Compliance of a Biocomposite Vascular Tissue in Longitudinal and Circumferential Directions as a Basis for Creating Artificial Substitutes
Source abstract
Abstract
Experimental studies on the mechanical behavior of 11 human common carotid arteries at different values of internal pressure and axial force were performed on a device allowing us to measure the internal pressure, axial force, and circumferential and longitudinal deformations of the vessel. The persons age ranged from 20 to 28 years. Two types of experiments were carried out. In the first series, cylindrical samples were gradually loaded by an internal pressure from 0 to 200 mmHg at different longitudinal stretch ratios. The second series included axial extension of the same samples at different circumferential stretch ratios. The undulation level of elastin membranes in longitudinal and circumferential directions for samples fixated at different values of internal pressure and longitudinal stretch ratio were determined from histological data.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Elasticity and Material Modeling · Electrospun Nanofibers in Biomedical Applications · Tribology and Lubrication Engineering
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (48)
Show source-provided references
1. H. W. Weizsacker and J. G. Pinto, “Isotropy and anisotropy of the arterial wall,” J. Biomech., 21, 477–487 (1988).10.1016/0021-9290(88)90240-0 ↗
2. J. Zhou and Y. C. Fung, “The degree of nonlinearity and anisotropy of blood vessel elasticity,” Proc. Nat. Acad. Sci. USA, 94, No. 26, 14255–14260 (1997).10.1073/pnas.94.26.14255 ↗
3. G. J. L'Italien, N. R. Chandrasekar, G. M. Lamuraglia, W. C. Pevec, S. Dhara, D. F. Warnock, and W. M. Abbott, “Biaxial elastic properties of rat arteries in vivo: influence of vascular wall cells on anisotropy,” Am. J. Physiol., 267, Pt. 2, H574–H579 (1994).
4. R. N. Vaishnav, J. T. Young, J. S. Janicki, and D. J. Patel, “Nonlinear anisotropic elastic properties of the canine aorta,” Biophys. J., No. 12, 1008–1027 (1972).10.1016/S0006-3495(72)86140-X ↗
5. R. H. Cox, “Anisotropic properties of the canine carotid artery in vivo,” J. Biomech., No. 8, 293–300 (1975).10.1016/0021-9290(75)90081-0 ↗
6. P. B. Dobrin and T. R. Canfield, “Elastase, collagenase, and the biaxial elastic properties of dog carotid artery,” Am. J. Physiol., 247, 124–131 (1984).
7. T. V. How, “Mechanical properties of arteries and arterial grafts,” in: G. Hastings (ed.), Cardiovascular Biomaterials Springer, London (1992), pp. 1–35.10.1007/978-1-4471-1847-3_1 ↗
8. A. T. Yokobori and T. Yokobori, “The mechanical test method of cardiovascular and related biomaterials,” Bio-Med. Mater. Eng., No. 1, 25–43 (1991).10.3233/BME-1991-1105 ↗
9. V. Kasyanov, B. Purinya, V. Kasyanenko, N. Kochurkova, and N. Sugonyaeva, “Biomechanical properties of novel compliant textile grafts,” in: Transactions of the Fourth World Biomaterials Congress, Berlin (1992), p. 42.
10. V. Kasyanov, V. Mironov, and R. R. Markwarld, “Cardiovascular tissue engineering: blood vessels and heart valves as composite biomaterials,” in: Abstracts of Second Biannual Meeting of the Tissue Engineering Society, Orlando, Florida (1998), p. 501.
11. L. E. Niklason, J. Gao, W. M. Abbott, K. K. Hirschi, S. Houser, R. Marini, and R. Langer, “Functional arteries grown in vitro,” Science, 228, 489–493 (1999).10.1126/science.284.5413.489 ↗
12. G. J. Langewouters, K. H. Wesseling, and W. J. Goedhard, “The pressure-dependent dynamic elasticity of 35 thoracic and 16 abdominal human aortas in vitro described by a five-component model,” J. Biomech., 18, 613–620 (1985).10.1016/0021-9290(85)90015-6 ↗
13. P. B. Dobrin, “Vascular mechanics,” in: J. Shepherd and F. Abboud (eds.), Handbook of Physiology. The Cardiovascular System, Springer, New York (1983), pp. 64–102.
