journal article·2000-06

Sieve Calibration - a New Simple but High Precision Approach

Graham R. Rideal, Jamie Storey, Thomas R. Morris

Publisher page

Abstract

This paper describes the preparation, measurement and use of microspheres for calibrating individual test sieves. Using a 63 µm sieve as an example, 2.5 million apertures, or 80% of the surface is examined in under 2 minutes. Because of the narrowness of the size distribution, a 5% difference in percent passing only results in a mean aperture difference of 1 µm, indeed, the measurement uncertainty for all 24 tests performed was only 0.7 µm. The calibration is independent of the method of shaking and can be used for most sieves.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Material Properties and Processing · Electrowetting and Microfluidic Technologies · Microwave and Dielectric Measurement Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References