journal article·2014-07Open access

Effect of Stent Radial Force on Stress Pattern After Deployment: A Finite Element Study

Alessandro Borghi, Olive Murphy, Reza Bahmanyar, Chris McLeod

Abstract

The present article presents a method for assessing the radial stiffness of nitinol stents. An idealized stent model was created, and its radial stiffness was calculated by means of finite element modeling. The calculations were validated against experimental measurements. The variation of radial stiffness with geometrical dimensions was calculated, and the effect of increasing radial stiffness on endovascular deployment was analyzed. Peak tensile and compressive stresses as well as stent penetration were calculated in the case of an idealized pulmonary artery model having realistic dimensions as well as stiffness. The results of stress calculations were compared with a second set of simulations, where an idealized behavior of the stent (uniform expansion to a theoretical contact diameter) was modeled. The results show how in reality nitinol stents behave in a non-ideal way, having a non-uniform expansion and exerting non-uniform pressure on the contact areas with the artery. Such non-ideality decreases though with the increase in radial stiffness. The radial force alone may be insufficient in describing the stent-artery interaction, and numerical modeling proves to be necessary for capturing such complexity.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Coronary Interventions and Diagnostics · Aortic aneurysm repair treatments · Vascular Procedures and Complications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (26)

Show source-provided references
1. T.W. Duerig, D.E. Tolomeo, and M. Wholey, An overview of Superelastic Stent design, Minim Invasive Ther Allied Technol, 2000, 9(3–4), p 235–24610.3109/13645700009169654 ↗
2. ISO 25539-2:2012, ISO 25539-2:2012, Cardiovascular implants—Endovascular devices—Part 2: Vascular stents. 2012
3. FDA, Guidance Documents (Medical Devices and Radiation-Emitting Products)—Guidance for Industry and FDA Staff—Non-Clinical Engineering Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems.
4. S.H. Cha, M.H. Han, Y.H. Choi, C.J. Yoon, S.K. Baik, S.J Kim, and K.H. Chang, Vascular Responses in Normal Canine Carotid Arteries: Comparison Between Various Self-Expanding Stents of the Same Unconstrained Size, Invest. Radiol., 2003, 38(2), p 95–10110.1097/00004424-200302000-00004 ↗
5. J.W. Freeman, P.B. Snowhill, and J.L. Nosher, A Link Between Stent Radial Forces and Vascular Wall Remodeling: The Discovery of an Optimal Stent Radial Force for Minimal Vessel Restenosis, Connect Tissue Res., 2010, 51(4), p 314–32610.3109/03008200903329771 ↗
6. D. Vorwerk, F. Redha, J. Neuerburg, C. Clerc, and R.W. Günther, Neointima Formation Following Arterial Placement of Self-Expanding Stents of Different Radial Force: Experimental Results, Cardiovasc. Interv. Radiol., 1994, 17(1), p 27–3210.1007/BF01102068 ↗
7. P.B. Snowhill, J.L. Nosher, R.L. Siegel, and F.H. Silver, Characterization of Radial Forces in Z Stents, Invest. Radiol., 2001, 36(9), p 521–53010.1097/00004424-200109000-00004 ↗
8. C.R. Johnston, K. Lee, J. Flewitt, R. Moore, G.M. Dobson, and G.M. Thornton, The Mechanical Properties of Endovascular Stents: An in Vitro Assessment, Cardiovasc. Eng., 2010, 10(3), p 128–13510.1007/s10558-010-9097-9 ↗
9. P.E. Labossiere and K.E. Perry, FEA of In-Vitro Test Methods for Radial Force Characterization. Proceedings of the International Conference on Shape Memory and Superelastic Technologies, May 7-11, 2006 (Pacific Grove, California, USA), p 645–650
10. F.J.H. Gijsen, F. Migliavacca, S. Schievano, L. Socci, L. Petrini, A. Thury, J.J. Wentzel, A.F.W. van der Steen, P.W.S. Serruys, and G. Dubini, Simulation of Stent Deployment in a Realistic Human Coronary Artery, Biomed. Eng. Online, 2008, 7(1), p 2310.1186/1475-925X-7-23 ↗
