Depth limitations for in vivo magnetic nanoparticle detection with a compact handheld device
Source abstract
Abstract
The increasing interest for local detection of magnetic nanoparticles (MNPs) during clinical interventions requires the development of suitable probes that unambiguously detect the MNPs at a depth of several centimeters in the body. The present study quantitatively evaluates the limitations of a conventional magnetometry method using a sinusoidal alternating field. This method is limited by the variability of the magnetic susceptibility of the surrounding diamagnetic tissue. Two different sensors are evaluated in a theoretical model of MNP detection in a tissue volume. For a coil that completely encloses the sample volume, the MNPs can be detected if the total mass contributing to the signal is larger than 4.1×10-7 times the tissue mass. For a handheld surface coil, intended to search for the MNPs in a larger tissue volume, an amount of 1 μg of iron oxide cannot be detected by sensors with a diameter larger than 15 mm. To detect a spot with MNPs at 5 cm depth in tissue, it should contain at least 325 μg iron oxide. Therefore, for high-sensitive clinical MNP detection in surgical interventions, techniques with increased specificity for the nonlinear magnetic properties of MNPs are indispensable.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Characterization and Applications of Magnetic Nanoparticles · Geomagnetism and Paleomagnetism Studies · Magnetic Field Sensors Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (12)
Show source-provided references
1. Sentinel node biopsy using a magnetic tracer versus standard technique
2. The central-european sentimag study10.1016/j.breast.2014.01.004 ↗
3. A novel method for sentinel lymph node mapping using magnetite in patients with non-small cell lung cancer10.1016/S0022-5223(03)00216-2 ↗
4. Intraoperative sentinel lymph node mapping using a new sterilizable magnetometer in patients with nonsmall cell lung cancer10.1016/j.athoracsur.2005.06.005 ↗
5. Subpleural injection of tracer improves detection of mediastinal sentinel lymph nodes in non-small cell lung cancer10.1016/j.ejcts.2007.07.012 ↗
6. Sentinel node mapping and micrometastasis in patients with clinical stage ia non-small cell lung cancer10.1510/icvts.2009.214197 ↗
7. Magnetic resonance lymphography of sentinel lymph nodes in patients with breast cancer using superparamagnetic iron oxide
8. Sentinel lymph node biopsy in patients with breast cancer using superparamagnetic iron oxide and a magnetometer10.1007/s12282-011-0327-9 ↗
9. Precise localization of sentinel lymph nodes and estimation of their depth using a prototype intraoperative mini γ-camera in patients with breast cancer10.2967/jnumed.106.036574 ↗
10. The role of magnetic susceptibility in magnetic resonance imaging10.1118/1.597854 ↗
11. Rotational diffusion in iron ferrofluids10.1021/la0346393 ↗
12. AC susceptibility of magnetic fluid in nonlinear brownian relaxation region10.1143/JJAP.49.053001 ↗