Investigation of Contact Galvanic Couples in the Processes of Disinfection of Medical Instruments
Source abstract
Abstract
It is shown that the contact of POS-61 solder with structural materials used for medical purposes exerts practically no influence on its corrosion rate in oxidizing disinfectant solutions. The negative effect of galvanic couples formed in these solutions is connected with rapid decomposition of disinfectants, i.e., with the process of reduction of oxidants on the cathode.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Medical Device Sterilization and Disinfection · Ergonomics and Human Factors · Mechanical Systems and Engineering
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.