journal article·2022-08-03Open access

Implant—Abutment Misfit after Cyclic Loading: An In Vitro Experimental Study

John Eversong Lucena de Vasconcelos, Jefferson David Melo de Matos, Daher Antonio Queiroz, Guilherme da Rocha Scalzer Lopes, Bruna Caroline Gonçalves Vasconcelos de Lacerda, Marco Antonio Bottino, Cecilia Pedroso Turssi, Roberta Tarkany Basting, Flávia Lucisano Botelho do Amaral, Fabiana Mantovani Gomes França

Materials· Volume 15· Issue 15

Abstract

This study aimed to evaluate the influence of thermomechanical cycling (TMC) and type of abutment on the misfit and compressive strength of the implant–abutment interface. Forty 3.75-mm external hexagon implants with 25° angled abutments were divided into four groups (N = 10). Group A: overcast plus TMC; Group B: overcast without TMC; Group C: completely cast plus TMC; Group D: completely cast without TMC. Abutments were fixed to the implants with 32-Ncm torque, and groups A and C specimens were cyclically loaded at 80 N with 2 Hz for 1 million cycles. The misfit on the implant–abutment interface was evaluated by optical microscope (100×) and the compressive strength test was performed in a universal test machine. For statistical analysis, a two-way ANOVA and post hoc Tukey test were used. There was no difference in misfit presented by all the abutments in the absence of TMC (p > 0.05). When TMC was performed, the completely cast abutments showed greater misfit than overcast ones (p = 0.001). Regarding compressive strength, irrespective of TMC performed, the overcast abutments showed higher compressive strength values than completely cast abutments (p = 0.003). Moreover, disregarding the type of abutment used, the absence of TMC provided higher compressive strength values (p < 0.001). It was concluded that thermomechanical cyclic loading aggravated the misfit, especially in completely cast abutments, regardless of material or fabrication technique, and reduced the compressive strength of the two types of abutments tested.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Dental Implant Techniques and Outcomes · Orthopaedic implants and arthroplasty · Dental materials and restorations

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Implantology, College of Dentistry CECAPE, Juazeiro do Norte 63024-015, CE, Brazil

Dental Research Center, São Leopoldo Mandic Institute, Campinas 13045-755, SP, Brazil

Center for Dental Biomaterials, Department of Restorative Dental Sciences, University of Florida (UF Health), Gainesville, FL 32611, USA

Department of Biomaterials, Dental Materials, and Prosthodontics, Institute of Science and Technology, São Paulo State University (Unesp), São José dos Campos 05508-070, SP, Brazil

The University of Texas Health Science Center, Department of Restorative Dentistry & Prosthodontics, Houston (UTHealth) School of Dentistry, Houston, TX 77054, USA

Funding

São Paulo Research Foundation

References (52)

