Design of a detachable bronchoscope for endotracheal intubation
Source abstract
Abstract
Abstract A novel detachable bronchoscope with a 3D printed micro-transmission is investigated. The micro-transmission controls the motion of the tip of the bronchoscope and is embedded in one end of the flexible insertion section of the bronchoscope. The micro-transmission allows manipulation of the distal end through the translation of guide nuts that are connected to the tip of the bronchoscope through so-called “angulation” wires. The displacement of the guide nuts on the left and right-handed screw sections and the control of the distal end of the bronchoscope are investigated. The maximum tensile force on the angulation wire before failure is determined experimentally for different screw pitches.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Medical Device Sterilization and Disinfection · Airway Management and Intubation Techniques · Anesthesia and Sedative Agents
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (9)
Show source-provided references
1. Ball RS (2019) Theory of screws: a study in the dynamics of a rigid body. Creative Media Partners, LLC
2. Carlson TP, Moen CW, Soli TD, Greminger MA, Martinelli JJ (2015) Detachable bronchoscope with a disposable insertion tube. ASME. J Med Devices 9(2):02093310.1115/1.4030189 ↗
3. Holland P, Shoop N (2000) Flexible endoscopes: structure and function–the air and water system. Gastroenterol Nurs 23(6):264–26810.1097/00001610-200011000-00004 ↗
4. Hunter IW, Chen Y, Oliverira JM (2015) U.S. Patent No. US 2015/0011830 A1. Washington, DC: U.S. Patent and Trademark Office
5. Kohanfars M, Li Y, Morris K, Ahadian F, Talke F (2021) Design of a 3D printed micro-transmission system for a detachable bronchoscope. Inform Storage Process Syst 1(1):3. https://doi.org/10.1115/ISPS2021-6536710.1115/ISPS2021-65367 ↗
6. Krupa JR, Leominster M (2008) Endoscope with Detachable Elongation Portion, Provisional application No. 60/879,718
7. Marini JL Abington M (1999) Endoscopic Instrument with Removable Front End, 6,077,290j
8. Sanovas (2017). Sanovas awarded seminal patent for imaging stylet for intubation. Sanovas. https://sanovas.com/index.php/about-us/press-releases/1052-sanovas-awarded-seminal-patent-for-imaging-stylet-for-intubation
9. Sofi K, El-Gammal K (2010) Endotracheal tube defects: Hidden causes of airway obstruction. Saudi J Anesthesia 4(2):108–11010.4103/1658-354X.65123 ↗