11. W. Wu, M. Qi, X.P. Liu, D.Z. Yang, and W.Q. Wang, Delivery and Release of Nitinol Stent in Carotid Artery and Their Interactions: A Finite Element Analysis, J Biomech., 2007, 40(13), p 3034–304010.1016/j.jbiomech.2007.02.024 ↗
12. P.B. Matthews, B. Kim, A.N. Azadani, T.S. Guy, J.M. Guccione, L. Ge, and E.E. Tseng, Biomechanical Perspective on the Porcine Pulmonary Root Prior to Ross Remodeling, J. Heart Valve Dis., 2009, 18(6), p 682–690
13. C. Lally, F. Dolan, and P.J. Prendergast, Cardiovascular Stent Design and Vessel Stresses: A Finite Element Analysis, J. Biomech., 2005, 38(8), p 1574–158110.1016/j.jbiomech.2004.07.022 ↗
14. L. Petrini, F. Migliavacca, P. Massarotti, S. Schievano, G. Dubini, and F. Auricchio, Computational Studies of Shape Memory Alloy Behavior in Biomedical Applications, J. Biomech. Eng., 2005, 127(4), p 716–72510.1115/1.1934203 ↗
15. R.A. Schatz, J.C. Palmaz, F.O. Tio, F. Garcia, O. Garcia, and S.R. Reuter, Balloon-Expandable Intracoronary Stents in the Adult Dog, Circulation, 1987, 76(2), p 450–45710.1161/01.CIR.76.2.450 ↗
16. D.J. Kereiakes, D.A. Cox, J.B. Hermiller, M.G. Midei, W.B. Bachinsky, E.D. Nukta, M.B. Leon, S. Fink, L. Marin, and A.J. Lansky, Usefulness of a Cobalt Chromium Coronary Stent Alloy, Am. J. Cardiol., 2003, 92(4), p 463–46610.1016/S0002-9149(03)00669-6 ↗
17. H. Tamai, K. Igaki, E. Kyo, K. Kosuga, A. Kawashima, S. Matsui, H. Komori, T. Tsuji, S. Motohara, and H. Uehata, Initial and 6-Month Results of Biodegradable Poly-l-Lactic Acid Coronary Stents in Humans, Circulation, 2000, 102(4), p 399–40410.1161/01.CIR.102.4.399 ↗
18. C. Di Mario, H. Griffiths, O. Goktekin, N. Peeters, J. Verbist, M. Bosiers, K. Deloose, B. Heublein, R. Rohde, V. Kasese, C. Ilsley, and R. Erbel, Drug-Eluting Bioabsorbable Magnesium Stent, J. Interv. Cardiol., 2004, 17(6), p 391–39510.1111/j.1540-8183.2004.04081.x ↗
19. E. Grenadier, R. Shofti, M. Beyar, H. Lichtig, D. Mordechowitz, O. Globerman, W. Markiewicz, and R. Beyar, Self-Expandable and Highly Flexible Nitinol Stent: Immediate and Long-Term Results in Dogs, Am. Heart J., 1994, 128(5), p 870–87810.1016/0002-8703(94)90582-7 ↗
20. V. Kolpakov, M.D. Rekhter, D. Gordon, W.H. Wang, and T.J. Kulik, Effect of Mechanical Forces on Growth and Matrix Protein Synthesis in the In Vitro Pulmonary Artery. Analysis of the Role of Individual Cell Types, Circ. Res., 1995, 77(4), p 823–83110.1161/01.RES.77.4.823 ↗
21. C. Capelli, F. Gervaso, L. Petrini, G. Dubini, and F. Migliavacca, Assessment of Tissue Prolapse After Balloon-Expandable Stenting: Influence of Stent Cell Geometry, Med. Eng. Phys., 2009, 31(4), p 441–44710.1016/j.medengphy.2008.11.002 ↗
22. R. Liao, N.E. Green, S.-Y. James Chen, J.C. Messenger, A.R. Hansgen, B.M. Groves, and J.D. Carroll, Three-Dimensional Analysis of In Vivo Coronary Stent: Coronary Artery Interactions, Int. J. Cardiovasc. Imaging, 2004, 20(4), p 305–31310.1023/B:CAIM.0000041950.84736.e6 ↗
23. J.J. Wentzel, R. Krams, J.C.H. Schuurbiers, J.A. Oomen, J. Kloet, W.J. van der Giessen, P.W. Serruys, and C.J. Slager, Relationship Between Neointimal Thickness and Shear Stress After Wallstent Implantation in Human Coronary Arteries, Circulation, 2001, 103(13), p 1740–174510.1161/01.CIR.103.13.1740 ↗
24. P.B. Matthews, A.N. Azadani, C.S. Jhun, L. Ge, T.S. Guy, J.M. Guccione, and E.E. Tseng, Comparison of Porcine Pulmonary and Aortic Root Material Properties, Ann. Thorac. Surg., 2010, 89(6), p 1981–198810.1016/j.athoracsur.2010.03.002 ↗
25. R. Ní Ghriallais and M. Bruzzi, Self-Expanding Stent Modelling and Radial Force Accuracy, Comput Methods Biomech. Biomed. Eng., 2014, 17(4), p 318–33310.1080/10255842.2012.683427 ↗
26. A.R. Pelton, V. Schroeder, M.R. Mitchell, X.-Y. Gong, M. Barney, and S.W. Robertson, Fatigue and Durability of Nitinol Stents, J. Mech. Behav. Biomed. Mater., 2008, 1(2), p 153–16410.1016/j.jmbbm.2007.08.001 ↗