Measurement of γ-dose rates by n- and p-type semiconductor detectors
Source abstract
Abstract
n-Si detectors prepared of n-type silicon, as described in the literature, were used initially for the measurement of dose rates of a high activity 60Co radiation source. It was stated that the radiation induced a rapid “ageing” in an n-type detector leading to the decrease of sensitivity. On the basis of calculations of the photocurrent of semiconductor photovoltaic cells, a more radiation resistant detector is prepared using p-type silicon. Its sensitivity remained constant within ±10% up to a dose of 30 MR. This is sufficient for practical purposes. The photocurrent of the p-Si detector is linear in the range of dose rate between 50 and 107 R/h, using a recorder with a resistance of RT ≦ 1 kω. The directional dependence of the response is within ±20%, the temperature coefficient is 0.4% deg−1 between 20 and 80 °C.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Silicon and Solar Cell Technologies · Radiation Detection and Scintillator Technologies · Advanced Semiconductor Detectors and Materials
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.