Detection of Mercury and Copper Ions Using Surface Plasmon Resonance Optical Sensor
Source abstract
Abstract
Key words: surface plasmon resonance, mercury and copper ions, chitosan, glutaraldehyde Mercury and copper ions, Hg 2+ and Cu 2+ , can be detected by measuring surface plasmon resonance signals with a thin chitosan layer deposited on a gold film. An amount of 0.55 ml of chitosan cross-linked glutaraldehyde solution was spin coated onto a glass cover slip at 6000 rev./min for 30 s. Changes in the resonance angle () are directly proportional to the concentration of heavy metal ions in solution (0.5-100 ppm). The sensitivities to Hg 2+ and Cu 2+ are 0.00743 and 0.00654 ppm -1 , respectively. The gold/chitosan interface is highly sensitive to Hg 2+ and Cu 2+ with detection limits as low as 500 ppb.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Analytical Chemistry and Sensors · Advanced biosensing and bioanalysis techniques · Biosensors and Analytical Detection
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References
Reference metadata was not supplied for this record.