journal article·2024-11-11Open access

A Measurement Device for the Normal Spectral Emissivity of Materials Under Low-Temperature and Vacuum Conditions

Xufeng Liu, Zhenyuan Zhang, Guojin Feng

Applied Sciences· Volume 14· Issue 22

Abstract

Based on parallel light, a device for measuring sample emissivity under vacuum and low temperature was established. In this paper, a new emissivity measurement formula was designed, which replaces the derivation of Kirchhoff’s law of thermal radiation. The device is designed with a light-weight blackbody, an improved cooling speed, and an improved PID temperature control system to achieve a good temperature stability, close to 1 mK. The device is capable of measuring within a wavelength range of 4 μm to 16 μm. The results of the uncertainty assessment show that the uncertainty of the normal emissivity is better than 1.6% in the range of 4 μm to 7 μm, and better than 0.6% in the range of 7 μm to 14 μm (k = 1) at 0 °C.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Calibration and Measurement Techniques · Thermography and Photoacoustic Techniques · Advanced Sensor Technologies Research

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

National Institute of Metrology of China, Beijing 100029, China

College of Optical and Electronic Technology, China Jiliang University, Hangzhou 310018, China

Beijing Key Laboratory of Performance Guarantee on Urban Rail Transit Vehicles, Beijing University of Civil Engineering and Architecture, Beijing 100044, China

Funding

National Key R&D Projects

References (27)

Show source-provided references
1. Infrared spectral emissivity characteritation facility at NIST
2. Short-pulse Laser Heating on Metals10.1016/0017-9310(92)90131-B ↗
3. Wright, D., and Grego, L. (2005). Physics of Space Security, American Academy of Arts & Science.
4. Comparative study of radiometric and calorimetric methods for total hemispherical emissivity measurements10.1007/s00231-017-2238-6 ↗
5. Zhang, P., Li, Y., Li, Y., Chen, G., and Hua, W. (2022). A Proposition to Correct Infrared Emissivity Measurements for Curved Surface Targets Based on the Irradiation Reflection Method. Photonics, 9.10.3390/photonics9100739 ↗
6. Li, Y., Zhang, P., Chen, G., Li, Y., Hua, W., Li, Y., and Jiao, Z. (2022). Study on Method for Measuring Coating Emissivity by Applying Active Irradiation Based on Infrared Thermal Imager. Sensors, 22.10.3390/s22062392 ↗
7. Study on the Multi-wavelength Emissivity of GCr15 Steel and its Application on Temperature Measurement for Continuous Casting Billets10.1007/s10765-016-2142-1 ↗
8. Research on the Temperature and Emissivity Measurement of the Metallic Thermal Protection Blanket10.1007/s10765-008-0448-3 ↗
9. Survey of emissivity measurement by radiometric methods10.1364/AO.54.000669 ↗
10. Hanssen, L.M., Rink, M., and Mekhontsev, S.N. (2004, January 28–30). Status of NIST near infrared emittance measurement system. Proceedings of the 12th IEEE International Conference on Advanced Thermal Processing of Semiconductors, RTP 2004, Portland, OR, USA.
11. Kaplan, S.G., Hanssen, L.M., Early, E.A., and Nadal, M.E. (2000). Intercomparison of regular spectral transmittance and reflectance measurements with FTIR- and monochromator-based spectrophotometers. Optical Diagnostic Methods for Inorganic Materials II, Proceedings of the International Symposium on Optical Science and Technology, San Diego, CA, USA, 30 July–4 August 2000, SPIE.10.1117/12.403570 ↗
12. Hanssen, L.M., Mekhontsev, S.N., and Khromchenko, V.B. (2004). Infrared spectral emissivity characterization facility at NIST. Thermosense XXVI, Proceedings of the Defense and Security, Orlando, FL, USA, 12–16 April 2004, SPIE.10.1117/12.542224 ↗
13. Wan, Y.Z. (2008). Design of Material Emissivity Testing Scheme Based on Vacuum Conditions, Harbin Institute of Technology.
14. Direct measurement of total emissivities at cryogenic temperatures: Application to satellite coatings10.1016/j.cryogenics.2008.07.007 ↗
15. Yu, K., Liu, Y., Jia, G., and Shi, D. (2011). An Experimental Apparatus for Normal Spectral Emissivity Measurement, Henan Normal University.10.1117/12.907003 ↗
16. Emissivity Measurement Under Vacuum from to and from to at PTB10.1007/s10765-014-1745-7 ↗
17. Measurement of Normal Emissivity of Materials at Low Temperature10.3788/OPE.20162401.0059 ↗
18. Infrared spectral normal emittance/emissivity comparison10.1088/0026-1394/53/1A/03001 ↗
19. An Accurate Device for Apparent Emissivity Characterization in Controlled Atmospheric Conditions Up To 1423 K10.1109/TIM.2019.2944504 ↗
20. An Apparatus for the Directional Spectral Emissivity Measurement in the Near Infrared Band10.1007/s10765-021-02813-0 ↗
21. New Emissivity Measuring System with High Accuracy Under Controlled Environment Conditions10.1007/s10765-022-03129-3 ↗
22. Experimental setup for direct measurements of spectral directional emissivity from 0° to 80° angles at temperatures as low as 40 °C10.1016/j.ijthermalsci.2022.108045 ↗
23. Development of An Apparatus for the Directional Spectral Emissivity Measurement from 50 °C to 1000 °C10.1007/s10765-024-03393-5 ↗
24. Sanchez-Marin, F.J., Calixto-Carrera, S., and Villaseñor-Mora, C. (2009). Novel approach to assess the emissivity of the human skin. J. Biomed. Opt., 14.10.1117/1.3086612 ↗
25. Experimental Investigation of Infrared Spectral Emissivity of Copper at 300~1123 K
26. Linearity Study of a Spectral Emissivity Measurement Facility10.1007/s10765-008-0446-5 ↗
27. Nonlinear measurement of spectral responsivity of Fourier transform infrared detection system based on luminous flux multiplication method