Fracture of Polymer-Coated Nitinol During Gamma Sterilization
Source abstract
Abstract
After gamma sterilization of a packaged medical device, fractures were discovered in the superelastic nitinol wire used as part of the assembly. The nitinol wire was encased in fluorinated ethylene propylene (FEP) shrink tube. The only fractures occurred where the encased wire was held under strain during gamma sterilization. A study was conducted to determine the susceptibility of nitinol to this type of failure. The variables studied included wire diameter, wire surface finish, wire oxide layer, quantity of wires encased, type of tubing, and strain level during gamma sterilization. The greatest susceptibility to fracture occurred to single wire samples with a light oxide layer held under high strain in FEP shrink tube. Gamma sterilization experiments were conducted to isolate and confirm this failure mechanism. Scanning electron microscopy was used to analyze the fractured samples. Chemical analysis was performed in an attempt to detect trace elements to determine the root cause of the failures. Stress corrosion cracking caused by the liberation of fluorine due to the degradation of the polymer during gamma sterilization is suspected.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Shape Memory Alloy Transformations · Titanium Alloys Microstructure and Properties
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (9)
Show source-provided references
1. D. Norwich and A. Fasching, A Study on the Effect of Diameter on the Fatigue Properties of NiTi Wire, SMST-2008 Proceedings of the International Conference on Shape Memory and Superelastic Technologies, A. Tuissi, M. Mitchel, and J. San Juanm stress corrosion cracking, Ed., Stresa, Italy, 2009, p 558–562.10.1007/s11665-009-9415-9 ↗
2. A. Thompson and N. Moody, Ed., Hydrogen Effects in Materials, TMA, Warrendale, PA, 1996, p 699–840
3. B. James, S. Murray, and S. Saint, Fracture Characterization in Nitinol, SMST-2003 Proceedings of the International Conference on Shape Memory and Superelastic Technologies, A. Pelton and T. Deurig, Ed., Pacific Grove, CA, 2004, p 321–324.
4. H. Chandler, Metallurgy for the Non-Metallurgist, ASM International, Materials Park, OH, 1998, p 185–196
5. S. Cramer and B. Covino, Ed., Corrosion: Fundamentals, Testing, and Protection, ASM Handbook, Vol. 13A, ASM International, Materials Park, OH, 2003, p 346–366
6. M. Wagner, T. Sawaguchi, G. Kaustrater, D. Hoffken, and G. Eggeler, Structural Fatigue of Pseudoelastic NiTi Shape Memory Wires, Materials Science and Engineering, 2004, A378, p 105–10910.1016/j.msea.2003.11.058 ↗
7. L. Massey, The Effects of Sterilization Methods on Plastics and Elastomers, William Andrew Publishing, Norwich, NY, 2005, p 1–6010.1016/B978-081551505-0.50002-0 ↗
8. P. Adler, S. Herreria, G. Kostur, S. Carpenter, P. Poncet, M. Wu, and W. DiScipio, The Role of Alkaline Cleaning Solutions and Cl- Containing Lubricants on Pitting Corrosion in NiTi Alloys, SMST-2003 Proceedings of the International Conference on Shape Memory and Superelastic Technologies, A. Pelton and T. Deurig, Ed., Pacific Grove, CA, 2004, p 385–392.
9. M. Drexel, S. Smith, B. Gause, and D. Plumley, Irradiation-Assisted Stress-Corrosion Cracking of Nitinol During eBeam Sterilization, The International Conference on Shape Memory and Superelastic Technologies, Hong Kong, China, 2011.