journal article·2010-08

In situ diagnostic capabilities for beam position and beam intensity monitoring at SyLMAND

Venkat Subramanian, Sven Achenbach, David Klymyshyn, Garth Wells, Wade Dolton, Vinay Nagarkal, Brian Yates, Curtis Mullin, Martin Augustin

Microsystem Technologies· Volume 16· Issue 8-9
Publisher page

Abstract

SyLMAND, the Synchrotron Laboratory for Micro and Nano Devices at the Canadian Light Source, consists of a dedicated X-ray lithography beamline on a bend magnet port, and process support laboratories in a cleanroom environment. The beamline comprises a double mirror system with flat, chromium-coated silicon mirrors operated at varying grazing angles of incidence for spectral adjustment by high energy cut-off. We present in this paper, the in situ diagnostic components inside the vacuum vessel upstream and downstream of the mirrors that allow for monitoring the incident beam and the reflected beam after first, second, and both mirrors. Four fly wire systems are used for beam position monitoring and intensity measurements. Additionally, four detector plates mounted on and moving with the mirror bodies are used to determine the position of the mirror surfaces with respect to the beam. First experimental results verify the capabilities of the system by showing good agreement between measured and calculated data.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Advanced Surface Polishing Techniques · Advancements in Photolithography Techniques · Advanced X-ray Imaging Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (11)

Show source-provided references
1. Achenbach S, Klymyshyn D, Subramanian V (2005) Conceptual design report for SyLMAND “Synchrotron laboratory for micro and nano devices”. The Canadian Light Source Inc., CLSI 32.2.1.1, Saskatoon, Canada
2. Achenbach S, Subramanian V, Klymyshyn D, Wells G (2009a) Synchrotron laboratory for micro and nano devices—facility concept and design. Microsyst Technol (accepted for publication)10.1007/s00542-010-1071-3 ↗
3. Achenbach S, Klymyshyn D, Subramanian V, Reuther F, Mullin C, Wells G, Nagarkal V (2009b) An ultra-large-area Beryllium vacuum window for the X-ray lithography beamline SyLMAND at the Canadian light source. Nucl Instr Methods in Phys Research A (NIMS A)
4. Becker EW, Ehrfeld W, Hagmann P, Maner A, Münchmeyer D (1986) Fabrication of microstructures with high aspect ratios and great structural heights by synchrotron radiation lithography, galvanoforming, and plastic moulding (LIGA process). Microel Eng 4(1):36–5610.1016/0167-9317(86)90004-3 ↗
5. Chou MC, Pan CT, Wu TT, Wu CT (2008) Study of deep X-ray lithography behaviour for microstructures. Sens Act A (141):703–71110.1016/j.sna.2007.10.009 ↗
6. Fajardo P, Ferrer S (1995) Ultrahigh-vacuum-compatible position and shape monitor for high brilliance synchrotron radiation beams. Rev Sci Instrum 66(2):1882–188410.1063/1.1145813 ↗
7. Feiertag G, Ehrfeld W, Lehr H, Schmidt A, Schmidt M (1997) Calculation and experimental determination of the structure transfer accuracy in deep X-ray lithography. J Micromech Microeng (7):323–33110.1088/0960-1317/7/4/008 ↗
8. Lai B, Cerrina F (1986) SHADOW—a synchrotron radiation ray tracing program. Nucl Instr Methods Phys Res A 246:337–34110.1016/0168-9002(86)90101-4 ↗
9. Nagarkal V, Mullin C, Achenbach S, Subramanian V, Wells G (2008) Mechanical design of a radiatively cooled intensity chopper. In: Proc 5th Int Workshop Mech Eng Des Synchrotron Radiat Equip Instrum (MEDSI), Saskatoon, Canada, June 10–13:105
10. Sanchez del Rio M, Dejus RJ (1997) XOP: a multiplatform graphical user interface for synchrotron radiation spectral and optics calculations. SPIE Proc 3152:148–15710.1117/12.295554 ↗
11. Subramanian V, Achenbach S (2006) Preliminary design report for SyLMAND “Synchrotron Laboratory for Micro and Nano Devices”, Canadian Light Source, CLSI 32.2.69.1, Saskatoon, Canada