, 499–516 (2018). https://doi.org/10.1007/s10237-017-0974-710.1007/s10237-017-0974-7 ↗
33. Special 510(k): Device Modification, K121057, Günther Tulip Vena Cava Filter Sets and Cook Celect Vena Cava Filter Sets, Cook Medical–William Cook Europe ApS (2012)
34. Nicolás, M., Lucea, B., Laborda, A., Peña, E., De Gregorio, M.A., Martínez, M.A., Malvè, M.: Influence of a commercial antithrombotic filter on the caval blood flow during neutra and Valsalva maneuver. J. Med. Devices, Trans. ASME. 11, 1–11 (2017). https://doi.org/10.1115/1.403598310.1115/1.4035983 ↗
35. Aycock, K.I., Campbell, R.L., Manning, K.B., Sastry, S.P., Shontz, S.M., Lynch, F.C., Craven, B.A.: A computational method for predicting inferior vena cava filter performance on a patient-specific basis. J. Biomech. Eng. 136, 1–14 (2014). https://doi.org/10.1115/1.402761210.1115/1.4027612 ↗
36. Pérez-Andrés, A., Peña, E.: Calibration of interface properties and application to a finite element model for predicting vena cava filter–induced vein wall failure. Int. J. Numer. Methods Biomed. Eng. 34, 1–13 (2018). https://doi.org/10.1002/cnm.309810.1002/cnm.3098 ↗
37. Ogden, R.W.: Large deformation isotropic elasticity – on the correlation of theory and experiment for incompressible rubberlike solids. Proc. R. Soc. Lond. Ser. A, Math. Phys. Sci. 326, 565–584 (1972). https://doi.org/10.1098/rspa.1972.002610.1098/rspa.1972.0026 ↗
38. Dassault Systèmes: Abaqus version 2019, Providence RI (2019)
39. Anonymous: Cobalt Monograph. Brussels: Centre d’information du cobalt (1960)
40. García, A., Lerga, S., Peña, E., Malve, M., Laborda, A., De Gregorio, M.A., Martínez, M.A.: Evaluation of migration forces of a retrievable filter: experimental setup and finite element study. Med. Eng. Phys. 34, 1167–1176 (2012). https://doi.org/10.1016/j.medengphy.2011.12.00510.1016/j.medengphy.2011.12.005 ↗
41. Kapila, S., Sachdeva, R.: Mechanical properties and clinical applications of orthodontic wires. Am. J. Orthod. Dentofac. Orthop. 96, 100–109 (1989). https://doi.org/10.1016/0889-5406(89)90251-510.1016/0889-5406(89)90251-5 ↗
42. Assefpour-Dezfuly, M., Bonfield, W.: Microplasticity in Elgiloy. J. Mater. Sci. 20, 3018–3026 (1985). https://doi.org/10.1007/BF0055306810.1007/BF00553068 ↗
43. Dowell, J.D., Castle, J.C., Schickel, M., Andersson, U.K., Zielinski, R., McLoney, E., Guy, G., Yang, X., Ghadiali, S.: Celect inferior vena cava wall strut perforation begets additional strut perforation. J. Vasc. Interv. Radiol. 26, 1510–1518.e3 (2015). https://doi.org/10.1016/j.jvir.2015.06.02010.1016/j.jvir.2015.06.020 ↗
44. Aycock, K.I., Campbell, R.L., Lynch, F.C., Manning, K.B., Craven, B.A.: The importance of hemorheology and patient anatomy on the hemodynamics in the inferior vena cava. Ann. Biomed. Eng. 44, 3568–3582 (2016). https://doi.org/10.1007/s10439-016-1663-x10.1007/s10439-016-1663-x ↗
45. Dunn, A.C., Zaveri, T.D., Keselowsky, B.G., Sawyer, W.G.: Macroscopic friction coefficient measurements on living endothelial cells. Tribol. Lett. 27, 233–238 (2007). https://doi.org/10.1007/s11249-007-9230-010.1007/s11249-007-9230-0 ↗
46. Mummert, J., Sirois, E., Sun, W.: Quantification of biomechanical interaction of transcatheter aortic valve stent deployed in porcine and ovine hearts. Ann. Biomed. Eng. 41, 577–586 (2013). https://doi.org/10.1007/s10439-012-0694-110.1007/s10439-012-0694-1 ↗
47. McGee, O.M., Sun, W., McNamara, L.M.: An in vitro model quantifying the effect of calcification on the tissue–stent interaction in a stenosed aortic root. J. Biomech. 82, 109–115 (2019). https://doi.org/10.1016/j.jbiomech.2018.10.01010.1016/j.jbiomech.2018.10.010 ↗
48. Fung, Y.C.: Biomechanics. Springer, Berlin (1993)10.1007/978-1-4757-2257-4 ↗
49. Shi, Y., Glaser, K.J., Venkatesh, S.K., Ben-Abraham, E.I., Ehman, R.L.: Feasibility of using 3D MR elastography to determine pancreatic stiffness in healthy volunteers. J. Magn. Reson. Imaging 41, 369–375 (2015). https://doi.org/10.1002/jmri.2457210.1002/jmri.24572 ↗
50. Patel, P.N., Smith, C.K., Patrick, C.W.: Rheological and recovery properties of poly(ethylene glycol) diacrylate hydrogels and human adipose tissue. J. Biomed. Mater. Res., Part A 73, 313–319 (2005). https://doi.org/10.1002/jbm.a.3029110.1002/jbm.a.30291 ↗
51. Ogneva, I.V., Ushakov, I.B.: The transversal stiffness of skeletal muscle fibers and cardiomyocytes in control and after simulated microgravity. In: Frewin, C. (ed.) At. Force Microsc. Investig. into Biol. – From Cell to Protein, pp. 325–354 (2012). https://doi.org/10.5772/3645710.5772/36457 ↗
52. Auricchio, F., Taylor, R.L.: Shape-memory alloys: modelling and numerical simulations of the finite-strain superelastic behavior. Comput. Methods Appl. Mech. Eng. 143, 175–194 (1997)10.1016/S0045-7825(96)01147-4 ↗
53. Auricchio, F., Taylor, R.L., Lubliner, J.: Shape-memory alloys: macromodelling and numerical simulations of the superelastic behavior. Comput. Methods Appl. Mech. Eng. 146, 281–312 (1997)10.1016/S0045-7825(96)01232-7 ↗
54. Cao, H., Wu, M.H., Zhou, F., McMeeking, R.M., Ritchie, R.O.: The influence of mean strain on the high-cycle fatigue of Nitinol with application to medical devices. J. Mech. Phys. Solids 143, 104057 (2020). https://doi.org/10.1016/j.jmps.2020.10405710.1016/j.jmps.2020.104057 ↗
55. Kossa, A., McMeeking, R.M.: Bending of a Nitinol cantilever and its fatigue performance. Extrem. Mech. Lett. 42, 101083 (2021). https://doi.org/10.1016/j.eml.2020.10108310.1016/j.eml.2020.101083 ↗