Carlos Vera, Fernando Garnacho, Joni Klüss, Christian Mier, Fernando Álvarez, Kari Lahti, Abderrahim Khamlichi, Alf-Peter Elg, Armando Rodrigo Mor, Eduardo Arcones, Álvaro Camuñas, Pertti Pakonen, Javier Ortego, José Ramón Vidal, Miran Haider, Jorge Rovira, Pascual Simon, Antonio Squicciarini

Applied Sciences· Volume 13· Issue 14

Abstract

The insulation condition of HVDC grids consisting of cable systems, GIS, and converters should be monitored by partial discharge (PD) analysers using artificial intelligence (AI) tools for efficient insulation diagnosis. Although there are many experiences of PD monitoring solutions developed for the supervision of the insulation condition of HVAC grids using PD analysers, there are no standardised requirements for their qualification available yet. The international technical specification TS IEC 62478 provides general rules for PD measurements using electromagnetic methods but does not define performance requirements for qualification tests. HVDC and HVAC PD analysers must be tested by unambiguous test procedures. This paper compiles experiences of using PD analysers with HFCT sensors in HVAC grids (cable systems, GIS, and AIS) to define a qualification procedure for HVAC systems. This procedure is applicable to HVDC grids (cable systems, GIS, AIS, and converters) because the particularities related to the insulation behaviour under HVDC voltage are also considered. Representative PD sources are discussed in HVAC and HVDC positive and negative polarity. The PD pulse trend of representative insulation defects in HVDC cable systems is quite different from that of HVAC grids. Special attention should be paid to the acquisition of PD signals in HVDC grids since few pulses appear in solid insulations, mainly during voltage changes (polarity reversals or surges), but rarely in continuous operation with constant direct voltage. A synthetic PD simulator has been developed to reproduce trains of PD pulses or noise signals, similar to those that can appear in the power network. A set of three functionality tests has been developed for qualification of the diagnostic capabilities of PD analysers working up to 30 MHz addressed to HVDC or HVAC grids: (1) PD recognition test, (2) PD clustering test, and (3) PD location test. This qualification procedure has been validated by means of a round-robin test performed by five research institutes (RISE, FFII, TUDelft, TAU, and UPM) using commercial and in-development AI PD recognition and clustering tools to demonstrate its robustness and applicability. Applying this qualification procedure, two PD methods for electrical detection and prevention of insulation defects have been approved, one for HVAC and the other for HVDC grids.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: High voltage insulation and dielectric phenomena · Power Transformer Diagnostics and Insulation · Thermal Analysis in Power Transmission

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

ETSIDI, Departamento de Ingeniería Eléctrica, Universidad Politécnica de Madrid, Electrónica, Automática y Física Aplicada, 28006 Madrid, Spain

Fundación para el Fomento de la Innovación Industrial, FFII-LCOE, Eric Kandel Street 1, Getafe, 28906 Madrid, Spain

RISE Research Institutes of Sweden, 501 15 Borås, Sweden

Electrical Sustainable Energy Department, Delft University of Technology, Mekelweg 4, 2628 CD Delft, The Netherlands

Electrical Engineering Department, Faculty of Information Technology and Communication Sciences, Tampere University (TAU), Korkeakoulunkatu 3, 33720 Tampere, Finland

Instituto de Tecnología Eléctrica, Universitat Politècnica de València, Camino de Vera s/n, 46022 Valencia, Spain

AMPACIMON SLU, C. de la Antracita, 7, Nave 15, 28045 Madrid, Spain

Funding

Participating States · European Union’s Horizon 2020 research and innovation programme

References (24)

Show source-provided references
1. Investigation of electrical tree growth characteristics and partial discharge pattern analysis using deep neural network10.1016/j.epsr.2023.109287 ↗
2. Best practices for Partial Discharge Monitoring of HVDC Cable Systems and Qualification Tests
3. Application of HFCT and UHF sensors in on-line partial discharge measurements for insulation diagnosis of high voltage equipment10.3390/s150407360 ↗
4. (2016). High Voltage Test Techniques—Measurement of Partial Discharges by Electromagnetic and Acoustic Methods. TC 42 High-Voltage and high-Current Test Techniques. Standard No. IEC TS 62478:2016.
5. Location and characterization of partial discharge sites in shielded power cables10.1109/61.53090 ↗
6. Guarding MV cables on-line: With travelling wave based temperature monitoring, fault location, PD location and PD related remaining life aspects10.1109/TDEI.2016.005566 ↗
7. A clustering technique for partial discharge and noise sources identification in power cables by means of waveform parameters10.1109/TDEI.2015.005037 ↗
8. Signal analysis and feature generation for pattern identification of partial discharges in high-voltage equipment10.1016/j.epsr.2012.08.016 ↗
9. 19ENG02 FutureEnergy “Metrology for future energy transmission” EURAMET H2020 Project, EMPIR Program, 2020-2023.
10. Insulation Materials and Systems for Power Electronics Modules: A Review Identifying Challenges and Future Research Needs10.1109/TDEI.2020.009041 ↗
11. Localization of Electrical-Insulation and Partial-Discharge Failures of IGB Modules10.1109/28.980373 ↗
12. PD measurements, failure analysis, and control in high-power IGBT modules10.1049/hve.2017.0186 ↗
13. Power converters installed on board rolling stock—Part 1: Characteristics and test methods. Standard No. IEC 61287-1:2014.
14. Synchronous multi-channel PD measurements10.1109/TDEI.2008.4712676 ↗
15. Wang, Y., Chen, P., Zhao, Y., and Sun, Y. (2022). A Denoising Method for Mining Cable PD Signal Based on Genetic Algorithm Optimization of VMD and Wavelet Threshold. Sensors, 22.10.3390/s22239386 ↗
16. Automated wavelet selection and thresholding for PD detection10.1109/57.995398 ↗
17. Partial discharge signal denoising with spatially adaptive wavelet thresholding and support vector machines10.1016/j.epsr.2010.10.030 ↗
18. Wavelet packet denoising for online partial discharge detection in cables and its application to experimental field results10.1088/0957-0233/17/9/001 ↗
19. A new methodology for the identification of PD in electrical apparatus: Properties and applications10.1109/TDEI.2005.1430391 ↗
20. Partial discharge and noise separation by means of spectral-power clustering techniques10.1109/TDEI.2013.6571466 ↗
21. Requirements of Artificial Intelligence Platform addressed to Automatic Assessment of Insulation Condition of Indoor and Outdoor Installations through Partial Discharge Monitoring
22. (2022, November 20). Delft University of Technology, PDFlex. Available online: http://pdflex.ewi.tudelft.nl/.
23. New clustering techniques based on current peak value, charge and energy calculations for separation of partial discharge sources10.1109/TDEI.2016.006352 ↗
24. Álvarez, F., Ortego, J., Garnacho, F., and Sánchez-Urán, M.A. (2014, January 8–11). Advanced techniques for on-line PD measurements in high voltage systems. Proceedings of the 2014 ICHVE International Conference on High Voltage Engineering and Application, Poznan, Poland.10.1109/ICHVE.2014.7035404 ↗