journal article·1990-01

The Linearity and Response of Focussed Apertures

Mark P. Cowan, John G. Harfield

Publisher page

Abstract

Abstract Practical experience of particle size analysers using the Coulter principle shows that for measurements made over a range of 2% to 60% of the aperture diameter, the aperture responds linearly to particle volume at least for spheres. In this paper it is shown that the response of a focussed aperture is not significantly different from an unfocussed aperture. It is also shown that a simple linear equation of response successfully applied to a COULTER COUNTER® model ZM in a previous paper, can be used to predict the calibration constant of a new wide range instrument, the COULTER® MULTISIZER.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Microfluidic and Bio-sensing Technologies · Microbial Inactivation Methods · Cyclone Separators and Fluid Dynamics

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (10)

Show source-provided references
1. Harfield J. G. · 1988 · 151
2. J. G.Harfield R. T.Wharton R. W.Lines:Response of the COULTER COUNTER model ZM to spheres. Part. Charact. l (1984)32.10.1002/ppsc.19840010106 ↗
3. J. G.Harfield M.Cowan:Further investigations into the linearity of response in the Coulter aperture. Prepr. 1st World Congr. Part. Technol.Part 1 PARTEC Nürnberg April 16 to 18 1986 p.351–366.
4. 10.1007/BF0205292210.1007/BF02052922 ↗
5. P. J.Lloyd:The Response of the COULTER COUNTER to Cylindrical and Disc shaped Particles. Prepr. 2nd Eur. Symp. Particle Charact. PARTEC Nürnberg1979pp.694–703.
6. B. A.Batch:The Application of an Electronic Partical Counter to Size Analysis of Pulverised Coal and Fly Ash. J. Inst. Fuel(1964)455–461.
7. B.Scarlett:Theoretical Deviation of the Response of a COULTER COUNTER. Prepr. 2nd Eur. Symp. Particle Charact. PARTEC Nürnberg1979pp.681–693.
8. 10.1002/ppsc.1986003020210.1002/ppsc.19860030202 ↗
9. Definitive particle size distribution measurements using focussed flow Coulter apertures and Multichannel analyser
10. Bernard J. G. · 1988 · 135