Modelling and calibration of microthermal sensor for underwater wall shear stress measurement
Source abstract
Abstract
To measure underwater wall shear stress, a calibration method for a microthermal sensor is studied. In CC (constant current) mode, a mathematical model which quantitatively describes the relationship between the output of the thermal sensor and the wall shear stress is presented. A well‐characterised water tunnel was set up to provide a series of known wall shear stresses in the range of 0–5 Pa. Calibration results prove the proposed model and characterise the thermal sensor precisely. The consistency of the parameters indicates that the performance of the thermal sensor keeps stable in water while driven by electricity. The R ‐squares in all calibrations exceed 0.995. The deviations between the predicted and the generated wall shear stress are about 5%.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Advanced Fiber Optic Sensors · Photoacoustic and Ultrasonic Imaging · Ultrasonics and Acoustic Wave Propagation
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Key Laboratory of Micro/Nano Systems for Aerospace Ministry of Education Northwestern Polytechnical University Xi'an Shaanxi 710072 People's Republic of China
References (17)
Show source-provided references
1. SchetzJ.A.: ‘Direct measurement of skin friction in complex flows using movable wall elements’.24th Aerodynamic Measurement Technology and Ground Testing Conf. Portland OR AIAA Paper.2004 211210.2514/6.2004-2112 ↗
2. 10.2514/2.163910.2514/2.1639 ↗
3. BaoX.BadescuM.Bar‐CohenY.et al.: ‘Sensor for direct measurement of the boundary shear stress in fluid flow’.SPIE Smart Structures and Materials Nondestructive Evaluation and Health Monitoring. International Society for Optics and Photonics 2011 pp.79814G–79814G‐810.1117/12.880901 ↗
4. 10.1016/j.flowmeasinst.2012.11.00410.1016/j.flowmeasinst.2012.11.004 ↗
5. 10.1088/0960‐1317/17/4/00210.1088/0960‐1317/17/4/002 ↗
6. LinQ.JiangF.WangX.Q.et al.: ‘MEMS thermal shear‐stress sensors: experiments theory and modeling’.Technical Digest of 2000 Solid‐State Sensor and Actuator Workshop (Hilton Head 2000) 2000 pp.304–30710.31438/trf.hh2000.74 ↗
7. 10.1063/1.189662110.1063/1.1896621 ↗
8. 10.2514/2.175810.2514/2.1758 ↗
9. 10.1109/JMEMS.2005.85664410.1109/JMEMS.2005.856644 ↗
10. 10.1109/84.74940810.1109/84.749408 ↗
11. SenzhiS.YuetingX.HuijingY.: ‘Hot Wire/Hot Film Anemometer’ 2003
12. Goldstein R.J. · 1996
13. Practical hot‐wire anemometer excitation modes
14. 10.1063/1.113727910.1063/1.1137279 ↗
15. 10.1115/1.281736610.1115/1.2817366 ↗
16. 10.1063/1.115017010.1063/1.1150170 ↗
17. Senzhi S. · 1987