Specimen temperature increases during transmission electron microscopy
Source abstract
Abstract
New measurements of temperature rises in electron irradiated transmission microscopy specimens have been made by employing AuNi thin film thermocouples prepared by vacuum evaporation onto thin mica discs. The extent of heating is proportional to the electron flux (beam current) and for typical 100 kV operating conditions in the microscope is less than 8 K. Under extreme conditions (100 μA) temperature increases on the order of 15 K were observed.