Abstract

The use of organic electronics for radiation sensing applications has become increasingly popular in recent years. The inherently low atomic number of the organic components of these devices is cited as motivation for their exploration as tissue-equivalent dosimeters. In this work we fabricate organic thin film transistors (OTFTs) with a polymer electret layer as part of the gate dielectric stack on flexible polyethylene terephthalate (PET) substrates to get a lower average Z, lighter weight, and flexible device. The response of these devices is evaluated by comparing to the same device made on silicon substrates that we have made in the past. To evaluate dose-rate response these devices were irradiated with a 6 MV photon beam with dose rates of 60–600 MU/min. Energy dependence was evaluated by irradiating with 100, 180, and 300 kVp photon beams from an orthovoltage unit and 6 and 10 MV beams from a linear accelerator. Real-time readout was achieved by measuring the drain current with the gate and drain voltage at −15 V. Silicon and PET devices both demonstrated excellent linearity with dose rate. Si devices had a sensitivity of 79.7 nA/Gy and PET devices had a sensitivity of 33.4 nA/Gy. At 100 kVp Si and PET both had an increase in signal compared to at 6 MV of 2.46 and 4.76 times respectively. The larger energy dependence in PET devices is likely due to the high atomic number of the Indium–Tin-Oxide and gold layers. These devices on PET need improvements in sensitivity and energy dependence to achieve similar or better performance than their Si counterparts.
Source: expertdata · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Analytical Chemistry and Sensors · Radiation Effects in Electronics · Conducting polymers and applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (20)

Show source-provided references
1. Radiation dose management in nuclear power plants10.1016/S0149-1970(01)00002-6 ↗
2. X-ray based methods for non-destructive testing and material characterization10.1016/j.nima.2008.03.016 ↗
3. In vivo dosimetry in external beam radiotherapy10.1118/1.4811216 ↗
4. Comparison study of MOSFET detectors and diodes for entrance in vivo dosimetry in 18 MV x-ray beams10.1118/1.1785452 ↗
5. Energy dependence corrections to MOSFET dosimetric sensitivity10.1007/BF03178623 ↗
6. The response of a MOSFET, p-type semiconductor and LiF TLD to quasi-monoenergetic x-rays10.1088/0031-9155/42/12/006 ↗
7. Dose response of various radiation detectors to synchrotron radiation10.1088/0031-9155/43/11/006 ↗
8. Toward low-voltage and bendable X-ray direct detectors based on organic semiconducting single crystals10.1002/adma.201503090 ↗
9. Enhanced x-ray detection sensitivity in semiconducting polymer diodes containing metallic nanoparticles10.1088/0022-3727/46/27/275102 ↗
10. Direct x-ray detection with conjugated polymer devices10.1063/1.2748337 ↗
11. Characterization of thick film poly(triarylamine) semiconductor diodes for direct x-ray detection10.1063/1.3225909 ↗
12. X-Ray detectors with ultrahigh sensitivity employing high performance transistors based on a fully organic small molecule semiconductor/polymer blend active layer10.1002/aelm.202200293 ↗
13. Radiation induced photocurrent in the active volume of P3HT/PCBM BHJ photodiodes10.1016/j.orgel.2020.105890 ↗
14. Characterization of an organic semiconductor diode for dosimetry in radiotherapy10.1002/mp.14229 ↗
15. Berggren · Org. Bioelectron. Adv. Mater. · 2007 · 19 · 3201
16. Preliminary characterization of the response of an organic field effect transistor to ionizing radiation10.1016/j.radmeas.2018.08.008 ↗
17. Organic field-effect transistors as flexible, tissue-equivalent radiation dosimeters in medical applications10.1002/advs.202001522 ↗
18. A highly sensitive, direct X-ray detector based on a low-voltage organic field-effect transistor10.1002/aelm.201600409 ↗
19. Low-operating-voltage operation and improvement in sensitivity with passivated OFET sensors for determining total dose radiation10.1109/LED.2010.2074179 ↗
20. Preliminary evaluation of pentacene field effect transistors with polymer gate electret as ionizing radiation dosimeters10.3390/app112311368 ↗