14. A. M. Brant, M. F. Teodori, R. L. Kormos, and H. S. Borovetz, “Effect of variations in pressure and flow on the geometry of isolated canine carotid arteries,” J. Biomech., 20, 831–838 (1987).10.1016/0021-9290(87)90143-6 ↗
15. K. Hayashi, “Experimental approaches on measuring the mechanical properties and constitutive laws of arterial walls,” J. Biomech. Eng., 115, No. 4B, 481–488 (1993).10.1115/1.2895528 ↗
16. G. A. Johnson, T. K. Hung, A. M. Brant, and H. S. Borovetz, “Experimental determination of wall shear rate in canine carotid arteries perfused in vitro,” J. Biomech., 22, 1141–1150 (1989).10.1016/0021-9290(89)90216-9 ↗
17. K. Pertold, M. Resch, and R. Peter, “Three-dimensional numerical analysis of pulsatile flow and wall shear stress in the carotid artery bifurcation,” J. Biomech., 24, 409–420 (1991).10.1016/0021-9290(91)90029-M ↗
18. H. Wolinsky and S. Glagov, “Structural basis for the static mechanical properties of the aortic media,” Circ. Res., 14, 400–413 (1964).10.1161/01.RES.14.5.400 ↗
19. K. B. Chandran, D. Gao, G. Han, H. Baraniewski, and J.D. Corson, “Finite-element analysis of arterial anastomoses with vein, Dacron, and PTFE grafts,” Med. Biol. Eng. Comput., 30, No. 4, 413–418 (1992).10.1007/BF02446179 ↗
20. K. Rhee and J. Tarbell, “A study of the wall shear rate distribution near the end-to-end anastomosis of a rigid graft and a compliant artery,” J. Biomech., 27, No.3, 329–338 (1994).10.1016/0021-9290(94)90009-4 ↗
21. A. Edwards, R. J. Carson, S. Bowald, and W. S. Quist, “Development of a microporous compliant small bore vascular graft,” J. Biomater. Appl., 10, No. 2, 171–187 (1995).10.1177/088532829501000207 ↗
22. W. Matthew, M. Weston, K. Rhee, and J. Tarbell, “Compliance and diameter mismatch affect the wall shear-rate distribution near an end-to-end anastomosis, “J. Biomech., 29, No. 2, 187–198 (1996).10.1016/0021-9290(95)00028-3 ↗
23. R. J. Zdrahala, “Small-caliber vascular grafts. Pt. I. State of the art,” J. Biomater. Appl., 10, No. 4, 309–329 (1996).10.1177/088532829601000402 ↗
24. K. Doi and T. Matsuda, “Significance of porosity and compliance of microporous, polyurethane-based microarterial vessel on neoarterial wall regeneration,” J. Biomed. Mater. Res., 37, No. 4, 573–584 (1997).10.1002/(SICI)1097-4636(19971215)37:4<573::AID-JBM17>3.0.CO;2-9 ↗
25. G. W. Bos, A. A. Poot, T. Beugeling, W. G. van Aken, and J. Feijen, “Small-diameter vascular graft prosthesis: current status,” Arch. Physiol. Biochem., 106, No. 2, 100–115 (1998).10.1076/apab.106.2.100.4384 ↗
26. W. M. Abbott and D. J. Bouchier-Hayes, “The role of mechanical properties in graft desing,” in: H. Dardik (ed.), Graft Material Vascular Surgery, Year Book Medical Publisher, Chicago (1978), pp. 59–78.
27. K. Hayashi, H. Handa, S. Nagasawa, A. Okumura, and K. Moritake, “Stiffness and elastic behavior of human intracranial and extracranial arteries,” J. Biomech., 13, 175–184 (1980).10.1016/0021-9290(80)90191-8 ↗
28. P. Boutouyrie, C. Glaser, A. Moryusef, Y. Bezie, J. N. Fabiani, S. Laurent, and P. Lacolley, “Associations between viscoelastic properties of large arteries and their extracellular matrix composition in abdominal aortic aneurysms in humans,” Therapie, 54, No. 1, 85–91 (1999).
29. T. Imura, K. Yamamoto, T. Satoh, K. Kanamori, T. Mikami, and H. Yasuda, “In vivo viscoelastic behavior of the human aorta,” Circ. Res., 66, No. 5, 1413–1419 (1990).10.1161/01.RES.66.5.1413 ↗
30. E. Rosset, C. Brunet, R. Rieu, P. Rolland, J. F. Pellissier, P. E. Magnan, P. Foulon, A. Drizenko, M. Laude, A. Branchereau, and A. Friggi, “Viscoelastic properties of human arteries. Methodology and preliminary results,” Surg. Radiol. Anat., 18, 2, 89–96 (1996).10.1007/BF01795225 ↗
31. S. M. Wells, B. L. Langille, and S. L. Adamson, “In vivo and in vitro mechanical properties of the sheep thoracic aorta in the perinatal period and adulthood,” Am. J. Physiol., 274, No. 5, Pt. 2, H1749–H1760 (1998).
