journal article·2002-11

Investigation of Contact Galvanic Couples in the Processes of Disinfection of Medical Instruments

O. K. Sevydova, I. D. Roi

Materials Science· Volume 38· Issue 6
Publisher page

Abstract

It is shown that the contact of POS-61 solder with structural materials used for medical purposes exerts practically no influence on its corrosion rate in oxidizing disinfectant solutions. The negative effect of galvanic couples formed in these solutions is connected with rapid decomposition of disinfectants, i.e., with the process of reduction of oxidants on the cathode.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Medical Device Sterilization and Disinfection · Ergonomics and Human Factors · Mechanical Systems and Engineering

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (5)

Show source-provided references
1. OST 42–21–2–85. Sterilization and Disinfection of Medical Instruments [in Russian], VNIIDiS and VNIIIMT, Moscow (1985).
2. OST 25.1–005–87. Resistance of Metallic Medical Instruments to Disinfectants, Presterilizing Cleaning, and Sterilization. Classification. Choice of the Method [in Russian], Ministry of Instrument Making, Automation, and Control Systems, Moscow (1988).
3. V. Ya. Belous, L. Ya. Gurvich, A. D. Zhirnov, et al., “Contact corrosion of high-strength stainless steels,” Zashch. Met., 34, No. 3, 266–272 (1998).
4. T. V. Chezhina, L. N. Gorchakov, and Yu. I. Archakov, “On the corrosion-electrochemical behavior of chromium steels in solutions of formic acid,” Zashch. Metal., 23, No. 2, 275–278 (1987).
5. P. S. Mel'nikov, A Handbook of Electroplating in Machine Building [in Russian], Mashinostroenie, Moscow (1979).