Development of Optical-fiber-based Neutron Detector Using Li-glass Scintillator for an Intense Neutron Field

Akihisa Ishikawa, Atsushi Yamazaki, Kenichi Watanabe, Sachiko Yoshihashi, Akira Uritani, Yoshinori Sakurai, Hiroki Tanaka, Ryo Ogawara, Mitsuru Suda, Tsuyoshi Hamano

Sensors and Materials· Volume 32· Issue 4

Abstract

The development of a neutron detector for real-time measurements in an intense neutron field is desired. We are developing a quartz optical-fiber-based neutron detector as a real-time neutron monitor. Thus far, Eu:LiCaAlF 6 crystals and LiF/Eu:CaF 2 eutectics have been used as the scintillators of the quartz optical-fiber-based neutron detectors. These scintillators have high light yields, but relatively long decay time constants, which are the limiting factors for the dynamic range of optical-fiber-based neutron detectors. In this work, we applied a small Li glass scintillator to a quartz optical-fiber-based neutron detector using an optical fiber with high numerical aperture and a UV-curable resin with high transmittance. The fabricated detector showed a clear peak corresponding to neutron events in the pulse height distribution. The radiation resistance was evaluated and no deterioration was observed up to 2 10 12 n/cm 2 . The detector output linearity was confirmed up to 450 kcps, which was about ten times larger than that of the detector using the conventional Eu:LiCaAlF 6 or LiF/Eu:CaF 2 scintillator.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Radiation Detection and Scintillator Technologies · Atomic and Subatomic Physics Research · Nuclear Physics and Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References