journal article·2012-06

Design and simulation of a thermo transfer type MEMS based micro flow sensor for arterial blood flow measurement

Kalyan Kumar Mistry, Abhijit Mahapatra

Microsystem Technologies· Volume 18· Issue 6
Publisher page

Abstract

Thermo transfer type MEMS (Micro Electro Mechanical System) based micro flow sensing device have promising potential to solve the limitation of implantable arterial blood flow rate monitoring. The present paper emphasizes on modeling and simulation of MEMS based micro flow sensing device, which will be capable of implantable arterial blood flow rate measurement. It describes the basic design and model architecture of thermal type micro flow sensor. A pair of thin film micro heaters is designed through MEMS micro machining process and simulated using CoventorWare; a finite element based numerical code. A rectangular cross section micro channel has been modeled where in micro heater and thermal sensors are embedded using the same CoventorWare tools. Some promising and interesting results of thermal dissipation depending upon very small amount of flow rate through the micro channel are investigated. It is observed that measuring the variation of temperature difference between downstream and upstream, the variation of fluid flow rate in the micro channel can be measured. The numerical simulation results also shows that the temperature distribution profile of the heated surface depends upon microfluidic flow rate i.e. convective heat transfer is directly proportional to the microfluidic flow rate on the surface of the insulating membrane. The simplified analytical model of the thermo transfer type flow sensor is presented and verified by simulation results, which are very promising for application in arterial blood flow rate measuring in implantable micro devices for continuous monitoring of cardiac output.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Advanced Sensor Technologies Research · Infrared Thermography in Medicine · Photoacoustic and Ultrasonic Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (36)

