High-power cold diodes for the protection of a 1.5 T superconducting MRI magnet system
Source abstract
Abstract
Abstract Cold diodes play a crucial role in various operational aspects of the superconducting magnet of magnetic resonance imaging (MRI) scanners. The forward and reverse characteristics and failure modes are studied for two types of high-power rectifier diode to find their suitability for application at 4.2 K in a whole-body 1.5 T superconducting MRI magnet system. The diodes are characterized at various temperatures (300 K, 77 K and 4.2 K) in the current range up to 600 A. The turn-on voltage of the diodes at 4.2 K is found to be increased by 7–12 times that of room temperature. The diodes show negative differential resistance (NDR) beyond 1 A of forward current at 4.2 K. The forward voltage stabilizes in the range of 1.22–1.24 V after attaining the turn-on voltage in the range of 5–8 V. The significance of NDR is analysed for the MRI magnet. The long-term stability of the forward voltage is studied at 77 K with 550 A of forward current. Two modes of diode failure are also studied to foresee any failure during the operation of the magnet.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Physics of Superconductivity and Magnetism · Silicon Carbide Semiconductor Technologies · Quantum and electron transport phenomena
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (19)
Show source-provided references
1. Wilson · 1983
2. Development of high homogeneity and high stability 1.5T superconducting magnet for whole-body MRI scanner10.5958/2349-2120.2019.00034 ↗
3. Quench protection system of a whole-body non-impregnated 1.5 T superconducting MRI magnet10.1109/TASC.2021.3076402 ↗
4. A 2-tesla active shield magnet for whole-body imaging and spectroscopy10.1109/20.133510 ↗
5. Design and quench protection analysis of a 9 T NbTi superconducting magnet for a large-bore EMPS10.1109/tasc.2020.2972216 ↗
6. Quench detection and protection of the wide bore 900 MHz NMR magnet at the national high magnetic field laboratory10.1109/tasc.2004.830547 ↗
7. Quench protection for the MICE cooling channel coupling magnet10.1109/tasc.2009.2018054 ↗
8. First implementation of the CLIQ quench protection system on a full-scale accelerator quadrupole magnet10.1109/tasc.2016.2529840 ↗
9. Superconducting magnet quench protection analysis and design10.1109/20.133573 ↗
10. Protection circuit and quench simulation of a 7 T superconducting magnet for animal MRI10.1109/tasc.2012.2232959 ↗
11. Quench protection design and simulation of a 7 tesla superconducting magnet10.1109/tasc.2010.2044783 ↗
12. The CLIQ quench protection system applied to the 16 T FCC-hh dipole magnets10.1109/tasc.2019.2930705 ↗
13. Reduced mass persistent switches for large superconducting magnets in space10.1109/77.233772 ↗
14. Quench protection of persistent current switches using diodes at cryogenic temperature
15. Voltage characteristics of diodes used for passive quench protection of low-current magnets10.1109/tasc.2014.2362574 ↗
16. Performance of the cold powered cold diodes and diode leads in the main magnets of the LHC10.1088/1757-899X/101//1/012076 ↗
17. Testing of high current by-pass diodes for the LHC magnet quench protection10.1109/20.511530 ↗
18. Overview of catastrophic failures of freewheeling diodes in power electronics circuits10.1016/j.microrel.2013.07.126 ↗
19. Rectifier diodes—power semiconductors. Dynex semiconductor. Dynex