journal article·2019-07Open access

Porous Ti-10Mo alloy fabricated by powder metallurgy for promoting bone regeneration

Wei Xu, Zhuo Liu, Xin Lu, Jingjing Tian, Gang Chen, Bowen Liu, Zhou Li, Xuanhui Qu, Cuie Wen

Science China Materials· Volume 62· Issue 7

Abstract

Porous Ti-10Mo alloys were fabricated by powder metallurgy using a space-holder method. The pore characteristics, microstructure, mechanical properties, in vitro biocompatibility, and in vivo osseointegration of the fabricated alloys were systematically investigated. The results show that with different weight ratios of the space-holder (NH 4 HCO 3 ) added, all of the porous Ti-10Mo alloys sintered at 1,300°C exhibited a typical Widmanstatten microstructure. The porosity and average pore size of the porous structures can be controlled in the range of 50.8%66.9% and 70.1381.4 μm, respectively. The Ti-10Mo alloy with 63.4% porosity exhibited the most suitable mechanical properties for implant applications with an elastic modulus of 2.9 GPa and a compressive yield strength of 127.5 MPa. In vitro, the alloyconditioned medium showed no deleterious effect on the cell proliferation. The cell viability in this medium was higher than that of the reference group, suggesting non-toxicity and good biological characteristics of the alloy specimens. In vivo, after eight weeks implantation, new bone tissue formed surrounding the alloy implants, and no noticeable inflammation was observed at the implantation site. The bone bonding strength of the porous Ti-10Mo alloy increased over time from 46.6 N at two weeks to 176.4 N at eight weeks. Suitable mechanical properties together with excellent biocompatibility in vitro and osteointegration in vivo make the porous Ti-10Mo fabricated by powder metallurgy an attractive orthopedic implant alloy.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Titanium Alloys Microstructure and Properties · Bone Tissue Engineering Materials · Orthopaedic implants and arthroplasty

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (52)