Show source-provided references
1. Aarnoudse W, van’t Veer M, Pijls NHJ, ter Woorst J, Vercauteren S, Tonino P, Geven M, Rutten M, van Hagen E, de Bruyne B, van de Vosse F (2007) Direct volumetric blood flow measurement in coronary arteries by thermodilution. J Am Coll Cardiol 50(24):2294–2304. doi: 10.1016/j.jacc.2007.08.04710.1016/j.jacc.2007.08.047 ↗
2. Baroncini M, Placidi P, Cardinali GC, Scorzoni A (2004) Thermal characterization of a microheater for micromachined gas sensors. Sens Actuators, A 115:8–14. doi: 10.1016/j.sna.2004.03.01210.1016/j.sna.2004.03.012 ↗
3. Bolz A, Lang V, Merkely B, Schaldach M (1997) First results of an implantable sensor for blood flow measurement. In: Proceedings 19th international conference on IEEE/EMBS, Chicago, pp 2341–234310.1109/IEMBS.1997.758834 ↗
4. Bork T, Hogy A, Lempen M, Muller D, Joss D, Bardyn T, Buchler P, Keppner H, Barun S, Tardy Y, Burgen J (2010) Development and in vitro characterization of an implantable flow sensing transducer for hydrocephalus. Biomed Microdevice 12(4):607–61810.1007/s10544-010-9413-6 ↗
5. Chung CK, Chen TC, Shihi TR, Chang WT (2009) Fabrication of a novel micro liquid flow sensor using a TaN thin film. In: Proceedings IEEE international conference on nano/micro engineering molecular systems, China10.1109/NEMS.2009.5068716 ↗
6. DeHennis AD, Wise KD (2006) A fully-integrated multi-site pressure sensor for wireless arterial flow characterization. J Microelectromech Syst 15(3):678–685. doi: 10.1109/JMEMS.2006.87666810.1109/JMEMS.2006.876668 ↗
7. Delaunois AL (1973) Thermal method for continuous blood-velocity measurements in large blood vessels, and cardiac output determination. Medical Biological Eng 11:201–20510.1007/BF02478152 ↗
8. Elwenspoek M (1999) Thermal flow micro sensors. Mechanical Microsensors, published at Springer IEEE 199910.1109/SMICND.1999.810580 ↗
9. Ernst H, Jachimowicz A, Urban GA (2002) High resolution flow characterization in Bio-MEMS. Sensors Actuators A 100:54–6210.1016/S0924-4247(02)00187-5 ↗
10. Haga Y, Esashi M (2004) Biomedical microsystems for minimally invasive diagnosis and treatment. Proc IEEE 92(1):98–114. doi: 10.1109/JPROC.2003.82054510.1109/JPROC.2003.820545 ↗
11. Jayanthy AK, Sujatha N, Reddy MR (2011) Measuring blood flow: techniques and applications. IJRRAS 6(2):203–216
12. Kahroba P, Mirzaee I, Sharifi P, Shirvani H (2008) The microcavity-based micro-heater: an optimum design for micro-heaters. Microsyst Technol 14:705–710. doi: 10.1007/s00542-007-0418-x10.1007/s00542-007-0418-x ↗
13. Kamiunten S, Nagayo H, Motosuke M, Honami S (2010) Coupled electrothermal analysis of a micro flow sensor with control circuit using spice. Electron Commun Japan 93(1):58–64. doi: 10.1002/ecj.1009510.1002/ecj.10095 ↗
14. Khanna VK, Prasad M, Dwivedi VK, Shekhar C, Pankaj AC, Basu J (2007) Design and electro thermal simulation of polysilicon microheater on a suspended membrane for use in gas sensing. Indian J Pure Appl Phys 45:332–335
15. Kim TH, Kim SJ (2006) Development of a micro-thermal flow sensor with thin-film thermocouples. J Micromech Microeng 16:2502–2508. doi: 10.1088/0960-1317/16/11/03510.1088/0960-1317/16/11/035 ↗
16. Konzelmann U, Hecht H, Lembke M (1995) Breakthrough in reverse flow detection a new mass air flow meter using micro silicon technology. Int Congress Exposition Detroit MI. doi: 10.4271/950433
17. Liu C, Luo X, Zhang Z, Wang X, Liu S (2010) A hot platinum thin film anemometer. In: IEEE international conference on electronic packaging technology and high density packaging
18. Locke SE, Gale TJ, Kilpatrick D (2004) Development of an implantable blood flow and pressure monitor for pulmonary hypertension. Comput Cardiol 31:713–71610.1109/CIC.2004.1443038 ↗
19. Locke SE, Gale TJ, Kilpatrick D (2005) Implantable blood flow measurement techniques for humans. In: IEEE proceeding engineering in medicine and biology 27th annual conference, Shanghai10.1109/IEMBS.2005.1615732 ↗
20. Ma RH, Wang YH, Chiang SL, Lee CY (2011) Fabrication and characterization of MEMS-based flow sensors based on hot films. Microsyst Technol 17:655–660. doi: 10.1007/s00542-010-1156-z10.1007/s00542-010-1156-z ↗
21. Meng E, Li PY, Tai YC (2008) A biocompatible Parylene thermal flow sensing array. Sensors Actuators A 144:18–28. doi: 10.1016/j.sna.2007.12.01010.1016/j.sna.2007.12.010 ↗
22. Ohlsson A, Kubo SH, Steinhaus D, Connelly DT, Adler S, Bitkover C, Nordlander R, Ryde L, Bennett T (2001) Continuous ambulatory monitoring of absolute right ventricular pressure and mixed venous oxygen saturation in patients with heart failure using an implantable haemodynamic monitor. Eur Heart J 22(11):942–954. doi: 10.1053/euhj.2000.241910.1053/euhj.2000.2419 ↗
23. Pijls NHJ, Bruyne BD, Smith L, Aarnoudse W, Barbato E, Bartunek J, Bech GW, Vosse FD (2002) Coronary thermodilution to assess flow reserve validation in humans. J Am Hear Assoc. doi: 10.1161/01.CIR.0000017199.09457.3D
24. Rasmussen R, Mavriplis C, Zaghloul ME, Mikulchenko O, Mayaram K (2001) Simulation and optimization of a micro fluidic fow sensor. Sensors Actuators A 88:121–13210.1016/S0924-4247(00)00503-3 ↗
25. Receveur RAM, Lindemans FW, de Rooij NF (2007) Microsystem technologies for implantable applications. J Micromech Microeng 17:R50–R80. doi: 10.1088/0960-1317/17/5/R0210.1088/0960-1317/17/5/R02 ↗
26. Riet RV, Huysing JH (1976) Integrated direction-sensitive flow-meter. IEEE Electron Lett 12:647–64810.1049/el:19760497 ↗
27. Sapuppo F, Bucolo M, Paul C, Johnson PC, Fortuna L, Arena P (2007) An Improved instrument for real-time measurement of blood flow velocity in microvessels. IEEE Trans Instrum Meas 56(6):2663–2671. doi: 10.1109/TIM.2007.90795910.1109/TIM.2007.907959 ↗
28. Tan SS, Lam CM, Lim WE (2009) Experimental study on blood flow rate measurement using intravascular catheter. IEEE international conference biomedical and pharmaceutical engineering, ICBPE 0910.1109/ICBPE.2009.5384073 ↗
29. Terao M, Akutsu T, Tanaka Y (2007) Non-wetted thermal micro flow sensor. SICE annual conference, Kagawa University, Japan
30. Van Oudheusden BV (1992) Silicon thermal flow sensors. Sens Actuators, A 30(1–2):5–2610.1016/0924-4247(92)80192-6 ↗
31. Van Putten AFP (1983) An integrated silicon double bridge anemometer. SensActuators 4:387–396. doi: 10.1016/0250-6874(83)85049-5
32. Van Putten AFP, Middelhoek S (1974) An integrated silicon anemometer. Electron Lett IEEE 10(21):425–426. doi: 10.1049/el:1974033910.1049/el:19740339 ↗
33. Vennemann P, Lindken R, Westerweel J (2007) In vivo whole-field blood velocity measurement techniques. Exp Fluids 42:495–511. doi: 10.1007/s00348-007-0276-410.1007/s00348-007-0276-4 ↗
34. Wang L, Wang B, Lin Q (2008) Demonstration of MEMS-based differential scanning calorimetry for determining thermodynamic properties of biomolecules. Sens Actuators, B 134:953–958. doi: 10.1016/j.snb.2008.06.05910.1016/j.snb.2008.06.059 ↗
35. Wu S, Lin Q, Yuen Y, Tai YC (2001) MEMS flow sensors for nano-fluidic applications. Sens Actuators, A 89:152–15810.1016/S0924-4247(00)00541-0 ↗
36. Xue RF, Cheong JH, Cha HK, Liu X, Li P, Lim HJ, Lim LS, Cheng MY, He C, Park WT, Je M (2011) Ultra-low-power wireless implantable blood flow sensing microsystem for vascular graft applications. In: IEEE international symposium on integrated circuits10.1109/ISICir.2011.6131918 ↗