32. M. L. Raghavan, M. W. Webster, and D. A. Vorp, “Ex vivo biomechanical behavior of abdominal aortic aneurysm: assessment using a new mathematical model,” Ann. Biomed. Eng., 24, 573–582 (1996).10.1007/BF02684226 ↗
33. P. B. Dobrin, R. Mirande, S. Kang, Q. S. Dong, and R. Mrkvicka, “Mechanics of end-to-end artery-to-PTFE graft anastomoses,” Vasc. Surg., 12, No. 4, 317–323 (1998).10.1007/s100169900161 ↗
34. M. H. Wu, Q. Shi, L. R. Sauvage, S. Kaplan, N. Hayashi, and M. D. Patel, “The direct effect of graft compliance mismatch per se on the development of host arterial intimal hyperplasia at the anastomotic interface,” Ann. Vasc. Surg., 7, No. 2, 156–168 (1993).10.1007/BF02001010 ↗
35. P. D. Ballyk, C. Walsh, J. Butary, and M. Ojha, “Compliance mismatch may promote graft-artery intimal hyperplasia by altering suture-line stresses,” J. Biomech., 31, 229–237 (1998).10.1016/S0197-3975(97)00111-5 ↗
36. A. Giudiceandrea, A. M. Seifalian, B. Krijgsman, and G. Hamilton, “Effect of prolonged pulsatile shear stress in vitro on endothelial cell seeded PTFE and compliant polyurethane vascular grafts,” Eur. J. Vasc. Endovasc. Surg., 15, No. 2, 147–154 (1998).10.1016/S1078-5884(98)80136-6 ↗
37. S. Y. Kim, T. J. Hinkamp, W. R. Jacobs, R. C. Lichtenberg, H. Posniak, and R. Pifarre, “Effect of an inelastic aortic synthetic vascular graft on exercise hemodynamics,” Ann. Thorac. Surg., 59, No. 4, 981–989 (1995).10.1016/0003-4975(95)00068-V ↗
38. R. Roeder, J. Wolfe, N. Lianakis, T. Hinson, L. A. Geddes, and J. Obermiller, “Compliance, elastic modulus, and burst pressure of small-intestine submucosa (SIS), small-diameter vascular grafts,” J. Biomed. Mater. Res., 47, No. 1, 65–70 (1999).10.1002/(SICI)1097-4636(199910)47:1<65::AID-JBM9>3.0.CO;2-F ↗
39. N. R. Tai, A. Giudiceandrea, H. J. Salacinski, A. M. Seifalian, and G. Hamilton, “In vivo femoropopliteal arterial wall compliance in subjects with and without lower limb vascular disease,” J. Vasc. Surg., 30, No. 5, 936–945 (1999).10.1016/S0741-5214(99)70020-0 ↗
40. R. Walden, G. J. L'Italien, J. Megrman, and W. M. Abbott, “Matched elastic properties and successful arterial grafting,” Arch. Surg., 115, No. 10, 1166–1169 (1980).10.1001/archsurg.1980.01380100018004 ↗
41. M. W. Weston, K. Rhee,and J. M. Tarbell, “Compliance and diameter mismatch affect the wall shear-rate distribution near an end-to-end anastomosis,” J. Biomech., 26, No. 2, 187–198 (1996).10.1016/0021-9290(95)00028-3 ↗
42. R. White, L. Goldberg, and F. Hirose, “Effect of healing on small internal diameter arterial graft compliance,” Biomater. Med. Dev. Artif. Organs, No. 11, 21–29 (1983).10.3109/10731198309118791 ↗
43. R. A. White, S. R. Klein, and F. Shorst, “Maintenance of compliance in a small-diameter arterial prosthesis,” Trans. Am. Soc. Artif. Intern. Organs, 32, 601–604 (1986).
44. W. Trubel, H. Schima, A. Moritz, F. Raderer, A. Windisch, R. Ullrich, U. Windberger, U. Losert, and P. Polterauer, “Compliance mismatch and formation of distal anastomotic intimal hyperplasia in externally stiffened and lumen-adapted venous grafts,” Eur. J. Vasc. Endovasc. Surg., 10, No. 4, 415–423 (1995).10.1016/S1078-5884(05)80163-7 ↗
45. G. K. Naughton, “Tissue engineering constructs: from lab bench to market,” in: Proc. 1st Smith Nephew International Symposium, York, UK (1997), p. 2.
46. G. Zund, S. P. Hoerster, A. Schoeberlein, Q. Ye, M. Lachat, P. Vogt, and M. I. Tuina, “Tissue engineering in cardiovascular surgery: where do we stand?,” in: Proc. VBiennial Meeting 1st Soc. Applied Cardiovascular Biology, Munich, Germany (1998), p. IV–05.
47. N. L'Heureux, S. Paquesr, R. Labbe, L. Geremain, and F.A. Anger, “A completely biological tissue-engineered human blood vessel,” FASEB J., No. 12, 47–56 (1998).
48. B. S. Gupta and V. A. Kasyanov, “Biomechanics of human common carotid artery and design of novel hybrid textile compliant vascular grafts,” J. Biomed. Mater. Res., 34, 341–349 (1997)10.1002/(SICI)1097-4636(19970305)34:3<341::AID-JBM9>3.0.CO;2-K ↗