Show source-provided references
1. Geetha M, Singh AK, Asokamani R, et al. Ti based biomaterials, the ultimate choice for orthopaedic implants—A review. Prog Mater Sci, 2009, 54: 397–42510.1016/j.pmatsci.2008.06.004 ↗
2. Long M, Rack HJ. Titanium alloys in total joint replacement—a materials science perspective. Biomaterials, 1998, 19: 1621–163910.1016/S0142-9612(97)00146-4 ↗
3. Banerjee D, Williams JC. Perspectives on titanium science and technology. Acta Mater, 2013, 61: 844–87910.1016/j.actamat.2012.10.043 ↗
4. Niinomi M, Nakai M, Hieda J. Development of new metallic alloys for biomedical applications. Acta Biomater, 2012, 8: 3888–390310.1016/j.actbio.2012.06.037 ↗
5. Nune KC, Li S, Misra RDK. Advancements in three-dimensional titanium alloy mesh scaffolds fabricated by electron beam melting for biomedical devices: mechanical and biological aspects. Sci China Mater, 2018, 61: 455–47410.1007/s40843-017-9134-x ↗
6. Aksakal B, Yildirim ÖS, Gul H. Metallurgical failure analysis of various implant materials used in orthopedic applications. J Fail Anal Preven, 2004, 4: 17–2310.1007/s11668-996-0007-9 ↗
7. Chang B, Song W, Han T, et al. Influence of pore size of porous titanium fabricated by vacuum diffusion bonding of titanium meshes on cell penetration and bone ingrowth. Acta Biomater, 2016, 33: 311–32110.1016/j.actbio.2016.01.022 ↗
8. Ho WF, Ju CP, Chern Lin JH. Structure and properties of cast binary Ti-Mo alloys. Biomaterials, 1999, 20: 2115–212210.1016/S0142-9612(99)00114-3 ↗
9. Zhao X, Niinomi M, Nakai M, et al. Beta type Ti-Mo alloys with changeable Young’s modulus for spinal fixation applications. Acta Biomater, 2012, 8: 1990–199710.1016/j.actbio.2012.02.004 ↗
10. González JEG, Mirza-Rosca JC. Study of the corrosion behavior of titanium and some of its alloys for biomedical and dental implant applications. J Electroanal Chem, 1999, 471: 109–11510.1016/S0022-0728(99)00260-0 ↗
11. Kumar S, Narayanan TSNS. Corrosion behaviour of Ti-15Mo alloy for dental implant applications. J Dentistry, 2008, 36: 500–50710.1016/j.jdent.2008.03.007 ↗
12. Yan M, Qian M, Kong C, et al. Impacts of trace carbon on the microstructure of as-sintered biomedical Ti-15Mo alloy and reassessment of the maximum carbon limit. Acta Biomater, 2014, 10: 1014–102310.1016/j.actbio.2013.10.034 ↗
13. ATI 15MoTM Titanium alloy technical data sheet. ATI Allvac, Monroe, NC
14. Jablokov V, Nutt M, Richelsoph M, et al. The application of Ti-15Mo beta titanium alloy in high strength structural orthopaedic applications. J ASTM Int, 2005, 2: 1303310.1520/JAI13033 ↗
15. Lin JH, Ju CP, Ho WF, inventors, J.H. Chern, assignee. Biocompatible low modulus titanium alloy for medical implant. United States patent, 6409852, 2002 Jun. 25
16. Disegi J. Wrought titanium-15% molybdenum implant material. SYNTHES® Instruments and Implants, Second Edition. 2009
17. Li Y, Wong C, Xiong J, et al. Cytotoxicity of titanium and titanium alloying elements. J Dent Res, 2010, 89: 493–49710.1177/0022034510363675 ↗
18. Zhou YL, Luo DM. Microstructures and mechanical properties of Ti-Mo alloys cold-rolled and heat treated. Mater Charact, 2011, 62: 931–93710.1016/j.matchar.2011.07.010 ↗
19. Cremasco A, Messias AD, Esposito AR, et al. Effects of alloying elements on the cytotoxic response of titanium alloys. Mater Sci Eng-C, 2011, 31: 833–83910.1016/j.msec.2010.12.013 ↗
20. Lee EB, Han MK, Kim BJ, et al. Effect of molybdenum on the microstructure, mechanical properties and corrosion behavior of Ti alloys. Int J Mater Res, 2014, 105: 847–85310.3139/146.111092 ↗
21. Liu Y, Zheng Y, Hayes B. Degradable, absorbable or resorbable— what is the best grammatical modifier for an implant that is eventually absorbed by the body? Sci China Mater, 2017, 60: 377–39110.1007/s40843-017-9023-9 ↗
22. Wang X, Xu S, Zhou S, et al. Topological design and additive manufacturing of porous metals for bone scaffolds and orthopaedic implants: A review. Biomaterials, 2016, 83: 127–14110.1016/j.biomaterials.2016.01.012 ↗
23. Vandendolder J, Farber E, Spauwen P, et al. Bone tissue reconstruction using titanium fiber mesh combined with rat bone marrow stromal cells. Biomaterials, 2003, 24: 1745–175010.1016/S0142-9612(02)00537-9 ↗
24. Marin E, Fusi S, Pressacco M, et al. Characterization of cellular solids in Ti6Al4V for orthopaedic implant applications: Trabecular titanium. J Mech Behav BioMed Mater, 2010, 3: 373–38110.1016/j.jmbbm.2010.02.001 ↗
25. Zardiackas LD, Parsell DE, Dillon LD, et al. Structure, metallurgy, and mechanical properties of a porous tantalum foam. J Biomed Mater Res, 2001, 58: 180–18710.1002/1097-4636(2001)58:2<180::AID-JBM1005>3.0.CO;2-5 ↗
26. Gao Z, Li Q, He F, et al. Mechanical modulation and bioactive surface modification of porous Ti-10Mo alloy for bone implants. Mater Des, 2012, 42: 13–2010.1016/j.matdes.2012.05.041 ↗
