journal article·2023-01

Study on relaxation phenomenon of CdZnTe photon counting detectors in X-ray imaging

Yang Kang, Rui Wu, Sen Wu, Tingting Tan, Yingrui Li, Gangqiang Zha

Publisher page

Abstract

During X-ray scanning imaging, the X-ray dose received by CdZnTe photon counting detectors varies dynamically because of the different absorption coefficients that occur at different locations in the object being imaged. This paper explores the phenomenon of count rate relaxation in CdZnTe detectors during X-ray dose mutations and studies the intrinsic relationship of this phenomenon with the polarization effect. The characterization of the count rate relaxation processes of CdZnTe detectors at three typical X-ray doses are studied in detail. It is concluded that the count rate relaxation process of the CdZnTe detectors correspond to the polarization process, which can be divided into three stages according to the accumulation rate of space charges , and the relaxation time depends on the settling time of the polarizing electric field. In addition, the effects of temperature and bias voltage on the relaxation phenomenon of CdZnTe detectors are studied. The results show the relaxation process of the count rate can be regulated by adjusting the temperature and bias voltage of the CdZnTe detectors. Changes in temperature and bias mainly act on the trapping and de-trapping processes of deep-level traps, thereby affecting the establishment of the polarization electric field, which further proves the close relationship between the count rate relaxation phenomenon and the polarization effect. Finally, the effects of count rate relaxation on DR image quality are discussed based on self-built DR imaging system.
Source: expertdata · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Advanced X-ray and CT Imaging · Medical Imaging Techniques and Applications · Advanced Semiconductor Detectors and Materials

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (17)

Show source-provided references
1. Tutorial on X-ray photon counting detector characterization10.3233/XST-16210 ↗
2. Tilted angle CZT detector for photon counting/energy weighting x-ray and CT imaging10.1088/0031-9155/51/17/010 ↗
3. Energy-sensitive photon counting detector-based X-ray computed tomography10.1007/s12194-017-0390-9 ↗
4. Improved quality using newly designed algorithms in gamma- and X-ray fusion images with a photon counting CZT detector: combining the median modified Wiener filter and edge detection method10.1016/j.ijleo.2021.167681 ↗
5. Effect of deep-level defects on the performance of CdZnTe photon counting detectors10.3390/s20072032 ↗
6. Research progress on semiconductor materials and devices for radiation detection
7. X-ray response of CdZnTe detectors grown by the vertical Bridgman technique: energy, temperature and high flux effects10.1016/j.nima.2016.08.029 ↗
8. Vision 20/20: single photon counting x-ray detectors in medical imaging10.1118/1.4820371 ↗
9. Investigation on X-ray photocurrent response of CdZnTe photon counting detectors10.3390/s20020383 ↗
10. Nature of polarization in wide-bandgap semiconductor detectors under high-flux irradiation: application to semi-insulating Cd1-xZnxTe10.1103/PhysRevB.77.035205 ↗
11. Effect and mechanism of temperature on photon counting performance of CdZnTe detectors
12. Polarization in cadmium telluride nuclear radiation detectors10.1109/TNS.1976.4328231 ↗
13. Flux-dependent electric field changes in semi-insulating CdZnTe10.1088/0022-3727/46/23/235306 ↗
14. Characterization of polarizing semiconductor radiation detectors by laser-induced transient currents10.1063/1.4997404 ↗
15. CdTe/CZT under high flux irradiation10.1088/1748-0221/6/01/C01055 ↗
16. Dynamic lateral polarization in CdZnTe under high flux X-ray irradiation10.1109/TNS.2007.905159 ↗
17. A mechanism for dynamic lateral polarization in CdZnTe under high flux x-ray irradiation10.1063/1.2883924 ↗