Heat-transfer-printed Glucose Biosensor for Use in Diapers
Source abstract
Abstract
Wearable biosensors that can monitor health conditions are currently attracting attention. In this study, we investigated a method to fabricate a glucose biosensor by heat transfer printing. A screen-printed glucose oxidase-modified electrode formed on a polyethylene terephthalate (PET) substrate was transferred onto a nonwoven fabric. Conditions for thermal transfer while retaining enzyme activity were investigated, and electrochemical evaluation of the glucose biosensor was conducted. Enzyme-catalyzed reactions were found to readily occur on the electrode surface after heat transfer printing. A linear relationship was obtained in the glucose concentration range from 0.1 to 20 mmol dm -3 . In general, the glucose concentration in urine is 3-6 mmol dm -3 in healthy persons and 7 mmol dm -3 or higher in diabetics. This fabrication method is expected to be widely applicable to future biosensors for monitoring urine sugar in diapers and other products.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Electrochemical sensors and biosensors · Advanced Sensor and Energy Harvesting Materials · Microfluidic and Capillary Electrophoresis Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References