A New Surgical Drill Bit Concept for Bone Drilling Operations
Source abstract
Abstract
Abstract During surgical interventions, bone tissue may be damaged by the achieved high cutting efforts and thermal damage, which should be avoided. In this study, a new surgical drill bit with two types of cutting edges is proposed and tested, reducing tissue damage. Large part of the material is removed by brittle crack propagation, and surface is finished precisely with small uncut chip thickness. Even if temperatures achieved in the bone are not reduced due to rubbing effects, lower cutting force and processing energy, and better surface roughness are obtained with the proposed tool compared to the conventional drill bit. Keywords: BoneCuttingDamageDrillingForcesInfraredSurgicalTemperaturesThermographyTissue
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Dental Implant Techniques and Outcomes · Bone Tissue Engineering Materials · Orthopaedic implants and arthroplasty
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (18)
Show source-provided references
1. Haswell M. · Implant Practice · 2009 · 2 · 44
2. Eriksson A. · Journal of Prosthetic Dentistry · 1983 · 50 · 10110.1016/0022-3913(83)90174-9 ↗
3. Lundskog , J. Heat and bone tissue. An Experimental Investigation of the Thermal Properties of Bone Tissue and Threshold Levels for Thermal Injury. Ph.D. Dissertation, University of Göteborg, Göteborg, Sweden, 1972.
4. Bertollo N.
5. Lee J. · Medical Engineering and Physics · 2011 · 33 · 123410.1016/j.medengphy.2011.05.014 ↗
6. Abouzgia M.B. · International Journal of Oral and Maxillofacial Surgery · 1996 · 25 · 39410.1016/S0901-5027(06)80040-8 ↗
7. Tu , Y.K. ; Chen , L.W. ; Chiang , C.H. ; Chen , Y.C. ; Lu , W.H. ; Tsai , H.H. Thermal contact simulation of drill bit and bone during drilling. In4th International Conference on Bioinformatics and Biomedical Engineering, iCBBE, Chengdu , China , June 18–20 , 2010 ; 1 – 4 .10.1109/ICBBE.2010.5517275 ↗
8. Alam , K. Experimental and Numerical Analysis of Conventional and Ultrassonically Assisted Cutting of Bone. Ph.D. Dissertation, Loughborough University, UK, 2009.
9. Alam K. · Medical Engineering and Physics · 2010 · 33 · 23410.1016/j.medengphy.2010.10.003 ↗
10. Soriano J. · Advanced Materials Research · 2012 · 498 · 14510.4028/www.scientific.net/AMR.498.145 ↗
11. Jacob C. · Journal of Biomechanics · 1976 · 9 · 34310.1016/0021-9290(76)90056-7 ↗
12. Albertini M. · Materials Letters · 2012 · 79 · 16310.1016/j.matlet.2012.04.006 ↗
13. Bertollo N. · Clinical Biomechanics · 2010 · 25 · 61310.1016/j.clinbiomech.2010.02.007 ↗
14. Sugita N. · CIRP Annals–Manufacturing Technology · 2009 · 58 · 11310.1016/j.cirp.2009.03.057 ↗
15. Sugita N. · Journal of Biomechanics · 2009 · 42 · 282610.1016/j.jbiomech.2009.08.017 ↗
16. Sugita , N. ; Watanabe , M. ; Sano , Y. ; Terashima , M. ; Mitsuishi , M. Cutting tool for bone machining to carry out roughing and finishing in a single pass. InProceedings of The 6th International Conference on Leading Edge Manufacturing in 21st Century (LEM21), Saitama, Japan, November 8–10, 2011; paper no. 3285.10.1299/jsmelem.2011.6._3285-1_ ↗
17. Wiggins K. · Transactions of the ASME, Journal of Biomechanical Engineering · 1978 · 100 · 12210.1115/1.3426202 ↗
18. Childs T. · Machining Science and Technology · 2011 · 15 · 20610.1080/10910344.2011.580699 ↗