Abstract

A portable electrophoresis device based on the SU‐8/Pyrex microfluidic chip with integrated three 100 μm platinum (Pt) thin‐film electrodes (working, reference, and auxiliary) was adapted for the analysis of silver nanoparticles (AgNPs). The microfluidic platform includes the main electrophoresis instrument, reusable holder, and microfluidic chip with integrated Pt electrodes. The electrochemical oxidation behaviours of AgNPs on Pt thin‐film electrode were investigated by using cyclic voltammetry, while the electrophoresis behaviours of AgNPs were studied on the ‘Lab‐on‐a‐chip’ platform. The influences of parameters including running buffer, separation voltage, and detection potential on the electrophoresis behaviours of AgNPs were all investigated. The results showed that AgNPs could be oxidised at 0.2–0.4 V [phosphate buffer solution buffer, pH 7.0] on the Pt thin‐film electrode. The peak current enhanced with the increase of separation voltage, while the retention time decreased. The peak current varied with the detection potential and reached the highest value at 0.4 V. This microfluidic platform could be a fast and portable detection system for the analysis of total AgNPs.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Electrochemical Analysis and Applications · Analytical Chemistry and Sensors · Microfluidic and Capillary Electrophoresis Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Key Laboratory of Coastal Environmental Processes and Ecological Remediation Yantai Institute of Coastal Zone Research (YIC) Chinese Academy of Sciences (CAS) Yantai 264003 Shandong People's Republic of China

Shandong Provincial Key Laboratory of Coastal Environmental Processes YICCAS Yantai 264003 Shandong People's Republic of China

University of Chinese Academy of Sciences Beijing 100049 People's Republic of China

Fuan Pharmaceutical Group Yantai Justaware Pharmaceutical Co., Ltd. Yantai 264002 Shandong People's Republic of China

School of Chemistry and Chemical Engineering Yantai University Yantai 264005 Shandong People's Republic of China

References (25)

Show source-provided references
1. 10.1016/j.chroma.2011.09.06210.1016/j.chroma.2011.09.062 ↗
2. 10.1016/0925‐4005(90)80209‐I10.1016/0925‐4005(90)80209‐I ↗
3. 10.1373/clinchem.2005.05960010.1373/clinchem.2005.059600 ↗
4. 10.1021/ac060560210.1021/ac0605602 ↗
5. 10.1002/elps.20120023610.1002/elps.201200236 ↗
6. Microfluidic chips for clinical and forensic analysis
7. 10.1002/elps.20090036510.1002/elps.200900365 ↗
8. 10.1002/elps.20110021810.1002/elps.201100218 ↗
9. 10.1002/elps.20090029710.1002/elps.200900297 ↗
10. 10.1016/j.talanta.2011.02.04010.1016/j.talanta.2011.02.040 ↗
11. 10.1002/elps.20080047810.1002/elps.200800478 ↗
12. 10.1002/1522-2683(200211)23:21<3659::AID-ELPS3659>3.0.CO;2-M10.1002/1522-2683(200211)23:21<3659::AID-ELPS3659>3.0.CO;2-M ↗
13. 10.1021/ac001424e10.1021/ac001424e ↗
14. 10.1016/S0021‐9673(02)00075‐410.1016/S0021‐9673(02)00075‐4 ↗
15. 10.1016/j.chroma.2005.07.05510.1016/j.chroma.2005.07.055 ↗
16. 10.1016/j.talanta.2004.02.01310.1016/j.talanta.2004.02.013 ↗
17. Microfluidic separation and electrochemical detection of serotonin using a portable lab‐on‐a‐chip device
18. 10.1002/elps.20130006010.1002/elps.201300060 ↗
19. 10.1002/elps.20150009210.1002/elps.201500092 ↗
20. 10.1039/c2ay26166a10.1039/c2ay26166a ↗
21. 10.1016/j.nano.2006.12.00110.1016/j.nano.2006.12.001 ↗
22. 10.1002/elps.20120042610.1002/elps.201200426 ↗
23. 10.1021/es901555310.1021/es9015553 ↗
24. 10.1016/j.snb.2016.02.04910.1016/j.snb.2016.02.049 ↗
25. 10.1016/j.snb.2014.01.04010.1016/j.snb.2014.01.040 ↗