Biomechanical Analysis of Locking Reconstruction Plate Using Mono- or Bicortical Screws
Source abstract
Abstract
This experimental in vitro study evaluated the influence of screw length on the mechanical properties of a locking reconstruction plate designed with locking rings inserted into plate holes. Synthetic bone cylinders with 10 mm fracture gap and seven-hole locking reconstruction plates were used. Two groups of bone-plate constructs were assembled: Group 1 – three monocortical screws on each fracture side, Group 2 - three bicortical screws on each fracture side. In each group nine bone-plate constructs were tested until failure, three each in bending, compression and torsion. In each group, 21 bone-plate constructs were tested for failure in fatigue testing, seven each for bending, compression and torsion. In all static testing no significant differences were found between G1 and G2, except ultimate moment in torsion test (G2>G1; P=0.008). Statistical analysis revealed significant differences between groups in axial compression fatigue testing (G1>G2; P<0.05) and four-point bending fatigue testing (G1<G2; P<0.05) in maximum load, minimum load, maximum moment, and minimum moment. In conclusion, screw length can affect the mechanical properties of locking reconstruction plate. Compared to bicortical screws, monocortical screws were less resistant to bending than axial compression. This must be considered when choosing implant, particularly in fractures under high axial loads.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Bone fractures and treatments · Orthopedic Surgery and Rehabilitation · Shoulder and Clavicle Injuries
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Universidade Estadual Paulista, Brazil
Laboratório de Ensaios Mecânicos e Metalográficos, Brazil
Distrito Industrial Uninorte, Brazil
References (41)
Show source-provided references
1. Biomechanics of locked plates and screws10.1097/00005131-200409000-00003 ↗
2. Biomechanical concepts applicable to minimally invasive fracture repair in small animals10.1016/j.cvsm.2012.07.007 ↗
3. Johnston SA · 2012 · 576
4. Reconstruction plates for stabilization of mid-shaft clavicle fractures: Differences between nonlocked and locked plates in two different positions10.1016/j.jse.2008.10.002 ↗
5. Johnson AL · 2013 · 1086
6. Piermattei DL · 2006 · 125
7. Guidelines for the clinical application of the LCP10.1016/j.injury.2003.09.026 ↗
8. General principles for the clinical use of the LCP10.1016/j.injury.2003.09.023 ↗
9. The use and abuse of locking plates10.1016/j.mporth.2009.07.002 ↗
10. The concept of locking plates
11. Biomechanical testing of the LCP- how can stability in locked internal fixator be controlled?10.1016/j.injury.2003.09.021 ↗
12. A review of locking compression plate biomechanics and their advantages as internal fixators in fracture healing10.1016/j.clinbiomech.2007.08.004 ↗
13. Effect of screw insertion torque on mechanical properties of four locking systems10.1111/j.1532-950X.2013.12023.x ↗
14. Advances in internal fixation locking plates10.1136/inp.g7695 ↗
15. A biomechanical study of conventional acetabular internal fracture fixation versus locking plate fixation
16. Locking plates: tips and tricks10.2106/00004623-200710000-00028 ↗
17. Biomechanical comparison of different stabilization constructs for unstable posterior wall fractures of acetabulum. A cadaveric study10.1371/journal.pone.0082993 ↗
18. Evolution of the internal fixation of long bone fractures. The scientifgic basis of biological internal fixation: choosing a new balance between stability and biology10.1302/0301-620X.84B8.13752 ↗
19. A biomechanical evaluation of methods of distal humerus fracture fixation using locking compression plates versus conventional reconstruction plates10.1097/00005131-200405000-00004 ↗
20. Fixation of extra-articular distal humerus fractures using one locking plate versus two reconstruction plates: a laboratory study10.1097/TA.0b013e318181e53c ↗
21. Internal fixation of the distal humerus: a comprehensive biomechanical study evaluating current fixation techniques10.1097/BOT.0b013e3182a6693f ↗
22. Does plate adaptation affect stability? A biomechanical comparison of locking and nonlocking plate10.1053/joms.2002.35732 ↗
23. Ex vivo biomechanical comparison of the 2.4 mm uniLOCK reconstruction plate using 2.4 mm locking versus standard screws for fixation of acetabular osteotomy in dogs10.1111/j.1532-950X.2008.00441.x ↗
24. Biomechanical comparison of mono- and bicortical screws in an experimentally induced gap fracture10.3415/VCOT-14-03-0040 ↗
25. Innovations in locking plate technology10.5435/00124635-200407000-00001 ↗
26. Revolution in plate osteosynthesis: new internal fixator systems10.1007/s007760300044 ↗
27. Fixation of diaphyseal fractures with a segmental defect: a biomechanical comparison of locked and conventional plating techniques10.1097/01.ta.0000195462.53525.0c ↗
28. Relative stability of conventional and locked plating fixation in a model of the osteoporotic femoral diaphysis10.1016/j.clinbiomech.2008.11.002 ↗
29. Biomechanical evaluation of locking plate radial shaft fixation: unicortical locking fixation versus mixed bicortical and unicortical fixation in a sawbone model10.1016/j.jhsa.2007.05.019 ↗
30. Comparison of the mechanical behaviors of semicontoured, locking plate–rod fixation and anatomically contoured, conventional plate–rod fixation applied to experimentally induced gap fractures in canine femora10.2460/ajvr.70.1.23 ↗
31. Locking reconstruction double plating of distal humeral fractures: how many screws in the distal ulnar column segment in A3 fracture provide superior stability?10.1097/BOT.0b013e3181a87725 ↗
32. Effect of plate working length on plate stiffness and cyclic fatigue life in a cadaveric femoral fracture gap model stabilized with a 12-hole 2.4 mm locking compression plate10.1186/1746-6148-9-125 ↗
33. Ex vivo evaluation of the biomechanical effect of varying monocortical screw numbers on a plate-rod canine femoral gap model10.3415/VCOT-12-05-0061 ↗
34. The effect of locked screw angulation on the biomechanical properties of the S.P.S. Free-Block plate10.3415/VCOT-12-03-0045 ↗
35. Use of a plate and screws to repair lateral humeral condylar fractures in 10 dogs
36. Fixation of humeral intercondylar fractures using a lateral plate in 14 dogs supported by finite element analysis of repair10.1160/VCOT-07-01-0002 ↗
37. The ComPact Unilock 2.0/2.4 system and its clinical application in small animal orthopedics10.1055/s-0038-1632936 ↗
38. Temporary transarticular stabilization with a locking plate for medial shoulder luxation in a dog10.3415/VCOT-07-07-0066 ↗
39. Repair of long-bone fractures in cats and small dogs with the Unilock mandible locking plate system10.3415/VCOT-08-09-0084 ↗
40. The evolution of locked plates
41. Locked plating: indications and current concepts