Show source-provided references
1. Tolerance analysis for cast vs machined dental implants10.1016/j.procir.2015.06.047 ↗
2. Influence of CAD/CAM on the fit accuracy of implant-supported zirconia and cobalt-chromium fixed dental prostheses10.1016/j.prosdent.2014.07.010 ↗
3. Influence of Three Types of Abutments on Preload Values Before and After Cyclic Loading with Structural Analysis by Scanning Electron Microscopy10.11607/jomi.2481 ↗
4. Internal vs. external connections for abutments/reconstructions: A systematic review10.1111/j.1600-0501.2012.02556.x ↗
5. Influence of Diamond like Carbon–Coated Screws on the Implant-Abutment Interface
6. A new restorative technique for the perishing implant due to abutment screw fracture10.1563/AAID-JOI-D-12-00231 ↗
7. Influence of lubricant on screw preload and stresses in a finite element model for a dental implant10.1016/j.prosdent.2013.10.016 ↗
8. Effect of cyclic loading and retightening on reverse torque value in external and internal implants10.4047/jap.2015.7.4.288 ↗
9. Torque loss of different abutment sizes before and after cyclic loading10.11607/jomi.3868 ↗
10. Prosthetic Failure in Implant Dentistry10.1016/j.cden.2014.08.008 ↗
11. Variation in the total lengths of abutment/implant assemblies gene rated with a function of applied tightening torque in external and internal implant-abutment connection10.1111/j.1600-0501.2010.02063.x ↗
12. Mechanical Evaluation of an Implant-Abutment Self-Locking Taper Connection: Finite Element Analysis and Experimental Tests10.11607/jomi.2058 ↗
13. Loss of Preload in Screwed Implant Joints as a Function of Time and Tightening/Untightening Sequences10.11607/jomi.3344 ↗
14. Implant–abutment connections: Influence of the design on the micro gap and their fatigue and fracture behavior of dental implants10.1007/s10856-014-5211-7 ↗
15. Influence of alveolar bone loss and different alloys on the biomechanical behavior of internal-and external-connection implants: A three-dimensional finite element analysis10.11607/jomi.3814 ↗
16. Characterization of the interface between cast-to Co-Cr implant cylinders and cast Co-Cr alloys
17. The behavior of a zirconia or metal abutment on the implant-abutment interface during cyclic loading10.1016/j.prosdent.2019.09.023 ↗
18. Effect of unilateral misfit on preload of retention screws of implant-supported prostheses submitted to mechanical cycling. W.G10.1016/j.jpor.2010.05.002 ↗
19. The effect of different implant-abutment connections on screwjoint stability10.1563/AAID-JOI-D-11-00032 ↗
20. Clinical Outcomes of Single Dental Implants with External Connections: Results After 2 to 13 Years
21. Evaluation of the bacterial leakage along with the implant-abutment interface10.4103/0974-6781.91280 ↗
22. Comparative study of torque resistance and micro gaps between a combined Octatorx-cone connection and an internal hexagon implant-abutment connection10.1016/j.prosdent.2014.12.006 ↗
23. The Peri-Implant Sulcus Compared with Internal Implant and Suprastructure Components: A Microbiological Analysis10.1111/j.1708-8208.2009.00220.x ↗
24. Micro gap between zirconia abutments and titanium implants
25. Screw joint stability after the application of retorquing in implant-supported dentures under simulated masticatory conditions10.1016/j.prosdent.2013.07.024 ↗
26. The influence of abutment angulation on screw loosening of implants in the anterior maxilla
27. The Effect of Mechanical Cycling and Different Misfit Levels on Vicker’s Micro hardness of Retention Screws for Single Implant-Supported Prostheses10.1111/j.1532-849X.2011.00753.x ↗
28. Impact of abutment rotation and angulation on marginal fit: Theoretical considerations
29. Stress Distribution in the Abutment and Retention Screw of a Single Implant Supporting a Prosthesis with Platform Switching10.11607/jomi.2813 ↗
30. Biomechanical Implant Treatment Complications: A Systematic Review of Clinical Studies of Implants with at Least 1 Year of Functional Loading
31. Does abutment collar length affect abutment screw loosening after cyclic loading?10.1563/AAID-JOI-D-D-14-00021 ↗
32. Implant screw fracture10.4103/0974-6781.118868 ↗
33. The Role of Implant/Abutment System on torque Maintenance of Retention Screws and vertical Misfit of Implant-Supported Crowns Before and After Mechanical Cycling10.11607/jomi.2727 ↗
34. Load to failure of different titanium abutments for an internal hexagon implant10.1016/j.prosdent.2015.04.012 ↗
35. Leakage of Saliva Through the Implant-Abutment Interface: In Vitro Evaluation of Three Different Implant Connections Under Unloaded and Loaded Conditions
36. Fatigue induced changes in conical implant-abutment connections10.1016/j.dental.2015.09.004 ↗
37. Marginal fit and microbial leakage along with the implant-abutment interface of fixed partial prostheses: An in vitro analysis using Checkerboard DNA-DNA hybridization10.1016/j.prosdent.2015.05.009 ↗
38. Dynamic fatigue performance of implant-abutment assemblies with different tightening torque values10.3233/BME-141025 ↗
39. Effects of Abutment Screw Coating on Implant Preload10.1111/j.1532-849X.2010.00595.x ↗
40. In Vitro Analysis of Post–fatigue Reverse-Torque Values the Dental Abutment/Implant Interface for a Unitarian Abutment Design10.1111/j.1532-849X.2011.00756.x ↗
41. Effect of the coefficient of friction and tightening speed on the preload induced at the dental implant complex with the finite element method10.1016/j.prosdent.2014.09.021 ↗
42. Coating Dental Implant Abutment Screws with Diamondlike Carbon Doped with Diamond Nanoparticles: The Effect on Maintaining Torque After Mechanical Cycling10.11607/jomi.3933 ↗
43. Evaluation of UCLA Implant-Abutment Sealing10.11607/jomi.3217 ↗
44. Dimensional Tolerances and Assembly Accuracy of Dental Implants and Machined Versus Cast-On Abutments10.1111/j.1708-8208.2009.00189.x ↗
45. Marginal and Internal Adaptation of Different Superstructure and Abutment Materials Using Two Different Implant Systems for Five-Unit Implant-Supported Fixed Partial Dentures: An In Vitro Study10.11607/jomi.2744 ↗
46. Changes in the abutment-implant interface in Morse taper implant connections after mechanical cycling: A pilot study10.11607/jomi.3113 ↗
47. A conservative approach to retrieve a fractured abutment screw–A case report
48. Dental Implants with Conical Implant-Abutment Interface: Influence of the Conical Angle Difference on the Mechanical Behavior of the Implant10.11607/jomi.2775 ↗
49. Two Implant/Abutment Joint Designs: A Comparative Finite Element Analysis10.11607/jomi.2433 ↗
50. Mechanical Testing of Implant-Supported Anterior Crowns with Different Implant /Abutment Connections10.11607/jomi.2443 ↗
51. Load fatigue performance of conical implant-abutment connections
52. Implant-abutment interface: A comparison of the ultimate force to failure among narrow-diameter implant systems10.1016/j.prosdent.2014.01.020 ↗