Development of a freestanding copper antiscatter grid using deep X-ray lithography
Source abstract
Abstract
A prototype freestanding copper antiscatter grid with parallel lamellar walls has been fabricated using deep X-ray lithography and electroforming. The freestanding copper grid has square shaped lamellar walls that are 25 μm thick and 1 mm tall with a 550 μm period. For mammography applications, the lamellar walls have to be aligned to a point X-ray source of the X-ray system. To achieve this goal, we investigated a dynamic double-exposure technique. Our progress in using stacked exposures and use of the photoresist SU-8, both to reduce fabrication cost, is also presented.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Radiation Shielding Materials Analysis · Advanced X-ray Imaging Techniques · Advanced X-ray and CT Imaging