27. Li S, Li X, Hou W, et al. Fabrication of open-cellular (porous) titanium alloy implants: osseointegration, vascularization and preliminary human trials. Sci China Mater, 2018, 61: 525–53610.1007/s40843-017-9063-6 ↗
28. Correa VL, Garza KM, Murr LE. Vascularization in interconnected 3D printed Ti-6Al-4V foams with hydrogel matrix for biomedical bone replacement implants. Sci China Mater, 2018, 61: 565–57810.1007/s40843-017-9091-1 ↗
29. Xie F, He X, Lu X, et al. Preparation and properties of porous Ti-10Mo alloy by selective laser sintering. Mater Sci Eng-C, 2013, 33: 1085–109010.1016/j.msec.2012.11.037 ↗
30. Xie F, He X, Lv Y, et al. Selective laser sintered porous Ti-(4–10) Mo alloys for biomedical applications: Structural characteristics, mechanical properties and corrosion behaviour. Corros Sci, 2015, 95: 117–12410.1016/j.corsci.2015.03.005 ↗
31. Xie FX, He XB, Cao SL, et al. Structural characterization and electrochemical behavior of a laser-sintered porous Ti-10Mo alloy. Corros Sci, 2013, 67: 217–22410.1016/j.corsci.2012.10.036 ↗
32. Wong KC, Scheinemann P. Additive manufactured metallic implants for orthopaedic applications. Sci China Mater, 2018, 61: 440–45410.1007/s40843-017-9243-9 ↗
33. Xu W, Lu X, Zhang B, et al. Effects of porosity on mechanical properties and corrosion resistances of PM-fabricated porous Ti- 10Mo alloy. Metals, 2018, 8: 188–20110.3390/met8030188 ↗
34. ISO 10993-5: 1999. Biological evaluation of medical devices-part 5: tests for cytotoxicity: in vitro methods. ANSI/AAMI, Arlington, VA
35. Wen CE, Yamada Y, Shimojima K, et al. Novel titanium foam for bone tissue engineering. J Mater Res, 2002, 17: 2633–263910.1557/JMR.2002.0382 ↗
36. Kawai N, Niwa S, Sato M, et al. Bone formation by cells from femurs cultured among three-dimensionally arranged hydroxyapatite granules. J Biomed Mater Res, 1997, 37: 1–810.1002/(SICI)1097-4636(199710)37:1<1::AID-JBM1>3.0.CO;2-W ↗
37. Krishna BV, Bose S, Bandyopadhyay A. Low stiffness porous Ti structures for load-bearing implants. Acta Biomater, 2007, 3: 997–100610.1016/j.actbio.2007.03.008 ↗
38. Cameron HU, Pilliar RM, Macnab I. The rate of bone ingrowth into porous metal. J Biomed Mater Res, 1976, 10: 295–30210.1002/jbm.820100210 ↗
39. Ryan GE, Pandit AS, Apatsidis DP. Porous titanium scaffolds fabricated using a rapid prototyping and powder metallurgy technique. Biomaterials, 2008, 29: 3625–363510.1016/j.biomaterials.2008.05.032 ↗
40. Li Y, Wen C, Mushahary D, et al. Mg-Zr-Sr alloys as biodegradable implant materials. Acta Biomater, 2012, 8: 3177–318810.1016/j.actbio.2012.04.028 ↗
41. Xu W, Li M, Wen C, et al. The mechanical properties and in vitro biocompatibility of PM-fabricated Ti-28Nb-35.4Zr alloy for orthopedic implant applications. Materials, 2018, 11: 53110.3390/ma11040531 ↗
42. Oliveira NTC, Guastaldi AC. Electrochemical behavior of Ti-Mo alloys applied as biomaterial. Corros Sci, 2008, 50: 938–94510.1016/j.corsci.2007.09.009 ↗
43. Zhou YL, Luo DM. Corrosion behavior of Ti-Mo alloys cold rolled and heat treated. J Alloys Compd, 2011, 509: 6267–627210.1016/j.jallcom.2011.03.045 ↗
44. Chen J, Paetzell E, Zhou J, et al. Osteoblast-like cell ingrowth, adhesion and proliferation on porous Ti-6Al-4V with particulate and fiber scaffolds. Mater Sci Eng-C, 2010, 30: 647–65610.1016/j.msec.2010.01.005 ↗
45. Xu W, Lu X, Wang LN, et al. Mechanical properties, in vitro corrosion resistance and biocompatibility of metal injection molded Ti-12Mo alloy for dental applications. J Mech Behav BioMed Mater, 2018, 88: 534–54710.1016/j.jmbbm.2018.08.038 ↗
46. Huang HH, Wu CP, Sun YS, et al. Surface nanoporosity of β-type Ti-25Nb-25Zr alloy for the enhancement of protein adsorption and cell response. Surf Coatings Tech, 2014, 259: 206–21210.1016/j.surfcoat.2014.02.037 ↗
47. Okazaki Y, Nishimura E, Nakada H, et al. Surface analysis of Ti-15Zr-4Nb-4Ta alloy after implantation in rat tibia. Biomaterials, 2001, 22: 599–60710.1016/S0142-9612(00)00221-0 ↗
48. Lin DJ, Chuang CC, Chern Lin JH, et al. Bone formation at the surface of low modulus Ti-7.5Mo implants in rabbit femur. Biomaterials, 2007, 28: 2582–258910.1016/j.biomaterials.2007.02.005 ↗
49. Majumdar P, Singh SB, Chakraborty M. The role of heat treatment on microstructure and mechanical properties of Ti-13Zr-13Nb alloy for biomedical load bearing applications. J Mech Behav BioMed Mater, 2011, 4: 1132–114410.1016/j.jmbbm.2011.03.023 ↗
50. Muller D, Chim H, Bader A, et al. Vascular guidance: microstructural scaffold patterning for inductive neovascularization. Stem Cells Int, 2011, 2011: 1–610.4061/2011/547247 ↗
51. Bandyopadhyay A, Espana F, Balla VK, et al. Influence of porosity on mechanical properties and in vivo response of Ti6Al4V implants. Acta Biomater, 2010, 6: 1640–164810.1016/j.actbio.2009.11.011 ↗
52. Caparrós C, Guillem-Martí J, Molmeneu M, et al. Mechanical properties and in vitro biological response to porous titanium alloys prepared for use in intervertebral implants. J Mech Behav BioMed Mater, 2014, 39: 79–8610.1016/j.jmbbm.2014.05.029 ↗