Sieve Calibration - a New Simple but High Precision Approach
Source abstract
Abstract
This paper describes the preparation, measurement and use of microspheres for calibrating individual test sieves. Using a 63 µm sieve as an example, 2.5 million apertures, or 80% of the surface is examined in under 2 minutes. Because of the narrowness of the size distribution, a 5% difference in percent passing only results in a mean aperture difference of 1 µm, indeed, the measurement uncertainty for all 24 tests performed was only 0.7 µm. The calibration is independent of the method of shaking and can be used for most sieves.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Material Properties and Processing · Electrowetting and Microfluidic Technologies · Microwave and Dielectric Measurement Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References
Reference metadata was not supplied for this record.